HP 8112A 50 MHz Pulse Generator |
The HP 8112A is a fully programmable 50 MHz pulse generator with 5 ns transitions and 32 V p-p (into open circuit) max output amplitude. All pulse parameters are variable including delay and double pulse spacing.
Besides the comprehensive trigger modes, external modulation capabilities extend applicability; upper level, width, period and delay-modulated signals are available. These can be combined with the trigger modes so that complex real-life signals like modulated bursts are simulated easily.
For multi-channel applications, higher frequencies or multi-level
signals, refer to the HP 8110A pulse generator.
Specifications apply with 50 Ohm load and temperatures in the range 0° to 55° C. Timing (Specifications apply for min transition times) Period: 20.0 ns to 950 ms Delay: 75.0 ns to 950 ms Double Pulse: 20.0 ns to 950 ms Width: 10.0 ns to 950 ms Accuracy:+/-5% of progr value +/-2 ns (Delay: +/-5 ns) Output Characteristics (Voltages double when driving into open circuit) Levels High level: -7.90 V to 8.00 V Low level: -8.00 V to 7.90 V Accuracy: +/-1% of progr value +/-3% amplitude +/-40 mV Output resistence: 50 Ohm Transition Times Fixed: 5 ns typical Linear and cosine: 6.5 ns to 95.0 ms (max edge ratio 1:20 within a 1.5-decade range. Ranges overlap by 0.5 decade) Accuracy: +/-5% of programmed value +/-2 ns Preshoot, Overshoot, Ringing: +/-5% +/-10 mV (variable transitions), +/-10% +/-10 mV (fixed transitions) Operating Modes: Normal, Trigger, Gate, Ext Width (pulse restoration), Ext Burst (1 to 1999 pulse) Control (Modulation) Modes Period, Delay, Width: Covered in 8 non-overlapping decades (max input frequency 8 kHz) High Level: -8 V to 8 V, independent of progr low level (min input transition 200 us) General HP-IB Environmental Storage temperature : -40° to +65°C Operating temperature: 0° to 55°C Humidity: 95% RH, 0° to 40°C Power: 100/120/220/140 V rms; +5%; -10%; 48 to 440 Hz; 120 VA max Weight: Net 5.9 kg (13 lb); shipping, 8.0 kg (18 lb) Size: 212.3 mm W x 89 mm H x 450mm D (8.36 in x 3.5 in x 17.7 in)
HP 8112A 50 MHz Pulse Generator Opt 910 Additional Operating and Service Manual Opt 1BP Mil Std 45662A Calibration with Test Data UK6 Commercial calibration with test data W30 3-year return repair service W32 3-year calibration return service W34 3-year standards-compliant calibration service W50 5-year return repair service W52 5-year calibration return service W54 5-year standards-compliant calibration service HP 5062-4001 Bail Handle Kit HP 5062-3972 Rack Mount Kit (single 8112A) HP 5062-3974 Rack Mount Kit (two 8112A) HP 5061-9694 Lock Link Kit (for use with HP 5062-3974) (*) For off-the-shelf shipment, call 800-452-4844.
Page Updated: Thursday June 19 16:17:55 UTC 1997