Wavelength
Range
700 to 1650 nm (182 to 428 THz)
Absolute accuracy, laser lines separated by >=30 GHz ±3 ppm
(±0.005 nm at 1550 nm, ±0.004 nm at 1310 nm)
Differential accuracy(characteristic) ±2 ppm
Minimum resolvable separation (equal power lines input)
20 GHz (0.16 nm at 1550 nm, 0.11 nm at 1300 nm)
Display resolution
0.001 nm, normal update mode
0.01 nm, fast update mode
Power
Calibration accuracy
±0.5 dB (at ±30 nm from 780, 1310, and 1550 nm)
Flatness, 30 nm from any wavelength
1200 to 1600 nm ±0.2 dB
700 to 1650 nm ±0.5 dB
Linearity ±0.3 dB
Polarization dependence,
1200 to 1600 nm ±0.5 dB
700 to 1650 nm ±1.0 dB
Display resolution 0.01 dB
Sensitivity
700 to 800 nm, single line input -20 dBm
800 to 1200 nm, single line input -25 dBm
1200 to 1600 nm, single line input -40 dBm
1600 to 1650 nm, single line input -30 dBm
700 to 1650 nm, multiple lines input
30 dB below total input power, but not
less than single line input sensitivity
Selectivity
2 lines input separated by >=100 GHz 25 dB
2 lines input separated by >=30 GHz 10 dB
Input Power
Maximum displayed level (sum of all lines input)+10 dBm
Maximum safe input level (sum of all lines input)+18 dBm
General Specifications:
Weight: 9 kg (19 lb)
Dimensions: (H x W x D) 140 x 340 x 465 mm
5.5 x 13.4 x 18.3 in
HP 86120B Multi-Wavelength Meter
Opt 011 Replace FC/PC connector interface with Diamond HMS-10/HP connector interface
Opt 013 Replace FC/PC connector interface with DIN 47256 connector interface
Opt 014 Replace FC/PC connector interface with ST connector interface
Opt 017 Replace FC/PC connector interface with SC connector interface
Opt 022 Replace flat physical contact interface with angled physical contact interface
Opt 412 Adds an external 10 dB attenuator with flat FC/PC connector
Opt 417 Adds an external 10 dB attenuator with angled FC/PC connector
(requires option 022)
Opt AXE Rack flange kit with handles
Opt IA4 Rackmount kit without handles
Opt UK5 Nylon carrying case with shoulder strap
Opt UK6 Commercial calibration certificate with test data
Opt UK7 Hard carrying case
Additional connector interfaces can be ordered separately
All orders include the following:
Standard instrument includes a front-panel FC/PC optical fiber connector interface and a user's manual.
HP 86120B Multi-Wavelength Meter
PN 86120-1
| If you do not have the Adobe Acrobat Reader necessary for viewing PDF documentation, download your free Acrobat Reader now. |
|
To help you make the right instrument choices
Related product families available online:

Page Updated: Monday October 05 19:16:26 UTC 1998