HP 85037B Precision Detector |
The HP 85037B precision detector is designed specifically for operation
with the HP 8757D scalar network analyzer and may be used in either ac
or dc detection modes. This dual diode detector contains internal
frequency correction factors in an internal EE PROM (read automatically
by the HP 8757D) for improved measurement accuracy versus frequency.
When used in conjunction with the HP 8757D's internal power calibrator
(Opt 002), these detectors provide the maximum absolute power
measurement accuracy. The HP 85037B is not compatible with the HP
8757E, 8757A, 8756A, or 8755.
Frequency Range: 10 MHz to 26.5 GHz Connector Type: 3.5 mm (m) Dynamic Range ac mode: +20 to -55 dBm dc mode: +20 to -50 dBm Return Loss 0.01 to 0.04 GHz: 10 dB 0.04 to 18.0 GHz: 20 dB 18.0 to 26.5 GHz: 18 dB Frequency Response 0.01 to 0.04 GHz: ±0.35 dB 0.04 to 18.0 GHz: ±0.18 dB 18.0 to 26.5 GHz: ±0.22 dB Dynamic Accuracy¹ Absolute Accuracy² Power @ 20 dBm: ±0.25 dB ±0.25 dB Power @ 10 dBm: ±0.11 dB ±0.11 dB Power @ -30 dBm: ±0.11 dB ±0.11 dB Power @ -50 dBm: ±0.85 dB ±0.85 dB ¹Dynamic Accuracy refers to measurement accuracy as power varies (in dB) from a 0 dBm reference. 25° ±5°C, 50 MHz ²DC mode, 25° ±5°C