HP 83206A TDMA Cellular Adapter

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]
The HP 83206A TDMA Cellular Adapter adds TDMA DAMPS/DCCH (IS-136) mobile measurement capability to the HP 8920B RF Communications Test Set. The HP 83206A TDMA Cellular Adapter (HP8920B Option 800) provides a complete Pi/4 DQPSK signal generator, Pi/4 DQPSK modulation analyzer, data source, and BER analyzer for TDMA measurements on DAMPS (IS-54) and DCCH (IS-136) phones. The HP 83206A replaces the HP 83201B TDMA cellular adapter (HP 8920B Option 500) for making measurements on TDMA dual-mode DAMPS phones, adding digital control channel test features to fully characterize DCCH phones.


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

The following specifications apply when the HP 83206A is operating with an HP 8920B.

TDMA Signal Generator

Frequency Range 824 MHz to 894 MHz Output Level Range: RF In/Out: -22 dBm to -127 dBm Duplex Out: +4 dBm to -127 dBm Residual Error Vector Magnitude: <3.0% Residual Phase Error: <2.6 degrees Residual Magnitude Error: <2.6% IQ Origin Offset: <-30 dBc within 15 degrees C of last calibration Frequency Error: +/-4 Hz plus reference error

TDMA Analyzer

Frequency Range: 824 MHz to 894 MHz Input Level Range: RF In/Out: 1 mW to 60 W (0 to +47.8 dBm) Antenna In: -36 dBm to +16 dBm Input Frequency Setting Error: 1 kHz Frequency Error Accuracy: +/- 2.5 Hz plus reference accuracy Residual Error Vector Magnitude: <1.3% Error Vector Magnitude Measurement Accuracy: +/- 0.4% + 2% of reading Residual Phase Error: <1.0 degree Residual Magnitude Error: <0.9% I/Q Origin Offset Accuracy: +/-0.5 dB for values to -40 dBc


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

HP 83206A           TDMA Cellular Adapter
   Opt W30 	    Three-Year Return Repair Service
   Opt W32 	    Three-Year Customer Return Calibration Service
   Opt W50 	    Five-Year Return Repair Service
   Opt W52 	    Five-Year Customer Return Calibration Service


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

No literature is available online at this time.

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

To help you make the right instrument choices

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

Related product families available online:



Privacy Statement    Use of this site indicates that you accept the Terms of Use
Contact HP Test & Measurement. (c) Copyright 1994 - 1998 Hewlett-Packard Company.

Page Updated: Monday October 05 20:03:25 UTC 1998