HP 8590L Portable Spectrum Analyzer, 9 kHz to 1.8 GHz![]()
|
Frequency Specifications
Frequency Range 50 ohm: 9 kHz to 1.8 GHz 75 ohm (Opt001):1 MHz to 1.8 GHz Frequency Accuracy Frequency Reference Aging: ±2 x 10e-6/year Temperature Stabililty: ±5 x 10e-6 Initial Achievable Accuracy:±0.5 x 10e-6 Frequency Readout Accuracy (start, stop, center, marker): ±(frequency readout x freq ref error + span accuracy + 1% of span + 20% of RBW + 100 Hz) Marker Frequency Counter Accuracy Span <=10MHz: ±(marker freq x freq ref error + counter resolution + 100 Hz) Span >10 MHz: ±(marker freq x freq ref error + counter resolution + 1 kHz) Counter Resolutlon Span <=10 MHz: Selectable from 10 Hz to 100 kHz Span >10 MHz: Selectable from 100 Hz to 100 kHz Frequency Span Range: 0 Hz (zero span) and 10 kHz to 1.8 GHz Resolution: Four digits or 20 Hz, whichever is greater Accuracy Span <=10 MHz: ±2% ofspan Span >10 MHz: ±3% of span Sweep Time Range Span = 0 Hz or >10 kHz: 20 ms to 100 s Accuracy 20 ms to 100 s: ±3% Sweep Trigger: Free run, single, line, video, external Bandwidth Filters Resolution Bandwidths: 1 kHz to 3 MHz (3 dB) in 1, 3, 10 sequence; 9 kHz and 120 kHz (6 dB) EMI bandwidths. Accuracy: ±20% Selectivity (characteristic) -60 dB/-3 dB: 3 to 10 kHz 15:1 100 kHz to 3 MHz 15:1 1 kHz, 30 kHz 16:1 Video Bandwidth Range: 30 Hz to 1 MHz in 1, 3, 10 sequence Stability Noise Sidebands (1 kHz RBW, 30 Hz VBW, sample detector) >10 kHz Offset from CW Signal: <=-90 dBc/Hz >20 kHz Offset from CW Signal: <=-100 dBc/Hz >30 kHz Offset from CW Signal: <=-105 dBc/Hz Residual FM 1 kHz RBW, 1 kHz VBW: <=250 Hz p-p in 100 ms System Related Sidebands (>30 kHz offset from CW signal): <=-65 dBc
Amplitude Specifications
Amplitude Range Measurement Range: Displayed Average Noise Level to +30 dBm Opt 001: Displayed average noise level to +75 dBmV Maximum Safe Input (input attenuator >=10 dB) Average Continuous Power: +30 dBm (1 W) Opt 001: +75 dBmV (0.4 W) Peak Pulse Power: +30 dBm (1 W) Opt 001: +75 dBmV (0.4 W) DC: 25 Vdc Opt 001: 100 Vdc Gain Compression (>10 MHz): <=0.5 dB (total power at input mixer = -10 dBm) Displayed Average Noise Level (input terminated, 0 dB atten, 30 Hz VBW, sample detector) 1kHz RBW 400 kHz to 1.5 GHz <=-115 dBm 1.5 GHz to 1.8 GHz <=-113 dBm Opt 001 1 MHz to 1.5 GHz <=-63 dBmV 1.5 GHz to 1.8 GHz <=-61 dBmV Display Range Log Scale: 0 to -70 dB from ref level is calibrated; 0.1, 0.2, 0.5 dB/div and 1 to 20 dB/div in 1 dB steps; 8 div displayed Linear Scale: 8 divisions Scale Units: dBm, dBmV, dBµV, V, W Marker Readout Resolution Log Scale: 0.05 dB Linear Scale: 0.05% of ref level Reference Level Range: Same as amplitude range Resolution: ±0.01 dB for log scale; ±0.12% of ref level for linear scale Accuracy: ±0.3 dB at -20 dBm 0 to -59.9 dBm: ±(0.3 dB + 0.01 x dB from -20 dBm) Spurious Responses Second Harmonic Distortion 5 MHz to 1.8 GHz: <-70 dBc for -45 dBm tone at input mixer Third-order Intermodulation 5 MHz to 1.8 GHz: <-70 dBc for two -30 dBm tones at input and >50 kHz separation Other Input-Related Spurious (>=30 kHz offset, -20 dBm tone at input mixer): <-65 dBc Residual Responses (input terminated, 0 dB attenuation) 150 kHz to 1.8 GHz: <-90 dBm 1 MHz to 1.8 GHz (Opt 001): <-38 dBmV Frequency Response (10 dB input attenuation) Absolute (referenced to 300 MHz CAL OUT): ±1.5 dB Relative (referenced to midpoint between highest and lowest frequency response deviations): ±1.0 dB Calibrator Output Amplitude: -20 dBm ±0.4 dB Opt 001: +28.75 dBmV ±0.4 dB Resolution Bandwidth Switching Uncertainty (ref to 3 kHz RBW, at ref level) 3 kHz to 3 MHz RBW: ±0.4 dB 1 kHz RBW: ±0.5 dB Log to Linear Switching: ±0.25 dB at ref level Display Scale Fidelity Log Incremental Accuracy (0 to -60 dB from ref level): ±0.4 dB/4 dB Log Maximum Cumulative (0 to -70 dB from ref level): ±(0.4 + 0.01 x dB from ref level) Linear Accuracy: ±3% of ref level
