HP 8487A Power Sensor, 50 MHz to 50 GHz |
The HP 8487A thermocouple power sensor is designed for use with the HP
435B, 436A, 437B, 438A, 70100A, and E1416A power meters and provides
extraordinary accuracy and stability. The HP 8487A power sensor gives
you extremely low SWR even at mm-wave frequencies, and is traceable to
the U.S. National Institute of Standards and Technology (NIST). A 50
MHz calibration port is included with the power sensor for calibration
with the power meter. This calibration provides traceability to NIST at
millimeter-wave frequencies, and it eliminates the uncertainties due to
temperature changes and the variance in making measurements with
different meter/sensor combinations.
Frequency Range: 50 MHz to 50 GHz Maximum SWR 50 to 100 MHz: 1.15 100 MHz to 2 GHz: 1.10 2 to 12.4 GHz: 1.15 12.4 to 18 GHz: 1.20 18 to 26.5 GHz: 1.25 26.5 to 40 GHz: 1.30 40 to 50 GHz: 1.50 Power Linearity¹: +10 to +20 dBm; +2, -4% Maximum Power: 300 mW avg., 15W peak, 30W 盜 per pulse Connector Type: 2.4mm (m) ¹Negligible deviation except for the power range noted
HP 8487A Power Sensor, 50 MHz to 50 GHz Opt W30 3 Year Return Repair Service^M Opt W32 3 Yr Calibration Customer Return Service Opt W34 3 Year Standards Compliant Cal Service^M Opt W50 5 Year Return Repair Service^M Opt W52 5 Year Cust Return Calibration Service^M Opt W54 5 Year Standards Compliant Cal Service^M
Page Updated: Thursday June 19 16:13:26 UTC 1997