HP 85025B Detector |
The HP 85025B detector is designed specifically for operation with the
HP 8757D scalar network analyzer and is not compatible with the HP 8756
or the 8755. The HP 85025B detects either a modulated (ac) or an
unmodulated (dc) microwave signal.
Frequency Range: 10 MHz to 26.5 GHz Connector Type: 3.5 mm (m) Dynamic Range ac mode: +16 to -55 dBm dc mode: +16 to -50 dBm Return Loss 0.01 to 0.04 GHz: 10 dB 0.04 to 4 GHz: 20 dB 4 to 18 GHz: 17 dB 18 to 26.5 GHz: 12 dB Frequency Response 0.01 to 0.04 GHz: ±0.8 dB 0.04 to 18.0 GHz: ±0.5 dB 4 to 18 GHz: ±0.5 dB 18 to 26.5 GHz: ±2.0 dB Dynamic Accuracy¹ Absolute Accuracy² Power @ 16 dBm: ±0.8 dB ±0.8 dB Power @ 6 dBm: ±0.4 dB ±0.4 dB Power @ -35 dBm: ±0.4 dB ±0.4 dB Power @ -50 dBm: ±1.3 dB ±1.3 dB ¹Dynamic Accuracy refers to measurement accuracy as power varies (in dB) from a 0 dBm reference. 25° ±5°C, 50 MHz ²DC mode, 25° ±5°C
HP 85025B Detector Requires HP 8757A firmware revision 2.0 or higher. To upgrade previous revisions, order the HP 11614A firmware enhancement.