HP 8596EM EMC Analyzer, 9 kHz to 12.8 GHz

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]
The HP 8596EM is a stand-alone, automated EMC precompliance analyzer that reduces the time it takes to find, measure and document radiated and conducted EMI emissions. It belongs to the HP 8590EM series of EMC analyzers, which do not require an external computer and automation software to perform automated EMC measurements. With an upper frequency of 12.8 GHz, the HP 8596EM can perform precompliance tests on products with clock frequencies up to 2.5 GHz. The HP 8596EM has two modes of operation: EMI analysis and spectrum analysis. With the press of a key, this analyzer converts from EMC analyzer to full-function spectrum analyzer for use in a wide variety of non-EMC applications. Based on our highly versatile HP8590 E-series spectrum analyzers, this EMC analyzer simplifies the EMC measurement process with fully integrated automation features.


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

Frequency Specifications

Frequency Range dc coupled: 9 kHz to 12.8 GHz ac coupled: 100 kHz to 12.8 GHz Band LO Harmonic= N 0 1 9 kHz to 2.9 GHz (dc coupled) 0 1 100 kHz to 2.9 GHz (ac coupled) 1 1 2.75 GHz to 6.5 GHz 2 2 6.0 GHz to 12.8 GHz Frequency Reference (Opt. 004) Aging: ±2 x 10-6/year ±1 x 10-7/year Temperature Stability: ±5 x 10-6 ±1 x 10-8 Initial Achievable Accuracy: ±0.5 x 10-6 ±2.2 x 10-8 Marker Count Accuracy Frequency Span <=10 MHz x N1: +(marker frequency readout x frequency reference error + counter resolution + 100 Hz x N1) Frequency Span >10 MHz x N1: +(marker frequency x frequency reference error + counter resolution + 1 kHz x N1) Counter Resolution Frequency Span <10 MHz x N1: Selectable from 10 Hz to 100 kHz Frequency Span >10 MHz x N1: Selectable from 100 Hz to 100 kHz Frequency Sweep Time Range Span = 0 Hz >1 kHz 20 ms to 100 s Span = 0 Hz (Opt. 101) 20 µs to 100 s Accuracy 20 ms to 100 s: ±3% 20 µs to <20 ms (Opt. 101): ±2% Sweep Trigger: Free Run, Single, Line, Video, External Resolution Bandwidth: 30 Hz to 3 MHz (3 dB) in 1-3-10 sequence. 200 Hz, 9 kHz and 120 kHz (6 dB) EMI bandwidths. Accuracy: ±20% Selectivity (Characteristic) -60 dB/-3 dB 3 kHz to 10 kHz: 15:1 100 kHz to 3 MHz: 15:1 1 kHz, 30 kHz: 16:1 -40 dB/-3 dB 30 Hz to 300 Hz 10:1 Video Bandwidth Range: 1 Hz to 1 MHz in 1, 3 Sequence Stability Noise Sidebands (1 kHz RBW, 30 Hz VBW and sample detector) >10 kHz offset from CW signal: <=-90 dBc/Hz + 20 Log N >20 kHz offset from CW signal: <=-100 dBc/Hz + 20 Log N >30 kHz offset from CW signal: <=-105 dBc/Hz + 20 Log N Residual FM 1 kHz RBW, 1 kHz VBW: <=(250 x N) Hz pk-pk in 100 ms 30 Hz RBW, 30 Hz VBW: <=(30 x N) Hz pk-pk in 300 ms System-Related Sidebands >30 kHz offset from CW signal: <=-65 dBc + 20 Log N (N= LO harmonic) 1N = LO harmonic, N=1 for 91E, 94E, 95E. Frequency reference error = aging rate x period of time since adjustment + initial achievable accuracy + temperature stability.

