HP 8592L Portable Spectrum Analyzer, 9 kHz to 22/26.5 GHz

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]
The HP 8592L is a low-cost, but full-featured, frequency accurate microwave spectrum analyzer designed to meet your general purpose measurement needs. The easy-to-use interface provides access to more than 200 built-in functions.


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

Frequency Specifications

Frequency Range Band LO Harmonic = N Range 0 1 9 kHz to 2.9 GHz 1 1 2.75 to 6.5 GHz 2 2 6.0 to 12.8 GHz 3 3 12.4 to 19.4 GHz 4 4 19.1 to 22 GHz 4 4 (Opt 026, 027) 19.1 to 26.5 GHz Frequency Accuracy Frequency Reference Aging: ±2 x 10e-6/year Temperature Stabililty: ±5 x 10e-6 Initial Achievable Accuracy:±0.5 x 10e-6 Frequency Readout Accuracy (start, stop, center, marker): ±(frequency readout x freq ref error + span accuracy + 1% of span + 20% of RBW + 100 Hz x N) Marker Frequency Counter Accuracy Span <=10MHz x N: ±(marker freq x freq ref error + counter resolution + 100 Hz x N) Span >10 MHz x N: ±(marker freq x freq ref error + counter resolution + 1 kHz x N) Counter Resolution Span <=10 MHz x N: Selectable from 10 Hz to 100 kHz Span >10 MHz x N: Selectable from 100 Hz to 100 kHz Frequency Span Range: 0 Hz (zero span) and [10 kHz x N] to 19.25 GHz (to 23 GHz Opt 026,027) Resolution: Four digits or 20 Hz x N, whichever is greater Accuracy Span <=10 MHz x N: ±2% of span Span >10 MHz x N: ±3% of span Sweep Time Range Span = 0 Hz or >10 kHz: 20 ms to 100 s Accuracy: ±3% Sweep Trigger: Free run, single, line, video, external Bandwidth Filters Resolution Bandwidths: 1 kHz to 3 MHz (3 dB) in 1, 3, 10 sequence; 9 kHz and 120 kHz (6 dB) EMI bandwidths. Accuracy: ±20% Selectivity (characteristic) -60 dB/-3 dB: 3 to 10 kHz 15:1 100 kHz to 3 MHz 15:1 1 kHz, 30 kHz 16:1 Video Bandwidth Range: 30 Hz to 1 MHz in 1, 3, 10 sequence Stability Noise Sidebands (1 kHz RBW, 30 Hz VBW, sample detector) >10 kHz offset from CW signal: <=-90 dBc/Hz + 20 log N >20 kHz offset from CW signal: <=-100 dBc/Hz + 20 log N >30 kHz oftset from CW signal: <=-105 dBc/Hz + 20 log N Residual FM 1 kHz RBW, 1 kHz VBW: <=(250 x N) Hz p-p in 100 ms System Related Sidebands (>30 kHz offset from CW signal): <=-65 dBc + 20 log N Comb Generator Fundamental: 100 MHz Frequency Accuracy: ±0.007%

Amplitude Specifications

Amplitude Range Measurement Range: Displayed average noise level to +30 dBm Maximum Safe Input (input attenuator >=10 dB) Average Continuous Power: +30 dBm (1 W) Peak Pulse Power: +50 dBm (100 W) for < 10 µs pulse width and <1% duty cycle, input atten >=30 dB DC: 0V Gain Compression (>10 MHz): <=0.5 dB (total power at input mixer = -10 dBm) Displayed Average Noise Level (input terminated, 0 dB atten, 30 Hz VBW or 1 Hz VBW with Opt 130, sample detector) 1kHz RBW 400 kHz to 2.9 GHz <=-112 dBm 2.75 GHz to 6.5 GHz <=-114 dBm 6.0 GHz to 12.8 GHz <=-102 dBm 12.4 GHz to 19.4 GHz <=-98 dBm 19.1 GHz to 22 GHz <=-92 dBm Option 026/027 19.1 to 26.5 GHz <=-87 dBm Display Range Log scale: 0 to -70 dB from ref level is calibrated; 0.1, 0.2, 0.5 dB/div and 1 to 20 dB/div in 1 dB steps; 8 div displayed Linear Scale: 8 divisions Scale Units: dBm, dBmV, dBµV, V, W Marker Readout Resolution Log Scale: 0.05 dB Linear Scale: 0.05% of ref level Reference Level Range: Same as amplitude range Resolution: ±0.01 dB for log scale; ±0.12% of ref level for linear scale Accuracy: ±0.3 dB at -20 dBm 0 to -59.9 dBm: ±(0.3 dB + 0.01 x dB from -20 dBm) Spurious Responses Second Harmonic Distortion >10 MHz to 2.9 GHz: <-70 dBc for -40 dBm tone at input mixer >2.75 GHz: <-100 dBc for -10 dBm tone at mixer (or below DANL) Third-order Intermodulation >10 MHz: <-70 dBc for two -30 dBm tones at input and >50 kHz separation Other Input-Related Spurious (>=30 kHz offset, -20 dBm tone at input mixer) Applied Freq <=18 GHz: <-65 dBc Applied Freq <=22 GHz: <-60 dBc Residual Responses (input terminated, 0 dB attenuation) 150 kHz to 6.5 GHz: <-90 dBm Frequency Response (10 dB input attenuation) Absolute (referenced to 300 MHz CAL OUT): ±1.5 dB to ±5.0 dB (preselector peaked) Relative (referenced to midpoint between highest and lowest frequency response deviations): ±1.0 dB to ±2.0 dB (preselector peaked) Calibrator Output Amplitude: -20 dBm ±0.4 dB Resolution Bandwidth Switching Uncertainty (ref to 3 kHz RBW, at ref level) 3 kHz to 3 MHz RBW: ±0.4 dB 1 kHz RBW: ±0.5 dB Log to Linear Switching: ±0.25 dB at ref level Display Scale Fidelity Log Incremental Accuracy (0 to -60 dB from ref level): ±0.4 dB/4 dB Log Maximum Cumulative (0 to -70 dB from ref level): ±(0.4 + 0.01 x dB from ref level) Linear Accuracy: ±3% of ref level

