HP 8481D Diode Power Sensor, 10 MHz to 18 GHz


[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

Product Summary

The HP 8481D diode power sensor is designed for use with the HP 435B, 436A, 437B, 438A, 70100A, and E1416A power meters and provides extraordinary accuracy and stability. The HP 8481D high-sensitivity power sensor acts as a true-RMS reading device over its entire -70 to -20 dBm dynamic range which gives you accurate readings even if your test signal is subjected to multitone environments, modulated carriers, or carriers with high harmonics. You also get extremely low SWR and it's traceable to the U.S. National Institute of Standards and Technology (NIST). A calibration port is included with the power sensor for calibration with the power meter. This calibration provides traceability to NIST and eliminates the uncertainties due to temperature changes and the variance in making measurements with different meter/sensor combinations.


[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

Product Features



[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

Product Specifications

Frequency Range:		10 MHz to 18 GHz
Maximum SWR			
	10 to 30 MHz:		1.40
	30 MHz to 4 GHz:	1.15
	4 to 10 GHz:		1.20
	10 to 15 GHz:		1.30
	15 to 18 GHz:		1.35
Power Linearity¹:		-30 to -20 dBm; ±1%
Maximum Power:			100 mW avg.; 100 mW peak
Connector Type:			Type-N (m)


¹Negigible deviation except for the power range noted



[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

Product Ordering Information

HP 8481D	Diode Power Sensor, 10 MHz to 18 GHz
   Opt 1BN      Mil std 45662A calibration certification^M
   Opt 910      Extra manual^M
   Opt W30      3 Year Return Repair Service^M
   Opt W32      3 Year Customer Return Calibration Service^M
   Opt W34      3 Year Standards Compliant Cal Service^M
   Opt W50      5 Year Return Repair Service^M
   Opt W52      5 Year Cust Return Calibration Service^M
   Opt W54      5 Year Standards Compliant Cal Service^M


ČIncludes HP 11708A 30 dB attenuator for calibrating 
 against a 9 dBm, 50 MHz power reference.  HP 11708A 
 is factory set to 30 dB ±0.05 dB at 50 MHz, 
 traceable to NIST.  SWR <1.05 at 50 MHz.