General Specifications
Environmental MIL-T-28800: Has been type-tested to the environmental specifications of MIL-T-28800 Class 5 Temperature Operating: 0°C to +55°C Storage: -40°C to +75°C EMI Compatibility: Conducted and radiated interference CISPR Pub.11/1990 Group 1 Class A Audible Noise: <37.5 dBA pressure and <5.0 Bels power (ISODP7779) Power Requirements On (line 1): 90 to 132 V rms, 47 to 440 Hz 195 to 250 V rms, 47 to 66 Hz Power consumption <500 VA; <180 W Standby (line 0): Power consumption <7 W Memory User Program Memory (nominal): 238 KB nonvolatile RAM Data Storage (nominal) Internal: 50 traces, 8 states External Memory Cards: HP 85700A (32 KB), 24 traces or 32 states HP 85702A (128 KB), 99 traces or 128 states Video Cassette Recorder: Continuous video recording of display supported through composite video output Size (nominal, without handle, feet, or front cover): 325 mm W x 163 mm H x 427mm D Weight: 15.2 kg Environmental: Type tested to the environmental specifications of MIL-T-28800 Class 5
Option 010 and 011 Built-In Tracking Generators
Frequency Range Opt 010: 100 kHz to 1.8 GHz Opt 011: 1 MHz to 1.8 GHz Output Level Range Opt 010: -15 dBm to 0 dBm Opt 011: +27.8 to +42.8 dBmV Resolution: 0.1 dB Absolute Accuracy at 300 MHz, -10 dBm (+38.8 dBm V, Opt 011): ±1.5 dB Vernier Range: 15 dB Accuracy: ±1 dB Output Flatness: ±1.75 dB Spurious Output Harmonic Spurs: 0 dBm (+ 42.8 dBmV, Opt 011) output; <-25 dBc Nonharmonic Spurs: <-30 dBc Dynamic Range (characteristic, max output level - TG feedthrough) Opt 010: 106 dB Opt 011: 100 dB Power Sweep Range Opt 010: -15 dBm to 0 dBm Opt 011: +28.7 to +42.8 dBmV Resolution: 0.1 dB
If you have Adobe's Acrobat Reader already available on your system, please click on the icon provided below to print or view the specifications.
If you do not have the Adobe Acrobat Reader necessary for viewing PDF documentation, download your free Acrobat Reader, and then click on the icon below.
HP 8590L Portable Spectrum Analyzer, 9 kHz to 1.8 GHz Options General Opt 003 Memory Card Reader Opt 015 Protective Soft Tan Carrying/Operating Case Opt 016 Protective Soft Yellow Carrying/Operating Case Opt 040 Front Panel Protective Cover with Storage and CRT Sun Shield Opt 041 HP-IB and Parallel Printer Interfaces Opt 042 Protective Soft Carrying Case/Backpack Opt 043 RS-232 and Parallel Printer Interfaces Opt 908 Rack Mount Kit without Handles Opt 909 Rack Mount Kit with Handles Component Test Opt 010 Tracking Generator (100 kHz to 1.8 GHz) Cable TV Opt 001 75 ohm Input Opt 011 Tracking Generator (75 ohm, 1 MHz to 1.8 GHz) Opt 711 50/75 ohm Matching Pad/100 Vdc Block Warranty and Support Opt 0Q8 Factory Service Training Opt UK6 Commercial Calibration Certificate with Test Data Opt AB* Quick Reference Guide in Local Languages Opt W30 Two Additional Years Return-to-HP Service Opt W32 Two Additional Years Return-to-HP Calibration Opt 915 Component Level Information and Service Guide Most options can be retrofitted. Please contact your local HP sales representative. Application Measurement Cards/Personalities (Requires Opt 003 Memory/Measurement Card Reader) HP 85700A Blank 32-KB Memory Card HP 85702A Blank 128-KB Memory Card HP 85714A Scalar Measurement Personality HP 85721A Cable TV Measurement Personality HP 85921A Cable TV PC Software for HP 85721A For complete information, please contact your local HP sales representative.For Scalar Measurements HP 86044A 50 Ohm trans./reflection test set (Option H10) HP 85630A Scalar 50 Ohm transmission/reflection test set HP 86205A 50 Ohm RF bridge (type-N(f), 300 kHz to 6 GHz, 1.5 dB +0.1 dB/GHz insertion loss) HP 86207A 75 Ohm RF bridge (type-N(f), 300 kHz to 3 GHz, 1.5 dB +0.1 dB/GHz insertion loss) For EMC Measurements HP 11945A