General Specifications

Environmental MIL-T-28800: Has been type-tested to the environmental specifications of MIL-T-28800 Class 5 Temperature Operating: 0°C to +55°C Storage: -40°C to +75°C EMI Compatibility: Conducted and radiated interference CISPR Pub.11/1990 Group 1 Class A Audible Noise: <37.5 dBA pressure and <5.0 Bels power (ISODP7779) Power Requirements On (line 1): 90 to 132 V rms, 47 to 440 Hz 195 to 250 V rms, 47 to 66 Hz Power consumption <500 VA; <180 W Standby (line 0): Power consumption <7 W Memory User Program Memory (nominal): 238 KB nonvolatile RAM Data Storage (nominal) Internal: 50 traces, 8 states External Memory Cards: HP 85700A (32 KB), 24 traces or 32 states HP 85702A (128 KB), 99 traces or 128 states Video Cassette Recorder: Continuous video recording of display supported through composite video output Size (nominal, without handle, feet, or cover): 325 mm W x 163 mm H x 427mm D Weight: 15.9 kg Environmental: Type tested to the environmental specifications of MIL-T-28800 Class 5

Declarations of Conformity


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

HP 8592L       Portable Spectrum Analyzer, 9 kHz to 22/26.5 GHz

Options
General
   Opt 003     Memory/Measurement Card Reader
   Opt 015     Protective Soft Tan Carrying/Operating Case
   Opt 016     Protective Soft Yellow Carrying/Operating Case
   Opt 026     26.5 GHz Frequency Extension, APC 3.5 mm Connector
   Opt 027     26.5 GHz Frequency Extension, Type-N Connector
   Opt 040     Front Panel Protective Cover with Storage and CRT Sun Shield
   Opt 041     HP-IB and Parallel Printer Interfaces
   Opt 042     Protective Soft Carrying Case/Backpack
   Opt 043     RS-232 and Parallel Printer Interfaces
   Opt 908     Rackmount Kit without Handles
   Opt 909     Rackmount Kit with Handles

Cable TV
   Opt 711     50/75 ohm Matching Pad/100 Vdc Block

Warranty and Support
   Opt 0Q8     Factory Service Training
   Opt UK6     Commercial Calibration Certificate with Test Data
   Opt AB*     Quick Reference Guide in Local Languages
   Opt W30     Two Additional Years Return-to-HP Service
   Opt W32     Two Additional Years Return-to-HP Calibration
   Opt 915     Component Level Information and Service Guide

Most options can be retrofitted. Please contact your local
HP sales representative.

Application Measurement Cards/Personalities
(Requires Opt 003 Memory/Measurement Card Reader)

HP 85700A     Blank 32-KB Memory Card
HP 85702A     Blank 128-KB Memory Card
HP 85713A     Digital Radio Measurement Personality
HP 85721A     Cable TV Measurement Personality
HP 85921A     Cable TV PC Software for HP 85721A

For complete information, please contact your local HP sales
representative.

