HP 71210C Microwave Spectrum Analyzer, 100 Hz to 22 GHz![]()
|
As part of the Modular Measurement System (MMS) family, you can choose a configuration that best meets your requirements in performance, size, and price. Adding accessories such as a preamplifier or tracking generator is very easy.
For automated testing, the HP 70000 series of spectrum analyzers are compatible with the HP 8566B. Their proven reliability and repeatability allow a three-year calibration cycle, giving you a very low lifetime cost.
Frequency Specifications
Frequency Range: 100 Hz to 26.5 GHz Tunable in 1 Hz increments Frequency Readout Accuracy Span <=10 MHz x N1: ±[(freq readout x freq ref accuracy) +1.0% of span +10 Hz] Span >=10 MHz x N1 Sweep >= 20 ms: ±[(freq readout x freq ref accuracy) +1.5% of span +10 Hz] 10 ms <= sweep <20 ms: ±[(freq readout x freq ref accuracy) +2.5% of span +10 Hz] Frequency Span Range: 0 to 26.5 GHz in 0.5% increments Accuracy Span <10 MHz x N1: ±[1% of span + (span x freq ref accuracy)] Span >10 MHz x N1 Sweep >= 50 ms: ±[1.5% of span + (span x freq ref accuracy)] Sweep >= 20 ms: ±[2.5% of span + (span x freq ref accuracy)] Sweep >= 10 ms: ±[4.0% of span + (span x freq ref accuracy)] Frequency Reference Accuracy W/ HP 70310A W/O HP 70310A Aging: <1 x 10-7/year, <3 x 10-6/year <5 x 10-10/day (7-day avg.) Temperature Drift: <7 x 10-9 <1x10-5 Sweep Time Range: 10 ms to 1000 s, continuous with HP 70700A: Swept Frequency Spans: 15 ms to 355 s Fixed Frequency Zero Span: 80 µs to 355 s with 800 point trace Resolution Bandwidth (3 dB, synchronously tuned) Range: 10 Hz to 300 kHz (HP 70902A) (1, 3, 10 sequence and 100 kHz to 3 MHz (HP 70903A) 10% increments except 3 kHz to 10 kHz) Accuracy: ±20% Video Bandwidth Range: 3 Hz to 300 kHz (HP 70902A) (1, 3, 10 sequence) 300 Hz to 3 MHz (HP 70903A) When set to maximum (300 kHz or 3 MHz), effective bandwidth is greater than specified. Accuracy: ±20%, characteristic
Amplitude Specifications
Frequency Response Band 20 °C to 3 0 °C 0 °C to 50 °C Peak Variation 100 Hz to 2.9 GHz ±1.5 dB ±2.0 dB (10 dB attenuation): 2.7 to 22 GHz ±2.0 dB ±2.5 dB Referenced to 100 Hz to 2.9 GHz ±2.3 dB 300 MHz, -10 dBm: 2.7 GHz to 22 GHz ±3.3 dB Calibrator (10 dB attenuation) Calibrator Uncertainty: ±0.3 dB (-10 dBm, 300 MHz) Input Attenuator Range: 0 to 70 dB in 10 dB steps Switching Repeatability: ±0.2 dB IF Gain Accuracy Gain 20 °C to 30 °C 0 °C to 50 °C 10 dB ±0.2 dB ±0.2 dB 20 dB ±0.2 dB ±0.2 dB 30 dB ±0.2 dB ±0.3 dB 40 dB ±0.2 dB ±0.5 dB 50 dB ±0.2 dB ±0.6 dB 60 dB ±0.2 dB ±0.8 dB Scale Fidelity Bandwidth Fidelity Log Corrected HP 70902A 10 Hz ±0.7 dB (0 to 90 dB) 30 Hz to 100 kHz ±0.5 dB HP 70903A 300 kHz to 1 MHz ±0.5 dB (0 to 75 dB) 3 MHz ±0.7 dB Resolution Bandwidth: in 1, 3, 10 sequence Switching Repeatability ±0.1 dB, ±3 dB, uncorrected Marker Resolution: ±0.03 dB
Noise
Spectral Purity Freq Range Noise Sideband Offset 100 Hz - 2.9 GHz -108 dBc/Hz 10 kHz 2.7 - 6.2 GHz -108 dBc/Hz 30 kHz 6.0 - 12.8 GHz -102 dBc/Hz 30 kHz 12.6 - 22 GHz -96 dBc/Hz 30 kHz Line and System Related Sidebands <-65 dBc + 20 Log M1 Displayed Average Noise Level Frequency DANL 10 Hz Res BW, 0 dB 10 MHz - 2.9 GHz -139 dBm attn 3 Hz Video BW 2.7 GHz - 12.8 GHz -136 dBm Ref Level ł -85 dBm 12.6 GHz - 22 GHz -133 dBm with HP 70620B 1.0 - 2.9 GHz -155 dBm 2.7 - 12.8 GHz -153 dBm 12.6 - 22.0 GHz -150 dBm
