HP 85093A RF Electronic Calibration (ECal) Module, 3.5 mm
| |||
![[ Summary ]](/tmo/datasheets/Graphics/tab-summary-high.gif)
![[ Features ]](/tmo/datasheets/Graphics/tab-features-reg.gif)
![[ Specifications ]](/tmo/datasheets/Graphics/tab-specs-reg.gif)
![[ Key Literature ]](/tmo/datasheets/Graphics/tab-keylit-reg.gif)
![[ Ordering ]](/tmo/datasheets/Graphics/tab-ordering-reg.gif)
![[ Assistance ]](/tmo/datasheets/Graphics/tab-assistance-reg.gif)
Electronic calibration (ECal) is a fast, easy and accurate method of calibrating vector network analyzers. The new RF ECal system is easily controlled via a PC running Windows® 95 or NT® 4.0.
ECal is an ideal solution for calibrating network analyzers. Performing a full 2-port calibration takes less than half the time and number of connections using ECal versus mechanical cal kits. Furthermore, the accuracy of the calibration is comparable between electronic & mechanical methods.
Traditional mechanical calibrations require intensive operator interaction which is prone to errors. With ECal, the operator simply connects the ECal module to the network analyzer and the software controls the rest.
An RF ECal module contains solid-state devices that switch in resistive, conductive & insulated elements to produce open, short, loads and thru states. Each of these states is calibrated at the Hewlett-Packard standards laboratory and the calibration data is stored in an EEPROM. A heater which is located in the RF ECal module regulates the temperature of the module to 28 deg C. The module has been designed to be both electrically & mechanically very stable and reliable.
Frequency Range: 30 kHz to 6 GHz
Maximum Power: +20 dBm
Minimum Power: -45 dBm
Connectors: 3.5 mm
---------------------------------------------------------------------------------------------------------------
RF ECal - HP 85093A
---------------------------------------------------------------------------------------------------------------
30 to 300 kHz¹ .3 to 300 MHz .3 to 3 GHz 3 to 6 GHz
---------------------------------------------------------------------------------------------------------------
Directivity (dB) -52 -50.5 -52 -50.5
Source Match (dB) -45 -45 -43.5 -40
Reflection
Tracking (dB) ±0.05 ±0.043 ±0.043 ±0.055
Transmission
Tracking (dB) ±0.14 ±0.050 ±0.045 ±0.090
Load Match (dB) -41 -48 -47 -44
¸RF ECal modules are specified to operate from 300 kHz, with typical
performance down to 30 kHz.
HP-IB (IEEE-488)
Use any of the following Interface cards:
HP 82341 C/D
National AT-GPIB/TNT
National AT-GPIB/TNT (Plug and Play)
National PCI-GPIB
Vector Network Analyzer Compatibility
The list below shows the HP Vector Network Analyzers that are compatible with ECal.
Network Analyzer Firmware Revision²
HP 8753C 4.13
HP 8753D 6.14
HP 8753E 7.12
HP 8719/20/22C 1.05
HP 8719/20/22D 6.14
HP 8501B 6.54
HP 8510C 7.00
² Firmware revision that ECal has been tested with. Other firmware revisions may need
to be upgraded.
HP 85093A Electronic Calibration Module - Connector Type 3.5 mm
Opt 1BN Mil-STD 45662 calibration certificate
Opt 1BP Mil-STD 45662 calibration certificate with measured data
Opt 910 Add an extra operating and service manual
Opt UK6 Commercial calibration certificate with measured data
Opt 00A Adds male-male and female-female connectors for non-
insertable device measurement
Opt 00F Replaces male-female connectors with female-female
Opt 00M Replaces male-female connnectors with male-male
ECal requires an HP 85097A PC Interface kit to operate.
Order HP 85097A seperately.
Trademarks:
Windows®95 and Windows NT® are U.S. registered trademarks of Microsoft Corporation
| If you do not have the Adobe Acrobat Reader necessary for viewing PDF documentation, download your free Acrobat Reader now. |
|
To help you make the right product choices
Related HP 85060 and HP 85090 Series Electronic Calibration (ECal) Kits product datasheets:
Related product families available online:

Page Updated: Thursday February 18 10:02:27 UTC 1999