HP 85044A Transmission/Reflection Test Set, 50 Ohm, 300 kHz to 3 GHz |
The HP 85044A test set provides the capability to measure the
reflection and transmission characteristics of 50 ohm devices. The
device must be physically turned around to measure its reverse direction
characteristics. This test set is used with the HP 8753A/B/C or the HP
8753D Option 011 only.
Impedance: 50 ohms Frequency Range: 300 kHz to 3 GHz Directivity¹: 35 dB to 1.3 GHz 30 dB to 3.0 GHz Typical Tracking Transmission Magnitude, Phase¹²³ 0.3 MHz to 2.0 MHz: ±1.5 dB, ±10 degrees 2.0 MHz to Fmax: ±1.5 dB, ±10 degrees Reflection Magnitude, Phase¹²³ 0.3 MHz to 2.0 MHz: ±1.5 dB, ±25 degrees 2.0 MHz to Fmax: ±1.5 dB, ±10 degrees Effective Source Match²³ (test ports) 0.3 MHz to 2.0 MHz: 14 dB 2.0 MHz to 1.3 GHz: 20 dB 1.3 GHz to Fmax: 16 dB RF Connectors Test Port: Precision 7 mm All Others: 50 ohm Type-N (f) Physical Characteristics Size: 191 mm W x 62 mm H x 204 mm D (7.5 in x 2.44 in x 8.0 in) Weight: Net, 1.7 kg (3.8 lb) Shipping, 2.0 kg (4.4 lb) ¹Degrees, specified as deviation from linear phase ²Fmax is the upper frequency limit of the associated test set ³Can be improved through accuracy enhancement
HP 85044A 50 Ohm Transmission/Reflection Test Set
Page Updated: Thursday June 19 16:14:01 UTC 1997