HP 8643A Synthesized Signal Generator, 252 kHz to 1030 MHz |
The HP 8643A signal generator has been optimized to provide the
performance required for out-of-channel receiver tests while maintaining
a low price. Many performance capabilities have been included as
standard, such as pulse modulation. Reliability is enhanced by an
electronic attenuator on the 1 GHz version. Instead of mechanical
relays, the HP 8643A uses solid state components accurate to within 1.0
dB for setting levels. An advanced internal modulation synthesizer
provides coverage to 400 kHz and two-tone capability with the selection
of sine, square, triangle, sawtooth, and white Gaussian noise waveforms.
Frequency Specifications
Range: 0.252 to 1030 MHz Opt 002: 0.252 to 2060 MHz Resolution: 0.01 Hz Stability: 3 x 10e-10/day (Opt 001) Switching Speed (typ): <90 ms; <200 ms with FM on Sweep Digital Sweep: Digitally stepped sweep over entire frequency range. Linear/log selection. 0.5 to 1000 s sweeps. Markers/Z-Axis Output: 3 markers available/Z-axis output nominally +5 V/X-axis output nominally 0 to 10 V. Phase Continuous Sweep: 40 MHz of span available at maximum carrier frequency. 20 ms to 10 s sweep times.
Spectral Purity Specifications
SSB Phase Noise @ 1 GHz (20 kHz offset): -130 dBc/Hz Nonharmonics (>10 kHz offset): -100 dBc, 0.252 to 1030 MHz < -94 dBc, 1030 to 2060 MHz Harmonics: < -27 dBc, output <= +8 dBm Subharmonics: None, 0.252 to 515 MHz < -60 dBc, 515 to 1030 MHz < -40 dBc, 1030 to 2060 MHz Residual FM @ 1 GHz: 0.3 to 3 kHz post det. BW: <2 Hz rms
Output Specifications
Range: +13 to -137 dBm Resolution: 0.1 dB Absolute Accuracy: ±1 dB, output >= -127 dBm Reverse Power Protection: 50 W
Modulation Specifications
Amplitude Modulation Depth (@: <= +7 dBm): 0 to 100% Resolution: 0.1% Bandwidth (3 dB) <= +7 dBm: dc to >100 kHz, > 128 MHz Accuracy (1 kHz rate): ±(7% of setting + 1%) up to 80% depth Distortion (30% depth, 1 kHz rate): <2% Opt 002: <4% Frequency Modulation Maximum Peak Deviation: 2 MHz, 1030 to 2060 MHz (Deviation halves per lower octave): 1 MHz, 515 to 1030 MHz Resolution: 2.5% of setting Bandwidth (3 dB): dc to 100 kHz Carrier Accuracy in FM: ±0.5% of setting Indicator Accuracy: <5%, <30 kHz rate: <10%, <100 kHz rates Distortion: <5%, 20 Hz to 100 kHz rates Pulse Modulation On/Off Ratio: >50 dB; >80 dB, >1030 MHz Rise/Fall Time, 10 to 90%: <100 ns Repetition Rate: dc to 1 MHz Internal Modulation Source Waveforms and Rates: Sine; white Gaussian noise (0.1 Hz to 400 kHz); Triangle, sawtooth, square (0.1 Hz to 50 kHz) Frequency Accuracy: Same as timebase Output Level (into 600 ohms): 2 V pk Output Resolutions: 2 mV pk
Remote Programming Specifications
Interface: HP-IB (IEEE 488.2-1987) Control Language: HP-SL; all functions controlled except power-line switch IEEE-488 Functions: SH1, AH1, T6, TE0, L4, LE0, SR1, RL1, PP0, DC1, DT0, C0, E2
General Specifications
Power Requirements: ±10% of 100 V, 120 V, 220 V, or 240 V; 48 to 440 Hz; 400 VA Operational Temperature: 0° to 55°C Leakage: Conducted and radiated interference meets MIL-STD-461B REQ2 and FTZ 1046 Calibration Interval: Recommended 3 years (MTBC) Weight: 23 kg Size: 177 mm H x 426 mm W x 624 mm D (7 in x 16.8 in x 24.6 in) Option 010 adds 35 mm (1.4 in) to D
HP 8643A Signal Generator Opt 001 High-Stability Timebase Opt 002 2 GHz Doubled Output Opt 003 Rear-Panel Input/Output Opt 009 Specified VOR/ILS¹ Opt 010 Reduced-Leakage Configuration Opt 011 2 GHz Internal Frequency Counter Opt 907 Front-Handle Kit Opt 908 Rack Flange Kit Opt W30 Add 3 Years to Return Warranty Service Kit HP 08645-61116 ¹Not compatible with Option 002
Page Updated: Thursday June 19 16:14:19 UTC 1997