HP 85050D 7 mm Economy Mechanical Calibration Kit

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]
The HP 85050D economy mechanical calibration kit contains precision standard devices to characterize the systematic errors of the HP 8720 series or 8510C network analyzer system in the 7 mm interface. This kit also contains adapters to change the sex of the test port and a torque wrench for proper connection. Included are standards definitions on disk for the 8510C. Connector gages may be ordered separately.


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

Frequency Range
     (f)minimum:              0.045 GHz

     (-f)maximum:             18 GHz

Connector Type:               7-mm	

Return Loss, Fixed Load:      >=38 dB, DC to 18 GHz

Residual Directivity (dB) at (f)maximum:
                              40

Residual Source Match (dB) at (f)maximum:
                              35


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

HP 85050D     7 mm Economy Mechanical Calibration Kit


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

No literature is available online at this time.

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

To help you make the right product choices

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

Related product families available online:



Privacy Statement    Use of this site indicates that you accept the Terms of Use
Contact HP Test & Measurement. (c) Copyright 1994 - 1998 Hewlett-Packard Company.

Page Updated: Thursday February 18 08:07:09 UTC 1999