Amplitude Specifications

Amplitude Range: Displayed Average Noise Level to +30 dBm Maximum Safe Input Level (Input attenuator >10 dB) Average Continuous Power: +30 dBm (1 W) Peak Pulse Power: +50 dBm (100 W) for <10 µs pulse width and <1% duty cycle, input attenuation >30 dB. dc: 0 V (dc coupled) 50 V (ac coupled) Gain Compression >10 MHz: <0.5 dB (total power at input mixer2 = -10 dBm) (Mixer Power Level (dBm) = Input Power (dBm) - Input Atten. (dB).) Displayed Average Noise Level (Input terminated, 0 dB attenuation, 1 Hz VBW, sample detector) 30 Hz RBW 1 kHz RBW 400 kHz to 2.9 GHz: <=-125 dBm <=-110 dBm 2.75 MHz to 6.5 GHz: <=-127 dBm <=-112 dBm 6.0 MHz to 12.8 GHz: <=-115 dBm <=-100 dBm Spurious Responses Second Harmonic Distortion >10 MHz <-70 dBc for -40 dBm tone at input mixer. >2.75 GHz <-100 dBc for -40 dBm tone at input mixer. (or below displayed average noise level) Third Order Intermodulation Distortion >10 MHz <-70 dBc for two -30 dBm tones at input mixer and >50 kHz separation. Other Input Related Spurious <=6.5 GHz <-65 dBc at >30 kHz offset, for -20 dBm tone at input mixer <=12.8 GHz Residual Responses (Input terminated and 0 dB attenuation) 150 kHz to 6.5 GHz <-90 dBm Frequency Response (10 dB attenuation) (dc coupled preselector peaked) Absolute3 Relative Flatness4 9 kHz to 2.9 GHz ±1.5 dB ±1.0 dB 2.75 GHz to 6.5 GHz ±2.0 dB ±1.5 dB 6.0 GHz to 12.8 GHz ±2.5 dB ±2.0 dB 2Mixer Power Level (dBm)= Input Power (dBm) - Input Atten. (dB). 3Absolute referenced to 300 MHz CAL OUT. 4Relative Flatness ref. to midpoint between highest and lowest frequency response deviations.

IF and Display Specifications

Display Range Log Scale: 0 to -70 dB from reference level is calibrated; 0.1, 0.2, 0.5 dB/division and 1 to 20 dB/division in 1 dB steps; eight divisions displayed. Linear Scale: Eight divisions Scale Units: dBm, dBmV, dBµV, V,and W Marker Readout Resolution 0.05 dB for log scale 0.05% of reference level for linear scale Fast Sweep Times for Zero Span: (Opt. 101 or 301) 20 µs to 20 ms <=1 GHz: 0.7% of reference level for linear scale >1 GHz: 1.0% of reference level for linear scale Reference Level Range: same as amplitude range Resolution: 0.01 dB for log scale 0.12% of reference level for linear scale Accuracy: ±0.3 dB @ -20 dBm 0 dBm to -59.9 dBm: ±(0.3 dB + .01 x dB from -20 dBm) Resolution Bandwidth Switching Uncertainty (Referenced to 3 kHz RBW, at ref. level) 3 kHz to 3 MHz RBW: ±0.4 dB 1 kHz RBW: ±0.5 dB 30 Hz to 300 Hz RBW: ±0.6 dB Linear to Log Switching: ±0.25 dB at reference level Display Scale Fidelity Log Maximum Cumulative 0 to -70 dB from reference level 1 kHz to 3 MHz RBW: ±(0.3 +0.01 x dB from reference level) 30 Hz to 300 Hz RBW: ±(0.4 +0.01 x dB from reference level) Log Incremental Accuracy ±0.4 dB/4 dB 0 to -60 dB from reference level Linear Accuracy: ±3% of reference level

