The HP 85110L test set is specially configured for operation in pulsed-RF measurement systems. Four 90 dB step attenuators protect each input of the fundamentally-mixed down converter to allow measurement of test devices with output power of 50W CW. Special options are available to 50 GHz.


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

General Specifications

Frequency Range: 0.045 to 2 GHz Test Ports (Port 1 or 2) Nominal Operating Power Level: 0 dBm Connector Type: 7 mm Impedance, DC Bias: 50 ohm nominal, 500 mA, 40 Vdc maximum Attenuation Range (Incident Signal): 0 to 90 dB, in 10 dB steps RF Input Connector (Rear Panel) Maximum Input Power: +14 dBm Connector Type: 3.5 mm, female Dynamic Range1 (for transmission measurements, independent of duty cycle) Frequency Range GHz 0.045 to 0.1 0.1 to 0.5 0.5 to 2 Maximum Power (dBm) Measured at Port 22: +20 +6 +5 Reference Power at Port 1 (nominal): 0 +1 0 Minimum Power Measured at Port 2 Pulsed: -53 -66 -66 CW: -81 -95 -96 Receiver Dynamic Dange (dB) Pulsed: 72 72 72 CW: 101 101 100 System Dynamic Range Pulsed: 53 67 66 CW: 81 95 95 1Limited by compression level and system noise floor. Noise floor is measured with full two-port error correction, 1024 averages. 2This maximum power measurement assumes that the HP 85110L test set has its internal step attenuators set to 0 dB. The test can handle up to 50W (+47 dBm) or power if the step attenuators are activated and proper precautions are taken within the open architecture loops.

Measurement Port Characteristics

Frequency Range GHz 0.045 to 0.1 0.1 to 0.5 0.5 to 2 Residual (dB) Directivity: 40 40 40 Source Match: 39 39 39 Load Match: 40 40 40 Reflection Tracking: ±0.01 ±0.01 ±0.01 Transmission Tracking: ±0.0082 ±0.0116 ±0.0184 Crosstalk: 54 67 66

Pulsed-RF Detectors

Equivalent Measurement Bandwidth (-3 dB): 1.5 MHz Aperature Uncertainty: <1 nanosecond, typical Transition Time (10% to 90%): 300 nanoseconds Trigger Level (external): TTL (falling edge) Trigger Width (minimum: 100 nanoseconds Minimum Time Display: 5 microseconds Maximum Time Display: 40 milliseconds

Declarations of Conformity


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

HP 85110L     Pulsed-RF S-Parameter Test Set
   Opt 001    Add IF switching


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

HP 85110L, Technical Data Sheet ( PDF)

If you do not have the Adobe Acrobat Reader necessary for viewing PDF documentation, download your free Acrobat Reader now. Get Acrobat

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

To help you make the right product choices

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]


Related HP 8510C Network Analyzer Coaxial Test Sets Family product datasheets:

Related product families available online:



Privacy Statement    Use of this site indicates that you accept the Terms of Use
Contact HP Test & Measurement. (c) Copyright 1994 - 1998 Hewlett-Packard Company.

Page Updated: Thursday February 18 07:55:56 UTC 1999