HP 8494B DC to 18 GHz Manual Step Attenuator


[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

Product Summary

The HP 8494B dc to 18 GHz manual step attenuator offers exceptional repeatability and reliability. Attenuation is provided in 1 dB steps from 0 to 11 dB. Precision gold-plated leaf springs ensure long life (over 5 million cycles) and very high attenuator repeatability (typically 0.01 dB). To improve measurement accuracy, NIST traceable data (SWR and attenuation) is available as Option 890. Custom step attenuator/switch combinations are possible. The HP 11716A/B/C interconnect kits provide quick connection of HP 8494/5/6/7 1 dB and 10 dB step attenuators in series for greater dynamic range with 1 dB steps.


[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

Product Features



[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

Product Specifications

Electrical Specifications

Frequency Range: dc to 18 GHz Incremental Attenuation: 0 to 11 dB 1 dB steps SWR Maximum (50 ohm nominal) dc to 8 GHz: 1.5 dc to 12.4 GHz: 1.6 dc to 18 GHz: 1.9 Insertion Loss (0 dB setting): 0.6 dB + 0.09 dB/Ghz Attenuation Accuracy (% in dB from attenuation setting) dc to 12.4 GHz ±0.3 dB: 1 to 2 dB ±0.4 dB: 3 to 4 dB ±0.5 dB: 5 to 6 dB ±0.6 dB: 7 to 10 dB ±0.7 dB: 11 dB dc to 18 GHz ±0.7 dB: 1 to 5 dB ±0.8 dB: 6 to 9 dB ±0.9 dB: 10 to 11 dB Power Rating, Minimum Life: 1W average, 100W peak, 10 µs max pulse width 5 million cycles per section

Option 890 Calibration Data Specifications

Frequency List (MHz): 100, 300, 500, 700, 900, 1000, 1250, 1500, 1750, 2000, 2500, 3000, 3500, 4000 and every 500 MHz to 16000 (plus 12400 MHz), every 250 MHz from 16000 to 18000



[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

Product Ordering Information

Each order must include basic model number, suffix letter, and connector option.

HP 8494B	Manual Step Attenuator
   Opt 890	Calibration Data

   Opt 001	N (f) connector
   Opt 002	SMA (f) connector
   Opt 003	APC-7 connector



[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

To help you make the right instrument choices



Contact HP Test & Measurement. (c) Copyright 1994, 1995, 1996, 1997 Hewlett-Packard Company.

Page Updated: Thursday June 19 16:17:29 UTC 1997