Option 010 Tracking Generator Specifications

Option 010 Tracking Generator Frequency Range: 300 kHz to 2.9 GHz Output Level Range: -1 to -66 dBm Resolution: 0.1 dB Absolute Accuracy: (@ 300 MHz, -20 dBm, +28.8 dBmV): ±0.75 dB Vernier Range: 8 dB Accuracy: ±0.8 dB Output Attenuator: Range: 0 to 56, 8 dB steps Output Flatness (>10 MHz): ±2.0 dB Spurious Output Harmonic Spurs (-1 dBmV Output): <-25 dBc Nonharmonic Spurs 300 kHz - 2.0 GHz: <=-27 dBc 2.0 GHz - 2.9 GHz: <=-23 dBc Dynamic Range (Characteristic) Dynamic Range TG Feedthrough (Maximum output level minus TG feedthrough) (>400 MHz) 111 dB <=-110 dBm Power Sweep Range: -1 dBm to -66 dBm in 8 dB increments Resolution: 0.1 dB General Specifications Temperature Range Operating: 0 °C to +55 °C Storage: -40 °C to +75 °C EMI Compatibility: Conducted and radiated interference CISPR Pub. 11 and Messempfaenger Postverfuegung 526/527/79. Audible Noise: <37.5 dBa pressure and <5.0 Bels power (ISODP7779) Power Requirements ON (Line 1): 86 to 127, or 195 to 250 Vrms, 47 to 66 Hz, 103 to 126 Vrms, 400 Hz +10%. Power consumption <300 VA; <100W Standby (Line 0): Power consumption <7 W User Memory (nominal): 32 Kbytes non-volatile RAM Data Storage (nominal): 50 states and traces, internal memory and 8 internal state registers; 24 states and traces, memory card (HP 85700A)

Declarations of Conformity


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

HP 8596EM	ECM Analyzer, 9 kHz to 12.8 GHz
   Opt 004	Precision frequency reference
   Opt 009	LO Output
   Opt 010	Tracking generator (9 kHz to 2.9 GHz)
   Opt 015      Soft Tan Operating/Carrying Case
   Opt 016      Soft Yellow Operating/Carrying Case
   Opt 040	Front panel cover
   Opt 042	Protective soft carrying case
   Opt 043	Replace HP-IB/parallel w/RS232/parallel port
   Opt 101	Fast time domain sweeps
   Opt 102	AM/FM demod standard, order this option is only for
                TV sync requirements
   Opt 301	Combination of 101 and 102
   Opt 703	Delete quasi-peak detector
   Opt 908	Rack mount w/o handles
   Opt 909	Rack mount with handles
   Opt 910	Additional operating manuals
   Opt 915	Support manuals
   Opt UK6      Commercial cal. certificate w/ test data
   Opt W30      3 Year Return Repair Service
   Opt W32      3 Year Calibration Customer Return Svc.
   Opt W34      3 Year Standards Compliant Cal Service
   Opt W50      5 Year Return Repair Service
   Opt W52      5 Year Cust Return Calibration Service
   Opt W54      5 Year Standards Compliant Cal Service

HP 11945A Opt E51 Probe kit (9 kHz to 1 GHz) HP 11947A Transient limiter HP 11966C Biconical antenna (30 to 300 MHz) HP 11966D Log periodic antenna (.2 to 1 GHz) HP 11955A Biconical antenna (typical cal. data) HP 11956A Log periodic antenna (typical cal. data) HP 11966L 10 meter N cable HP 11967C Line impedance stabilization network (25 amp) HP 11967D LISN (10 amp) Opt 001 Replaces NEMA connector with SCHUKO connector Opt 002 Replaces NEMA connector with British connector HP 11966E Ridged horn (1 to 18 GHz) HP 11966N Log periodic antenna (.2 to 5 GHz) HP 11966P Broadband antenna (30 MHz to 1 GHz) HP 11968C Tripod HP 85420E Opt 1EM RF filter section


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

If you do not have the Adobe Acrobat Reader necessary for viewing PDF documentation, download your free Acrobat Reader now. Get Acrobat

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]
Additional help is a click away!

Want to talk to a knowledgeable engineer about this product at our HP Test and Measurement Call Center? Do you have a question you want to send in by e-mail? A measurement set-up you need to fax in? Follow the link above and you will be given a choice to contact an engineer by telephone, e-mail or fax.

This service is available only in the United States. For assistance outside the U.S., contact your local HP DIRECT or HP Sales Office.

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

Related Information:


Related HP 8590 E-Series Portable Spectrum Analyzer Family product datasheets:

Related product families available online:



Privacy Statement    Use of this site indicates that you accept the Terms of Use
Contact HP Test & Measurement. (c) Copyright 1994 - 1998 Hewlett-Packard Company.

Page Updated: Monday January 25 19:22:34 UTC 1999