For Scalar Measurements HP 86044A 50 Ohm trans./reflection test set (Option H10) HP 85630A Scalar 50 Ohm transmission/reflection test set HP 86205A 50 Ohm RF bridge (type-N(f), 300 kHz to 6 GHz, 1.5 dB +0.1 dB/GHz insertion loss) HP 86207A 75 Ohm RF bridge (type-N(f), 300 kHz to 3 GHz, 1.5 dB +0.1 dB/GHz insertion loss) For EMC Measurements HP 11945A Close field probe set (9 kHz to 1 GHz) (with Option E51)17 HP 11947A Transient limiter (BNC(f) to type-N(m), 9 kHz to 200 MHz, 10 dB insertion loss) HP 11951A Option 12D infrared controller Printers HP C2634A HP DeskJet 320 portable color printer (parallel interface) HP C2162A HP DeskJet 540 monochrome/color printer (parallel interface) HP C2164A HP DeskJet 660C color printer (parallel interface) HP C2003A HP LaserJet 4L printer (parallel interface) HP C2005A HP LaserJet 4P printer (parallel & serial interfaces) Cables (Serialinterface requires HP 82221P cable, Option 023 only) HP 10833A HP-IB cable (1 meter) HP 24542U RS-232 cable 3 meter (9 pin F to 9 pin F) Option 043 only or RS-232 9 pin PC connection to analyzer) HP 24542G RS-232 cable 3 meter (25 pin M to 9 pin F) Option 043 only or RS-232 25 pin PC or printer connection to analyzer) HP 24542M RS-232 cable 3 meter (25 pin M to 9 pin F) Option 043 only or RS-232 25 pin modem connection to analyzer) HP C2932A RS-232 cable 3 meter (9 pin M to 9 pin F) Option 043 only or serial 9 vin LaserJet 4P/4Plus connection to analyzer) HP 13242G RS-232 cable (5 meter) (for Option 023 only) HP C2950A Parallel printer cable (2 meter) Converters HP ITEL-45CHVU HP-IB/Parallel converter (North America, Japan, Korea, and Taiwan) HP ITEL-45CHVU HP-IB/Parallel converter (Continental Europe, U.K., Australia, and South Africa) (requires HP F1011A AC adapter) Amplifiers and Preamplifiers HP 8447D Preamplifier (100 kHz to 1.3 GHz, >25 dB gain) HP 8447E Power amplifier (100 kHz to 1.3 GHz, 22 dB gain) HP 8447F Preamplifier-power amplifier (HP 8447D & 8447E in a single package) HP 87405A Preamplifier (10 MHz to 3.0 GHz, 24 dB gain) HP 10855A Broadband preamplifier (2 MHz to 1300 MHz, 22 dB gain) HP 85905A 75 Ohm preamplifier (45 MHz to 1 GHz, 20 dB gain) Other Accessories Adaptors/Limiters HP 11852B Order with Option C04, 75/50 Ohm min. loss pad (type-N(f) to type-N(m), DC to 2 GHz, 5.7 dB insertion loss) HP 11867A RF Limiter (type-N(f) to type-N(m), DC to 1.8 GHz, <.75 dB insertion loss) 08590-60369 RS-232/parallel connector assembly to convert Opt 041 (HP-IB) to Opt 043 (RS-232) 08590-60368 HP-IB/parallel connector assembly to convert Opt 043 (RS-232) to Opt 041 (HP-IB) HP 11693A Limiter (type-N(f) to type-N(m), 0.1 GHz to 12.4 GHz, <2.0dB insertion loss) HP 11694A 75/50 Ohm matching transformer (BNC(f) to BNC(m), 3 MHz-500 MHz, <.75 dB insertion loss) 08590-60090 75 Ohm matching pad (same as Option 711) (BNC(f) to BNC(m), 1 MHz to 1.8 GHz, 5.7 dB insertion loss) HP11742A High frequency blocking capacitor (3.5mm(f to m)), 10 MHz to 26.5 GHz, .35 dB IL <12.4 GHz, .7 dB IL <26.5 GHz) Probes HP 41800A Active probe (5 Hz to 500 MHz) HP 85024A High frequency probe (300 kHz to 3 GHz) Rack Mounts 5062-4841 Rackmount without handles23 (centers instrument in rack) 5062-4840 Rackmount with handles23 (centers instrument in rack) 5062-8240 Rackmount with handles23 (mounts instrument left of center) 5062-8241 Rackmount without handles23 (mounts instrument left of center) 1494-0060 Rack slides for rack mount Miscellaneous HP 85901A Portable AC power source HP 85902A Burst carrier trigger 1420-0383 Memory card battery (CR2016) 9211-5604 Transit case 5180-9055 CRT viewing hood (sun shield) HP C1405B Keyboard (requires HP C1405-60015 adapter) HP 346B Noise source (BNC(f) to APC 3.5(m), 10 MHz to 18 GHz)


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

HP 8590 L-Series Portable Spectrum Analyzers, Product Overview

If you do not have the Adobe Acrobat Reader necessary for viewing PDF documentation, download your free Acrobat Reader now. Get Acrobat

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

To help you make the right instrument choices

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

Related Information:


Related HP 8590 E-Series Portable Spectrum Analyzer Family product datasheets:

Related product families available online:



Privacy Statement    Use of this site indicates that you accept the Terms of Use
Contact HP Test & Measurement. (c) Copyright 1994 - 1998 Hewlett-Packard Company.

Page Updated: Monday October 05 19:56:59 UTC 1998