HP 85025A Detector


[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

Product Summary

The HP 85025A detector is designed specifically for operation with the HP 8757D scalar network analyzer and is not compatible with the HP 8756 or the 8755. The HP 85025A detects either a modulated (ac) or an unmodulated (dc) microwave signal.


[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

Product Features



[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

Product Specifications

Frequency Range:		10 MHz to 18 GHz
Connector Type:			Type-N (m), 7 mm²
Dynamic Range
	ac mode:		+16 to -55 dBm
	dc mode:		+16 to -50 dBm
Return Loss
	0.01 to 0.04 GHz:	10 dB
	0.04 to 4 GHz:		20 dB
	4 to 18 GHz:		17 dB
Frequency Response
	0.01 to 0.04 GHz:	±0.8 dB
	0.04 to 18.0 GHz:	±0.5 dB
	4 to 18 GHz:		±0.5 dB

			Dynamic Accuracy¹	Absolute Accuracy²		
Power @  16 dBm:	±0.8 dB			±0.8 dB
Power @   6 dBm:	±0.4 dB			±0.4 dB
Power @ -35 dBm:	±0.4 dB			±0.4 dB
Power @ -50 dBm:	±1.3 dB			±1.3 dB

¹Dynamic Accuracy refers to measurement accuracy as power 
 varies (in dB) from a 0 dBm reference.  25° ±5°C, 50 MHz
²DC mode, 25° ±5°C



[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

Product Ordering Information

HP 85025A	Detector
   Opt 001	7 mm Connector

Requires HP 8757A firmware revision 2.0 or higher.  To upgrade previous
revisions, order the HP 11614A firmware enhancement.



[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

To help you make the right instrument choices



Contact HP Test & Measurement. (c) Copyright 1994, 1995, 1996, 1997 Hewlett-Packard Company.

Page Updated: Thursday June 19 16:17:24 UTC 1997