HP 85902A Burst Carrier Trigger for Digital Wireless Communications![]()
|
The HP 85902A burst carrier trigger provides a TTL timing reference that allows your spectrum analyzer to reliably trigger directly off the RF signal when performing transmitter tests. With an input range of 60 dB and a separate built-in preamplifier for greater sensitivity, you have the ability to trigger on a wide range of transmitter power levels.
For detailed specifications refer to the HP 85902A Product Overview literature number 5091-7292E.
Electrical Characteristics
Trigger Frequency Range: 10 MHz to 2000 MHz Burst Burst RF Rate: 25 Hz to 500 Hz Minimum Burst Width: 100 µs Trigger Output Delay: <20 µs for RF input >–25 dBm Jitter: <2 µs (on a consistent burst signal) Level: TTL/CMOS logic level RF Input Amplitude Range: –30 dBm to +30 dBm Preamplifier Frequency Range: 10 MHz to 2000 MHz Gain Versus Frequency 10 to 350 MHz: >18 dB 350 to 1000 MHz: >16 dB 1000 to 2000 MHz: >12 dB Noise Figure: <10 dB Output Power at 1 dB Gain Compression: >+10 dBm
Front and Rear Panel Characteristics
Front Panel Inputs and Outputs Preamp In: 50 ohm, BNC female Preamp Out: 50 ohm, BNC female RF Input: 50 ohm, BNC female Rear Panel Outputs and Features Trigger Out: TTL/CMOS, BNC female Input Trigger Threshold Adjustment: Screwdriver adjustable potentiometer Trigger Indicator: LED, red
Environmental
Temperature Range Operating: 0 °C to +55 °C Storage: –40 °C to +75 °C Humidity: 95% relative humidity (MIL-T28800E class 3)
General Specifications1
+15 Vdc at 100 mA –12.6 Vdc at 40 mA Dimensions (nominal) Length: 150 cm (6 in.) Width: 80 cm (3 in.) Height: 30 cm (1 in.) Weight (nominal): 341 grams (12 oz) 1Power is supplied by the probe power connector on the front panel of the HP spectrum analyzer.
HP 85902A Burst Carrier Trigger for Digital Wireless Communications Opt 001 Adds mounting flanges Opt OBN Extra manual Recommended Spectrum Analyzers HP 8590A Series with option 105 HP 8590E Series with option 105 HP 8560E Series Mobile communication measurement personalities for HP 8590 series spectrum analyzers HP 85715B GSM HP 85717A CT2-CAI HP 85718B NADC-TDMA HP 85720C PDC HP 85722B DCS-1800 HP 85723A DECT HP 85725B CDMA HP 85726B PHSAll orders include the following:
The HP 85902A includes 122-mm long and 23-mm long BNC cables and a operating/service manual.
HP 85902A Burst Carrier Trigger for Digital Wireless Communications, Product Overview
If you do not have the Adobe Acrobat Reader necessary for viewing PDF documentation, download your free Acrobat Reader now. |
![]() |
Related Information:
Related HP 8560 E-Series High-Performance Spectrum Analyzers product datasheets:
Page Updated: Monday October 05 19:05:14 UTC 1998