HP 8596E Portable Spectrum Analyzer, 9 kHz to 12.8 GHz![]()
|
Frequency Specifications
Frequency Range Band LO Harmonic = N Range 0 1 9 kHz to 2.9 GHz (dc-coupled) 0 1 100 kHz to 2.9 GHz (ac-coupled) 1 1 2.75 to 6.5 GHz 2 2 6.0 to 12.8 GHz Frequency Reference Aging: ±2 x 10-6/year; ±l x 10-7/year (Opt 004) Temperature Stabililty: ±5 x 10-6, ±l x 10-8 (Opt 004) Initial Achievable Accuracy: ±0.5 x 10-6; ±2.2 x 10-8 (Opt 004) Frequency Readout Accuracy (start, stop, center, marker): ±(frequency readout x freq ref error + span accuracy + 1% of span + 20% of RBW + 100 Hz x N) Marker Frequency Counter Accuracy Span <=10 MHz x N: ±(marker freq x freq ref error + counter res + 100 Hz x N) Span >10 MHz x N: ±(marker freq x freq ref error + counter res + 1 kHz x N) Counter Resolution Span <=10 MHz x N: Selectable from 10 Hz to 100 kHz Span >10 MHz x N: Selectable from 100 Hz to 100 kHz Frequency Span Range: 0 Hz (zero span) and [10 x N] kHz to 12.8 GHz; [1 x N] kHz minimum (Opt 130) Resolution: Four digits or 20 Hz x N, whichever is greater Accuracy Span <=10 MHz x N: ±2% of span Span >10 MHz x N: ±3% of span Sweep Time Range Span = 0 Hz or >1 kHz: 20 ms to 100 s Span = 0 Hz (Opt 101): 20 µs to 100 s Accuracy 20 ms to 100 s: ±3% 20 µs to <20 ms (Opt 101): ±2% Sweep Trigger: Free run, single, line, video, external Bandwidth Filters Resolution Bandwidths: 1 kHz to 3 MHz (3 dB) in 1, 3, 10 sequence; 9 kHz and 120 kHz (6 dB) EMI bandwidths. Option 130 adds 30, 100, and 300 Hz (3 dB) bandwidths and 200 Hz (6 dB) EMI bandwidth. Accuracy: ±20% Selectivity (characteristic) -60 dB/-3 dB: 3 to 10 kHz 15:1 100 kHz to 3 MHz 15:1 1 kHz, 30 kHz 16:1 -40 dB/-3 dB: 30 Hz to 300 Hz 10:1 Video Bandwidth Range: 30 Hz to 1 MHz in 1, 3 sequence (1 Hz to 1 MHz with Opt 130) Stability Noise Sidebands (1 kHz RBW, 30 Hz VBW, sample detector) >10 kHz offset from CW signal: <=-90 dBc/Hz + 20 log N >20 kHz offset from CW signal: <=-100 dBc/Hz + 20 log N >30 kHz oftset from CW signal: <=-105 dBc/Hz + 20 log N Residual FM 1 kHz RBW, 1 kHz VBW: <=(250 x N) Hz p-p in 100 ms 30 Hz RBW, 30 Hz VBW: <=(30 x N) Hz p-p in 300 ms System Related Sidebands (>30 kHz offset from CW signal): <=-65 dBc + 20 log N Comb Generator Fundamental: 100 MHz Frequency Accuracy: ±0.007%
Amplitude Specifications
Amplitude Range Measurement Range: Displayed average noise level to +30 dBm Maximum Safe Input (input attenuator >=10 dB) Average Continuous Power: +30 dBm (1 W) Peak Pulse Power: +50 dBm (100 W) for < 10 µs pulse width and <1% duty cycle, input atten >=30 dB DC: 0 V (dc-coupled); 50 V (ac-coupled) Gain Compression (>10 MHz): <=0.5 dB (total power at input mixer = -10 dBm) Displayed Average Noise Level (input terminated, 0 dB atten, 30 Hz VBW or 1 Hz VBW with Opt 130, sample detector) 30 Hz RBW 1 kHz RBW 400 kHz to 2.9 GHz <=-125 dBm <=-110 dBm 2.75 GHz to 6.5 GHz <=-127 dBm <=-112 dBm 6.0 GHz to 12.8 GHz <=-115 dBm <=-100 dBm Display Range Log Scale: 0 to -70 dB from ref level is calibrated; 0.1, 0.2, 0.5 dB/div and 1 to 20 dB/div in 1 dB steps; 8 div displayed Linear Scale: 8 divisions Scale Units: dBm, dBmV, dBµV, V, W Marker Readout Resolution Log Scale: 0.05 dB Linear Scale: 0.05% of ref level Fast Time Sweep For Zero Span (Opt 101 or 301, 20 µs to 20 