Close field probe set (9 kHz to 1 GHz) (with Option E51)17 HP 11947A Transient limiter (BNC(f) to type-N(m), 9 kHz to 200 MHz, 10 dB insertion loss) HP 11951A Option 12D infrared controller Printers HP C2634A HP DeskJet 320 portable color printer (parallel interface) HP C2162A HP DeskJet 540 monochrome/color printer (parallel interface) HP C2164A HP DeskJet 660C color printer (parallel interface) HP C2003A HP LaserJet 4L printer (parallel interface) HP C2005A HP LaserJet 4P printer (parallel & serial interfaces) Cables (Serial interface requires HP 82221P cable, Option 023 only) HP 10833A HP-IB cable (1 meter) HP 24542U RS-232 cable 3 meter (9 pin F to 9 pin F) Option 043 only or RS-232 9 pin PC connection to analyzer) HP 24542G RS-232 cable 3 meter (25 pin M to 9 pin F) Option 043 only or RS-232 25 pin PC or printer connection to analyzer) HP 24542M RS-232 cable 3 meter (25 pin M to 9 pin F) Option 043 only or RS-232 25 pin modem connection to analyzer) HP C2932A RS-232 cable 3 meter (9 pin M to 9 pin F) Option 043 only or serial 9 vin LaserJet 4P/4Plus connection to analyzer) HP 13242G RS-232 cable (5 meter) (for Option 023 only) HP C2950A Parallel printer cable (2 meter) Converters HP ITEL-45CHVU HP-IB/Parallel converter (North America, Japan, Korea, and Taiwan) HP ITEL-45CHVU HP-IB/Parallel converter (Continental Europe, U.K., Australia, and South Africa) (requires HP F1011A AC adapter) Amplifiers and Preamplifiers HP 8447D Preamplifier (100 kHz to 1.3 GHz, >25 dB gain) HP 8447E Power amplifier (100 kHz to 1.3 GHz, 22 dB gain) HP 8447F Preamplifier-power amplifier (HP 8447D & 8447E in a single package) HP 87405A Preamplifier (10 MHz to 3.0 GHz, 24 dB gain) HP 10855A Broadband preamplifier (2 MHz to 1300 MHz, 22 dB gain) HP 85905A 75 Ohm preamplifier (45 MHz to 1 GHz, 20 dB gain) Other Accessories Adaptors/Limiters HP 11852B Order with Option C04, 75/50 Ohm min. loss pad (type-N(f) to type-N(m), DC to 2 GHz, 5.7 dB insertion loss) HP 11867A RF Limiter (type-N(f) to type-N(m), DC to 1.8 GHz, <.75 dB insertion loss) 08590-60369 RS-232/parallel connector assembly to convert Opt 041 (HP-IB) to Opt 043 (RS-232) 08590-60368 HP-IB/parallel connector assembly to convert Opt 043 (RS-232) to Opt 041 (HP-IB) HP 11693A Limiter (type-N(f) to type-N(m), 0.1 GHz to 12.4 GHz, <2.0 dB insertion loss) HP 11694A 75/50 Ohm matching transformer (BNC(f) to BNC(m), 3 MHz-500 MHz, <.75 dB insertion loss) 08590-60090 75 Ohm matching pad (same as Option 711) (BNC(f) to BNC(m), 1 MHz to 1.8 GHz, 5.7 dB insertion loss) HP11742A High frequency blocking capacitor (3.5mm(f to m)), 10 MHz to 26.5 GHz, .35 dB IL <12.4 GHz, .7 dB IL <26.5 GHz) Probes HP 41800A Active probe (5 Hz to 500 MHz) HP 85024A High frequency probe (300 kHz to 3 GHz) Rackmounts 5062-4841 Rackmount without handles23 (centers instrument in rack) 5062-4840 Rackmount with handles23 (centers instrument in rack) 5062-8240 Rackmount with handles23 (mounts instrument left of center) 5062-8241 Rackmount without handles23 (mounts instrument left of center) 1494-0060 Rack slides for rack mount Miscellaneous HP 85901A Portable AC power source HP 85902A Burst carrier trigger 1420-0383 Memory card battery (CR 2016) 9211-5604 Transit case 5180-9055 CRT viewing hood (sun shield) HP C1405B Keyboard (requires HP C1405-60015 adapter) HP 346B Noise source (BNC(f)to APC 3.5(m), 10 MHz to 18 GHz)
HP 8590 L-Series Portable Spectrum Analyzers, Product Overview
If you do not have the Adobe Acrobat Reader necessary for viewing PDF documentation, download your free Acrobat Reader now. |
![]() |
Related Information:
Printer Compatibility Guide for Network and Spectrum Analyzers
Related HP 8590 E-Series Portable Spectrum Analyzer Family product datasheets:
Related product families available online:
Page Updated: Monday October 05 19:20:24 UTC 1998