False Signals
Spurious Responses Input Spurious for input level <=-40 dBm 100 Hz - 10 MHz <-60 dBc at the input mixer, 10 MHz - 22 GHz <-70 dBc except as listed, are less than these values (10 dB attn) Second Harmonic Distortion for mixer input Band Second Harmonic <=-40 dBm (10 dB attn) 100 Hz - 20 MHz <-60 dBc 20 MHz - 3.5 GHz <-70 dBc 3.5 GHz - 22 GHz <-100 dBc Third-Order Intermodulation Distortion for two signals, Center Intermod Equiv each <=-30 dBm at Frequency Products TOI the input mixer 100 Hz - 10 MHz <-66 dBc +3dBm (10 dB attn) 10 MHz - 22 GHz <-80 dBc +10 dBm Image Responses for RF input Center Image Responses (dBc) levels <= 0 dBm Freq (GHz) 6 MHz 42.8 MHz 642.8 MHz 0 - 2.9 <-85 <-85 NA 2.7 - 6.2 <-85 <-85 <-83 6.0 - 12.8 <-85 <-85 <-73 12.6 - 16.0 <-85 <-85 <-70 16.0 - 22.0 <-85 <-85 <-58 Residual Responses Range Responses (0 dB attenuation 10 MHz to 22 GHz <-100 dBm input terminated) Multiple Responses <-60 dBc for inputs <= 22 GHz and RF levels <= 0 dBm (>=10 dB attn)
If you have Adobe's Acrobat Reader already available on your system, please click on the icon provided below to print or view the specifications.
If you do not have the Adobe Acrobat Reader necessary for viewing PDF documentation, download your free Acrobat Reader, and then click on the icon below.
HP 71210C Microwave Spectrum Analyzer, 100 Hz to 22 GHz RF Related Options Opt 006 Delete HP 70902A 10 Hz to 300 kHz BW IF section Opt 007 Delete HP 70903A 100 kHz to 3 MHz BW IF section Opt 110 Delete HP 70310A precision frequency reference Opt 121 Add distribution amplifier to HP 70310A Rack Mounting Options Opt 810 Slide kit for both HP 70001A mainframe and HP 70004A display/mainframe Opt 908 Flange kit to mount without handles Opt 913 Flange kit to mount with handlesAll orders include the following:HP 70620B preamplifier for better sensitivity (1 GHz to 26.5 GHz) HP 70620B option 001 preamp (100 kHz to 26.5 GHz) HP 70300A RF tracking generator (20 Hz to 2.9 GHz) HP 70301A microwave tracking generator (2.7 GHz to 18 GHz)
HP 70001A, HP 70004A, HP 70310A, HP 70900B, HP 70902A, HP 70903A, HP 70908A, and user documentation
HP 70875A Noise Figure Measurement Personality, Product Overview
( PDF)
If you do not have the Adobe Acrobat Reader necessary for viewing PDF documentation, download your free Acrobat Reader now. |
![]() |
Additional help is a click away!
Want to talk to a knowledgeable engineer about this product at our HP Test and Measurement Call Center? Do you have a question you want to send in by e-mail? A measurement set-up you need to fax in? Follow the link above and you will be given a choice to contact an engineer by telephone, e-mail or fax.
This service is available only in the United States. For assistance outside the U.S., contact your local HP DIRECT or HP Sales Office.
Related MMS Spectrum Analyzer Family product datasheets:
Related product families available online:
Page Updated: Thursday February 18 07:51:17 UTC 1999