HP 8642A Synthesized Signal Generator, 100 kHz to 1050 MHz |
The HP 8642A synthesized signal generator is a high-performance,
programmable signal generator for demanding out-of-channel RF receiver
measurements and other stringent RF applications. It covers the
frequency range from 100 kHz to 1057.5 MHz. This signal generator
provides low SSB phase noise and spurious, high output level and level
accuracy, and a full range of modulation types. The low-distortion
internal modulation oscillator can be used to modulate the HP 8642A up
to 100 kHz rates. This internal oscillator can also be used as an
independent audio source with variable rates and levels.
Frequency Specifications
Range: 100 kHz to 1057.5 MHz Bands: The full range is covered in one continuous span. However, many specifications are dependent on carrier frequency. To simplify such specifications, the carrier frequency ranges are divided into bands as shown below: Band Carrier Frequency (MHz) 9 528.750001 to 1057.5 8 264.375001 to 528.75 7 132.187501 to 264.375 6 66.093751 to 132.1875 5 33.046876 to 66.09375 4 16.523438 to 33.046875 3 8.261719 to 16.523437 2 4.130860 to 8.261718 1 0.1 to 4.130859 HET 0.1 to 132.1875 Sweep Modes: Start-stop, span, and phase continuous X-Axis Output: 0 to 10 Vdc, ±10% Y-Axis Output: TTL positive true for display blanking during retrace Resolution: 1 Hz, 0.1 Hz with special function Stability Standard: Aging rate: ±2 ppm/year Option 001: <10e-9/day aging rate after 8 days warmup
Spectral Purity Specifications
Residual FM (in CW, AM or Angle Modulation <1/3 max dev.) 500 MHz: <1.2 Hz (0.3 to 3 kHz BW), <2 Hz (0.05 to 15 kHz BW) 1000 MHz: <2 Hz (0.3 to 3 kHz BW), <5 Hz (0.05 to 15 kHz BW) SSB Phase Noise at 20 kHz Offset (CW, AM or FM/PhiM <1/60 max dev) 125 MHz: -144 dBc/Hz 200 MHz: -141 dBc/Hz 500 MHz: -137 dBc/Hz 1000 MHz: -134 dBc/Hz Residual AM: <0.01% AM rms, 0.3 to 3 kHz BW Spurious Harmonics: -30 dBc at <= +10 dBm Subharmonics: None, fc <=1057.5 MHz Nonharmonics (>10 kHz offsets): -100 dBc Third Order Intermodulation: < -50 dBc at +10 dBm, 2 generators 25 kHz apart into a resistive combiner. Typically decreases 10 dB for every 5 dB of combined level decrease.
Output Specifications
Level Range: From maximum available to -140 dBm (0.023 µV) Maximum Level Available bands 1 thru 7: +20 dBm (2.24 V) bands 8 and HET: +18 dBm (1.78 V) band 9: +16 dBm (1.41 V) Resolution: 0.01 dB Absolute Accuracy: ±1 dB, output level >= -127 dBm Flatness: <= ±0.75 dB, +10 dBm output level Impedance: 50 ohm nominal SWR: <1.5:1, level<0 dBm; <2.0:1, level >=0 dBm Reverse Power Protection: 50W, from a 50 ohm source 25 Vdc
Modulation Specifications
Amplitude Modulation AM Depth: 0 to 99.9%, output level <= +10 dBm AM Resolution: 0.1% AM Indicator Accuracy at 1 kHz Rate and Up to 90% AM: ±(3.5% of setting + 1% AM), fc <=528.75 MHz ±(5% of setting + 1% AM), fc >528.75 MHz AM Distortion at 1 kHz Rate: Depth Distortion fc <=528.75 MHz fc >528.75 MHz 0 to 30% AM <1% <2% 30 to 70% AM <2% <4% 70 to 90% AM <4% <6% AM 3 dB Bandwidth, Depth <=90% External DC/AC Coupling:dc/20 Hz to 100 kHz, fc: 01 to 4.13 MHz, 33.04 to 2115 MHz, HET; dc/20 Hz to 20 kHz, fc: 4.13 to 33.04 MHz Internal: Same as external AC Incidental