HP 85108L Pulsed-RF Network Analyzer System, 45 MHz to 2 GHz

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]
The HP 85108L Pulsed-RF network analyzer is a factory-integrated system that provides the entire instrument configuration required to make pulsed-RF measurements from 45 MHz to 2.3 GHz. Simply add the desired measurement accessories for a complete system. The system is built around the HP 8510C with the pulsed-RF measurement option (Opt 008) and includes two HP 83620B Option H80 synthesizers. The system is optimized for component testing in applications such as cellular, direct broadcast satellite, and VHF/UHF. The HP 85108L has excellent pulsed-RF and CW performance with system dynamic range greater than70 dB when using the sideband detectors and better than 95 dB when using the standard narrowband detectors. Custom pulsed bias configurations for high (<100W) and narrow pulses (<1usec) can be integrated into any pulsed-RF system.


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

Dynamic Range1

For transmission measurements, independent of duty cycle. Frequency Range (GHz) 0.045 0.1 2 Maximum Power (dBm)2 measured at port 2 +20 +6 +5 Reference Power (dBm) at port 1 (nominal) +0 +1 0 Minimum Power (dBm) measured at port 2 Pulsed: -53 -66 -66 CW: -81 -95 -96 Receiver Dynamic Range (dB) Pulsed: 72 72 72 CW: 101 101 100 System Dynamic Range (dB) Pulsed: 53 67 66 CW: 81 95 95 1Limited by compression level and system noise floor. Noise floor is measured with full two-port error correction. 1024 averages. 2This maximum power measurement assumes that the HP 85110A test set has its internal step attenuators set to 0 dB. The test can handle up to 20 W (+43 dBm) or power if the step attenuators are activated and an isolator is installed - in the port 2 real panellink.

Measurement Port Characteristics

Frequency Range GHz 0.045 0.1 2 Residual (dB) Directivity: 40 40 40 Source Match: 39 39 39 Load Match: 40 40 40 Reflection Tracking: ±0.01 ±0.01 ±0.01 Transmission Tracking: ±0.082 ±0.0116 ±0.0184 Crosstalk3: 54 67 66 3Does not include noise.

Pulsed-RF Detectors

Equivalent Measurement Bandwidth (-3 dB): 1.5 MHz Aperature Uncertainty: <1 nanosecond, typical Transition Time (10% to 90%): 300 nanoseconds Trigger Level: TTL (falling edge), external Trigger Width: 100 nanoseconds, minimum Minimum Time Display: 5 microseconds Maximum Time Display: 40 milliseconds


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

HP 85108L     Pulsed-RF Network Analyzer System (45 MHz to 2 GHz)
   Opt 010    Add time domain capability to HP 8510C
   Opt 230    220/240V line voltage operation
   Opt W31    Add two years additional on-site service

Custom pulsed bias configurations for high power (>100W) and
narrow pulses (<1usec) can be integrated into any pulsed-RF system.

All orders include the following:

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

HP 8510C Family Network Analyzer, Technical Specifications ( PDF)
HP 85108 Series Network Analyzer Systems, Product Overview ( PDF)

If you do not have the Adobe Acrobat Reader necessary for viewing PDF documentation, download your free Acrobat Reader now. Get Acrobat

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]
Additional help is a click away!

Want to talk to a knowledgeable engineer about this product at our HP Test and Measurement Call Center? Do you have a question you want to send in by e-mail? A measurement set-up you need to fax in? Follow the link above and you will be given a choice to contact an engineer by telephone, e-mail or fax.

This service is available only in the United States. For assistance outside the U.S., contact your local HP DIRECT or HP Sales Office.

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

Related Information:


Related HP 8510C Network Analyzer Family, 45 MHz to 110 GHz product datasheets:

Related product families available online:



Privacy Statement    Use of this site indicates that you accept the Terms of Use
Contact HP Test & Measurement. (c) Copyright 1994 - 1998 Hewlett-Packard Company.

Page Updated: Thursday February 18 07:55:44 UTC 1999