HP 85047A S-Parameter Test Set, 50 Ohm, 300 kHz to 6 GHz |
The HP 85047A test set provides the capability to measure
transmission and reflection characteristics (including S-parameters) of
two-port devices in either direction with a single connection. The test
set is controlled from the analyzer and includes programmable step
attenuators. It includes a frequency doubler that can be switched in to
measure 3 MHz to 6 GHz in a single sweep or switched out to measure 300
kHz to 3 GHz in a single sweep. The HP 8753B/C controls the frequency
doubler. (The HP 8753D Option 006 and 011 with built-in 6 GHz source
does not use the frequency doubler.) Option 006 (6 GHz receiver) is
required to activate the HP 85047A.
Impedance: 50 ohms Frequency Ranges: 300 kHz to 3 GHz 3 MHz to 6 GHz 300 kHz to 6 GHz Directivity: 35 dB to 1.3 GHz 30 dB to 3.0 GHz 25 dB to 6.0 GHz Typical Tracking Transmission Magnitude, Phase 300 kHz to 3 GHz: ±1.5 dB, ±10 degrees 3 to 6 GHz: ±0.5, -2.5 dB, ±20 degrees Reflection Magnitude, Phase 300 kHz to 3 GHz: ±1.5 dB, ±10 degrees 3 to 6 GHz: ±1.5 dB, ±20 degrees Effective Source Match 300 kHz to 1.3 GHz: 20 dB 1.3 GHz to 3 GHz: 16 dB 3 GHz to 6 GHz: 14 dB RF Connectors Test Port: Precision 7 mm All Others: 50 ohm Type-N (f) Includes: Four 190 mm (7.5 in) cables with type-N (m) connectors for connection to the HP 8753, one HP 8753 test set interconnect cable. Physical Characteristics Size: 426 mm W x 90 mm H x 508 mm D (16.75 in x 3.5 in x 20 in) Weight: Net, 10 kg (22 lb) Shipping, 15 kg (33 lb)
HP 85047A 50 Ohm S-Parameter Test Set - 6 GHz Opt 009 Mechanical Test Port Switch Opt 913 Rack Mount Kit HP 5062-4069
All orders include the following:
Four 190 mm (7.5 in) cables with type-N (m) connectors for connection to the HP 8753 and one HP 8753 test set interconnect cable.
Page Updated: Thursday June 19 16:14:07 UTC 1997