PhiM at 1 kHz Rate and 30% AM: <0.2 radians peak Frequency Modulation Maximum FM Deviation: Carrier Maximum deviation Maximum deviation frequency dc-coupled ac-coupled or internal band (the smaller of) 9 1.5 MHz 1.5 MHz or f mod X 1080 8 750 MHz 750 kHz or f mod X 540 7 375 kHz 375 kHz or f mod X 270 6 187 kHz 187 kHz or f mod X 135 5 93.8 kHz 93.8 kHz or f mod X 67.5 4 46.9 kHz 46.9 kHz or f mod X 33.75 3 23.4 kHz 23.4 kHz or f mod X 16.88 2 11.7 kHz 11.7 kHz or f mod X 8.44 1 93.8 kHz 93.8 kHz or f mod X 67.5 HET 1.5 MHz 1.5 MHz or f mod X 1080 FM Resolution: 0.7% of setting or 0.0004% of maximum deviation, whichever is larger FM Indicator Accuracy, Rates <= 100 kHz:±(5% of setting + 10 Hz) FM Distortion, Rates 20 Hz to 100 kHz: 4% for max. dev., 2% for 1/2 max. dev., 0.4% for 1/15 maximum dc-coupled deviation FM 3 dB Bandwidth (dc/ac coupling): dc/20 Hz to 200 kHz (external) Incidental AM: 0.3%, 20 kHz peak dev., 1 kHz rate, fc >400 kHz Phase Modulation Maximum Phase Deviation: Carrier frequency band Maximum deviation (radians) 9 100 8 50 7 25 6 12.5 5 6.25 4 3.13 3 1.56 2 0.78 1 6.25 HET 100 PhiM Accuracy: ±(5% of setting + 0.09 radians), 1 kHz rate Phi Resolution: greater of 0.7% of setting or 0.0004% of maximum deviation PhiM Distortion: <0.4%, 1 kHz rate PhiM 3 dB Bandwidth: DC/20 Hz to 15 kHz Pulse Modulation (for output levels <= +15 dBm) Pulse On/Off Ratio: >40 dB; >80 dB, fc >1057.5 MHz Rise/Fall Time: <400 ns, 10 % to 90% Maximum Repetition Frequency: 100 kHz Minimum Pulse Width: 2 µs Internal Modulation Oscillator Rates: 20 Hz to 100 kHz Frequency Resolution: 1% of setting Accuracy: 2% ofsetting Output Level Range: 0 to 3 V peak into 600 ohns Resolution: 4 mV Accuracy: ±(4% + 15 mV) within 1 second Distortion: <0.02%, 0.02 kHz to 15.8 kHz; <0.15%, >15.8 kHz Output Impedance: 600 ohms ±10%
Remote Programming Specifications
Interface: HP-IB (IEEE-488-1978) HP-IB Functions: Listener, talker, and controller. SH1, AH1, T5, TE0, L3, LE0, SR1, RL1, PP1, DC1, DT1, C1, C3, C28, E2
General Specifications
Operating Temperature Range: 0° to 55°C Storage Temperature Range: -55° to +75°C Leakage:Conducted and radiated interference is within the requirments of MIL-STD-461V method RE02. Interference is also within the standards set by FTZ-1046. Also, RF leakage of <0.5 µV is induced in a 2 turn loop 2.5 cm in diameter, held 2.5 cm away from any surface for ouptut levels <=0 dBm. Power Requirements: 100, 120, 220, or 240 V; +5%, -10%; 48 to 440 Hz; 300 VA max Size: 425 mm W x 133 mm H x 617 mm D (16.75 in x 5.25 in x 24.3 in) HP System II Module Size: 1 W x 5 1/4 H x 23 D Weight: Net, 32.7 kg (71.5 lb); Shipping, 43 kg (95 lb)
HP 8642A Synthesized Signal Generator¹ Opt 001 High-Stability Time Base Opt 002 Input/Output Connectors on Rear Panel Only Opt 1BN MIL-STD 45662A Certificate of Calibration Opt 1BP MIL-STD 45662A Calibration w/ Test Data Opt 710 On-Site Repair Manual Opt 907 Front Handle Kit Opt 908 Rack Flange Kit with Front Handles Opt 910 Additional Operation and Calibration Manual Opt 915 Add Service Manual HP 8642A Opt W30 Extended Repair Service HP 11801A On-Site Repair Kit for HP 8642A ¹HP-IB cables not supplied
Page Updated: Thursday June 19 16:14:27 UTC 1997