ms) <=1 GHz: 0.7% of ref level for linear scale >1 GHz: 1.0% of ref level for linear scale Reference Level Range: Same as amplitude range Resolution: ±0.01 dB for log scale; ±0.12% of ref level for linear scale Accuracy: ±0.3 dB at -20 dBm 0 to -59.9 dBm: ±(0.3 dB + 0.01 x dB from -20 dBm) Spurious Responses Second Harmonic Distortion >10 MHz: <-70 dBc for -40 dBm tone at input mixer >2.75 GHz: <-100 dBc for -10 dBm tone at mixer (or below DANL) Third-Order Intermodulation >10 MHz: <-70 dBc for two -30 dBm tones at input and >50 kHz separation Other Input-Related Spurious (>=30 kHz offset, -20 dBm tone at input mixer) <-65 dBc Residual Responses (input terminated, 0 dB attenuation) 150 kHz to 6.5 GHz: <-90 dBm Frequency Response (10 dB input attenuation) Absolute (referenced to 300 MHz CAL OUT): ±1.5 dB to ±2.5 dB (preselector peaked) Relative Flatness (referenced to midpoint between highest and lowest frequency response deviations): ±1.0 dB to ±2.0 dB Calibrator Output Amplitude: -20 dBm ±0.4 dB Resolution Bandwidth Switching Uncertainty (ref to 3 kHz RBW, at ref level) 3 kHz to 3 MHz RBW: ±0.4 dB 1 kHz RBW: ±0.5 dB 30 Hz to 300 Hz RBW: ±0.6 dB Log to Linear Switching: ±0.25 dB at ref level Display Scale Fidelity Log Incremental Accuracy (0 to -60 dB from ref level): ±0.4 dB/4 dB Log Maximum Cumulative (0 to -70 dB from ref level) 3 kHz to 3 MHz RBW ±(0.3 + 0.01 x dB from ref level) 30 Hz to 3 kHz RBW ±(0.4 + 0.01 x dB from ref level) Linear Accuracy: ±3% of ref level
General Specifications
Environmental MIL-T-28800: Has been type-tested to the environmental specifications of MIL-T-28800 Class 5 Temperature Operating: 0 °C to +55 °C Storage: -40 °C to +75 °C EMI Compatibility: Conducted and radiated interference CISPR Pub.11 and Messempfaenger Postverfuegung 526/527/79 Audible Noise: <37.5 dBA pressure and <5.0 Bels power (ISODP7779) Power Requirements On (line 1): 90 to 132 V rms, 47 to 440 Hz 195 to 250 V rms, 47 to 66 Hz Power consumption <500 VA; <180 W Standby (line 0): Power consumption <7 W Memory User Program Memory (nominal): 238 KB nonvolatile RAM Data Storage (nominal) Internal: 50 traces, 8 states External Memory Cards: HP 85700A (32 KB), 24 traces or 32 states HP 85702A (128 KB), 99 traces or 128 states Video Cassette Recorder: Continuous video recording of display supported through composite video output Size (nominal, without handle, feet, or cover): 325 mm W x 163 mm H x 427mm D Weight: 16.4 kg
Option 010 and 011 Built-In Tracking Generators
Frequency Range Opt 010: 9 kHz to 2.9 GHz Output Level Range Opt 010: -1 to 66 dBm Resolution: 0.1 dB Absolute Accuracy: ±0.5 dB Vernier Range: 9 dB Accuracy: ±0.8 dB Output Flatness: ±2 dB, >10 MHz Spurious Output Harmonic Spurs: -1 dBm output; <-25 dBc Nonharmonic Spurs: <-30 dBc Dynamic Range (characteristic, max output level - TG feedthrough) Opt 010: 111 dB, >400 kHz Power Sweep Range Opt 010: -66 dBm to -1 dBm in 8 dB increments Resolution: 0.1 dB
If you have Adobe's Acrobat Reader already available on your system, please click on the icon provided below to print or view the specifications.
If you do not have the Adobe Acrobat Reader necessary for viewing PDF documentation, download your free Acrobat Reader, and then click on the icon below.
