HP 85060A 7 mm Electronic Calibration Module, 1 to 18 GHz


[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

Product Summary

The electronic calibration system consists of a control unit and a connector specific calibration module. The control unit provides communication and control for the calibration module. The HP 85060A (7 mm) is an electronic calibration module that must be controlled by the control unit (HP 85060C).

Electronic calibration (ECal) is a precision, single-connection, one- or two-port calibration technique that uses fully traceable and verifiable electronic standards. ECal provides repeatable, accurate measurements while bringing convenience and simplicity to your daily calibration routine. ECal replaces the traditional calibration technique which uses mechanical standards. With mechanical standards, you are required to make numerous connections to the test ports for a single calibration. These traditional calibrations require intensive operator interaction, which is prone to errors.

With ECal, a full two-port calibration can be accomplished with a single connection of the ECal module and minimal operator interaction. This results in more repeatable calibrations and less wear on the connectors--and on you. Calibrations for noninsertable devices are equally convenient and straight-forward.


[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

Product Features



[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

Product Specifications

Frequency Range
	Standard:			1 - 18 GHz
	Opt 001:			.045 to 2 GHz
Maximum Power:			        +20 dBm	
Connectors:		                7 mm

Measurement Port Characteristics¹²
HP 85060A
					Frequency Range (GHz)
Residual			1 to 2³ 	2 to 8		8 to 18
----------------------------------------------------------------------
Directivity (dB)
        Typical		        49.5		49.5		47.5
	Specification	        48		47.5		44.5
Source Match (dB)
	Typical		        47.5		46		43
	Specification	        46		44		39.5
Reflection Tracking (±dB)
	Typical		        0.026		0.04		0.05
	Specification	        0.032		0.046		0.073

HP 85060A Opt 001
				Frequency Range (GHz)
Residual			0.045 to 2³
------------------------------------------------------
Directivity (dB)
	Typical		        54
	Specification	        51
Source Match (dB)
	Typical		        46.5
	Specification	        44.5
Reflection Tracking (±dB)
	Typical		        0.03
	Specification	        0.038


¹ Typical numbers exclude the effect of Shock and Humidity.  Typical performance 
  can be achieved if the module is not exposed to severe humidity or dropped 
  from greater than 16 inches after factory calibration.  Specifications include
  the effects of all environmental conditions:  Sine Vibration, Random Vibration,
  Storage Survival, Operating Temperature Stability, Shock, Humidity.
² Port A and B have different specifications.  Difference is typically 0 to 2 dB 
  for source match and directivity, and 0 to 0.006 dB for reflection tracking.  
  Data Shown is worst case (port B).
³ When a broadband calibration is performed (using a lowband module), Option 001 
  specifications are achieved.



[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

Product Ordering Information

HP 85060A     7 mm Electronic Calibration Module, 1 to 18 GHz
   Opt 001    Add Lowband Calibration Module, .045 to 2 GHz

HP 85060C     Electronic Calibration Control Unit (required to drive 
              HP 85060A, 85062A, or 85064A calibration modules)