HP 83751A Synthesized Sweeper, 2 GHz to 20 GHz

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]
The HP 83751A sweeper brings outstanding synthesized performance to the component-test marketplace. It delivers the best performance for the price in general purpose benchtop, swept test or scalar applications. This sweeper provides superior accuracy and stability while maintaining the speed of an analog source. Fully synthesized CW, stepped and ramp sweep modes are available in broadband and narrowband operation. The synthesis capability is particularly useful for the characterization of narrowband devices, in which the frequency instabilities of open-loop sources become most apparent. Excellent output power flatness and accuracy can be translated to the input port of the device under test with the power flatness correction feature of this source.


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

Frequency Specifications

Frequency Range: 2 to 20 GHz Timebase Stability Standard 10 MHz Time Base: ±10 ppm High Stability Time Base (Option 1E5): Accuracy = Calibration ± Aging Rate ±Temperature Effects ± Line Voltage Effects Aging Rate: <5 x 10-10/day, 1 x 10-7/year

CW Mode

Resolution: 1 Hz Accuracy Same as timebase Switching Time: 70 ms, maximum

Stepped Sweep Mode

Resolution: Settable 1 Hz, Display 1 kHz Accuracy: Same as timebase Minimum Step Size: Same as resolution Number of Points: 2 to 1601 Switching Time/Point1: 7 ms + 8 ms/GHz step Dwell Time/Point: 1 ms to 1 s 1Up to 50 ms switching times can occur when crossing the 2 GHz band switch point.

Ramp (Analog, Phase Locked) Sweep Mode

Resolution: 1 kHz Accuracy (25 ±5 °C): for 100 ms Sweeps the Greater of: ±0.01% of Span ± timebase or ±75 kHz ± timebase Sweep Time: 10 ms to 100s; 400 MHz/ms max

RF Output2

Maximum Leveled Power3 (25 ±5 °C)+10 dBm Minimum Settable Power: -15 dBm With Option 1E1: -85 dBm Resolution Settable: 0.01 dB Display: 0.1 dB Accuracy and Flatness4 ------------------------------------------------ Power Level Accuracy (25 ±5 °C) Flatness5 ------------------------------------------------ >-10 dBm ±1.0 dB ±0.7 dB >-80 dBm ±1.5 dB ±1.2 dB Power Sweep: ±25 dB/sweep Power Slope: 0 to ±2 dB/GHz, up to power sweep limit External Leveling External Detector: Range: -0.2 mV to -0.5 V Source Match: <1.7:1 SWR, (internally leveled) 2With type N connector (option 1ED), performance is typical above 18 GHz 3Power typically degrades <2.5 dB over 0 to 55 °C 4Specifications apply for coupled attenuator mode and ALC level >-10 dBm 5Below 50 MHz, flatness is specified over 25 ±5 °C range

Spectral Purity

Harmonics (At Max Leveled Power): 10 MHz to 1.5 GHz: -30 dBc 1.5 to 20 GHz: -45 dBc Subharmonics None Residual FM (0.05 to 15 kHz Bandwidth)6 1 kHz RMS in CW mode 6Residual FM is typically <10 kHz in unlocked FM mode

Modulation

Pulse On/Off Ratio: 60 dB Rise/Fall Times: 0.5 to 2 GHz: 15 ns 2 to 20 GHz: 100 ns rise, 50 ns fall Minimum Leveled Width7: 2 µs Internal Pulse Generator: Width Range: 1 µs to 65 ms Period Range: 2 µs to 65 ms Resolution: 1 µs Internal Square Wave: 1.0 kHz and 27.8 kHz, scalar mode 7For frequencies <500 MHz in CW mode, or <2 GHz in swept mode

AM

Sensitivity: 1 dB/V Bandwidth (3 dB): >100 kHz, Usable to 1 MHz Depth: 20 dB; +10 dBm to -10 dBm Input Impedance: 3.5 kohms

FM

AC/Locked Mode: Rates: 50 kHz to 10 MHz Maximum Deviation: Same as unlocked mode up to 25 X rate DC/Unlocked Mode: Rates: DC to 10 MHz Maximum Deviation: DC to 100 Hz Rates: ±75 MHz 100 Hz to 1 MHz Rates: ±7 MHz 1 MHz to 2 MHz Rates: ±5 MHz 2 MHz to 10 MHz Rates: ±1 MHz Sensitivity: -6 or -20 MHz/V Input Impedance: 1 kohm

Declarations of Conformity


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

HP 83751A     Synthesized Microwave Sweeper, 2 GHz to 20 GHz
   Opt 0B2    Operation manual
   Opt 0B3    Service manual
   Opt 1BN    Mil-std 45662A calibration certification
   Opt 1BP    Mil-std 45662A calibration w/ test data
   Opt 1CM    Rackmount kit
   Opt 1CP    Rackmount and handle kit
   Opt 1CR    Rack slide kit
   Opt 1E1    Add step attenuator
   Opt 1E4    Add rear panel output
   Opt 1E5    Add high stability timebase
   Opt 1ED    Type-N output connector
   Opt 1EE    Add mm source interface
   Opt AB0    Taiwan - Chinese localization
   Opt AB1    Korea - Korean localization
   Opt AB2    China - Chinese localization
   Opt ABD    Germany - German localization
   Opt ABF    France - French localization
   Opt ABJ    Japan - Japanese localization
   Opt ABZ    Italy - Italian localization
   Opt AX2    Add bail handle and front protective cover
   Opt UK6    Commercial calibration certificatew/ test data
   Opt W30    3 Years Return Repair Service
   Opt W32    3 Years Customer Return Calibration Service
   Opt W34    3 Years Customer Return Standards Compliant Calibration Service
   Opt W50    5 Years Return Repair Service
   Opt W52    5 Years Customer Return Calibration Service
   Opt W54    5 Years Standards Compliant Calibration Service


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

No literature is available online at this time.

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

To help you make the right product choices

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]