HP 8596E Portable Spectrum Analyzer, 9 kHz to 12.8 GHz General Options Opt 004 Precision Frequency Reference Opt 009 Lo and Sweep + Tune Opt 015 Soft Tan Carrying/Operating Case Opt 016 Soft Yellow Carrying/Operating Case Opt 040 Front Panel Protective Cover With Storage Opt 041 HP-IB and Parallel Printer Interfaces Opt 042 Protective Soft Carrying Case/Backpack Opt 043 RS-232 and Parallel Printer Interfaces Opt 101 Fast Time Domain Sweeps and Analog + Display Opt 102 AM/FM Demodulator and TV Sync Trigger (TV Sync requires Opt 101) Opt 130 Narrow Resolution Bandwidths (30 to 300 Hz and 200 Hz EMI) Opt 140 Narrow Bandwidths and Precision Frequency Reference Opt 301 TV Sync Trigger, Fast Time Domain Sweeps, AM/FM Demodulator, Analog + Display Opt 908 Rackmount Kit without Handles Opt 909 Rackmount Kit with Handles Communications Bundling Opt BD0 CDMA Bundle Opt 004 Precision Frequency Reference Opt 053 Improved Amplitude Accuracy for CDMA Opt 105 Time-Gated Spectrum Analysis Opt 151 DSP, Fast ADC and Digital Demodulator Opt 160 PDC/PHS/NADC/CDMA FW for Opt 151 HW 85725C CDMA Test Personality for 8590 E-Series Opt BD1 GSM Bundle Opt 004 Precision Frequency Reference Opt 105 Time-Gated Spectrum Analysis Opt 151 DSP, Fast ADC and Digital Demodulator Opt 163 GSM/DCS Firmware for Phase/Frequency Error 85727A GSM Multi-Band Transmitter Measurement Personality Opt BD2 TDMA/NADC Bundle Opt 004 Precision Frequency Reference Opt 050 Improved Amplitude Accuracy for NADC/PCS Opt 105 Time-Gated Spectrum Analysis Opt 151 DSP, Fast ADC and Digital Demodulator Opt 160 PDC/PHS/NADC/CDMA FW for Opt 151 HW 85718B NADC/PCS-TDMA Measurement Personality Opt BD3 PDC Bundle Opt 004 Precision Frequency Reference Opt 051 Improved Amplitude Accuracy for PDC Opt 105 Time-Gated Spectrum Analysis Opt 151 DSP, Fast ADC and Digital Demodulator Opt 160 PDC/PHS/NADC/CDMA FW for Opt 151 HW 85720C BDC Test Personality for 8590 E-Series Opt BD4 PHS Bundle Opt 004 Precision Frequency Reference Opt 052 Improved Amplitude Accuracy for PHS Band Opt 105 Time-Gated Spectrum Analysis Opt 151 DSP, Fast ADC and Digital Demodulator Opt 160 PDC/PHS/NADC/CDMA FW for Opt 151 HW 85726B PHS Measurement Personality for 8590 E-Series Component Test Options Opt 010 Tracking Generator (9 kHz to 2.9 GHz) Opt 119 Noise Figure Cable TV Options Opt 011 Tracking Generator (75 ohm) Opt 107 TV Receiver Video Tester Opt 180 TV Picture NTSC/PAL/SECAM Opt 711 50/75 ohm Matching Pad/100 Vdc Block Communications Options Opt 012 Source for DECT Receiver Test Opt 050 Improved Amplitude Accuracy (NADC - TDMA bands) Opt 051 Improved Amplitude Accuracy for PDC Bands Opt 052 Inproved Amplitude Accurancy for PHS Band Opt 053 Improved Amplitude Accuracy for CDMA Bands Opt 105 Time-Gated Spectrum Analysis Opt 110 CT2 Demodulator Opt 111 Group Delay and Amplitude Flatness (also requires the HP 11770A Link Measurement Personality) Opt 112 DECT Demodulator Opt 151 DSP, FAST ADC, and Digital Demodulator Opt 160 PDC, PHS, NACD, and CDMA Firmware for Option 151 Opt 163 GSM/DCS1800 Firmware for Option 151 EMC Options Opt 103 Quasi-Peak Detector, AM/FM Demodulator Warranty and Support Options Opt 0Q8 Factory Service Training Opt UK6 Commercial Calibration Certificate with Test Data Opt AB* Quick Reference Guide in Local Languages Opt W30 Two Additional Years Return-to-HP Service Opt W32 Two Additional Years Return-to-HP Calibration Opt 915 Component Level Information and Service Guide Most options can be retrofitted. Please contact your local HP sales representative. Application Measurement Cards/Personalities HP 85700A Blank 32-KB Memory Card HP 85702A Blank 128-KB Memory Card HP 11770A Link Measurement Personality HP 85712D EMC Measurement Personality HP 85713A Digital Radio Measurement Personality HP 85714A Scalar Measurement Personality HP 85715B GSM Measurement Personality HP 85717A CT2-CAI Measurement Personality HP 85718B NADC-TDMA Measurement Personality HP 85719A Noise Figure Measurement Personality HP 85720C PDC Measurement Personality HP 85721A Cable TV Measurement Personality HP 85722B DCS1800 /PCS1900 Measurement Personality HP 85723A DECT Measurement Personality HP 85724A Broadcast Measurement Personality HP 85725C CDMA Measurement Personality HP 85726B PHS Measurement Personality HP 85727A GSM Multi-band Measurement Personality HP 85921A Cable TV PC Software for HP 85721A Some measurement personalities are not supported by all HP 8590 series models. For complete information, please contact your local HP sales representative.
Available accessories for this product include:
HP 8590 E-Series Portable Spectrum Analyzers, Technical Specifications
HP 8590L Series Portable Spectrum Analyzers, Product Overview
HP 8590 E-Series Spectrum Analyzers and the Options and Attachments that are Compatible with them, Configuration Guide
If you do not have the Adobe Acrobat Reader necessary for viewing PDF documentation, download your free Acrobat Reader now. |
![]() |
Related Information:
Printer Compatibility Guide for Network and Spectrum Analyzers
Related HP 8590 E-Series Portable Spectrum Analyzer Family product datasheets:
Related product families available online:
Page Updated: Thursday November 12 21:44:42 UTC 1998