HP 8644B High-Performance Synthesized Signal Generator, 252 kHz to 1030 MHz

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]
The HP 8644B signal generator represents the highest overall performance in HP's line of 1 GHz and 2 GHz signal generators. This generator has very low SSB phase noise (-136 dBc/Hz) and spurious (-105 dBc). Instead of mechanical relays, the HP 8644B uses solid state components accurate to within 1.0 dB for setting levels. An advanced internal modulation synthesizer provides coverage to 400 kHz and two-tone capability with the selection of sine, square, triangle, sawtooth, and white Gaussian noise waveforms.


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

Frequency Specifications

Range: 0.252 to 1030 MHz Opt 002: 0.252 to 2060 MHz Resolution: 0.01 Hz Accuracy: Timebase stability x fc Switching Speed: <350 ms, typical Sweep Digital Sweep: Digitally stepped sweep over entire frequency range. Linear/log selection. 0.5 to 1000 s sweeps. Markers/Z-Axis Output: 3 markers available/Z-axis output nominally +5 V/X-axis output nominally 0 to 10 V. Phase Continuous Sweep: 40 MHz of span available at maximum carrier frequency. 20 ms to 10 s sweep times.

Spectral Purity Specifications

SSB Phase Noise @ 1 GHz (20 kHz offset): -136 dBc/Hz Nonharmonics (>10 kHz offset): -105 dBc, 0.252 to 1030 MHz <-100 dBc, 1030 to 2060 MHz Harmonics: <-25 dBc, output <=+10 dBm Subharmonics: None, 0.252 to 515 MHz <-52 dBc, 515 to 1030 MHz <-40 dBc, 1030 to 2060 MHz Residual FM @ 1 GHz: 0.3 to 3 kHz post det. BW: <1 Hz rms

Output Specifications

Range: +16 to -137 dBm +13 dBm (Opt 002, 005) Resolution: 0.1 dB Absolute Accuracy: ±1 dB, output >=-127 dBm Reverse Power Protection: 50 W

Modulation Specifications

Amplitude Modulation Depth (@: <=+7 dBm): 0 to 100% Resolution: 0.1% Bandwidth (3 dB) <=+7 dBm: dc to 100 kHz, 128 MHz <fc <1030 mHz Accuracy (1 kHz rate): ±(7% of setting + 1%) up to 80% depth Distortion (30% depth, 1 kHz rate): <2% Opt 002: <4% Frequency Modulation Maximum Peak Deviation: 20 MHz/200 kHz1, <1030 MHz (Deviation halves per lower octave): 10 MHz/100 kHz1, <515 MHz Resolution: 2.5% of setting Bandwidth (3 dB): dc to 100 kHz Carrier Accuracy in FM: ±0.5% of setting Indicator Accuracy: <5%, <30 kHz rate: <10%, <100 kHz rates Distortion: <5%, 20 Hz to 100 kHz rates 1Low-noise mode Pulse Modulation On/Off Ratio: >35 dB; >80 dB, >1030 MHz Rise/Fall Time, 10 to 90%: <100 ns Repetition Rate: dc to 1 MHz Internal Modulation Source Waveforms and Rates: Sine; white Gaussian noise (0.1 Hz to 400 kHz); Triangle, sawtooth, square (0.1 Hz to 50 kHz) Frequency Accuracy: Same as timebase Output Level (into 600 ohms): 2 V pk Output Resolutions: 2 mV pk 1Low-noise mode

Remote Programming Specifications

Interface: HP-IB (IEEE 488.2-1987) Control Language: HP-SL; all functions controlled except power IEEE-488 Functions: SH1, AH1, T6, TE0, L4, LE0, SR1, RL1, PP0, DC1, DT0, C0, E2

General Specifications

Power Requirements: ±10% of 100 V, 120 V, 220 V, or 240 V; 48 to 440 Hz; 400 VA Operational Temperature: 0 °C to 55 °C Leakage: Conducted and radiated interference meets MIL-STD-461B REQ2 and FTZ 1046 Calibration Interval: Recommended 3 years (MTBC) Weight: 30 kg Size: 177 mm H x 426 mm W x 624 mm D (7 in x 16.8 in x 24.6 in) Option 010 adds 35 mm (1.4 in) to D

Declarations of Conformity


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

HP 8644B	High-Performance Synthesized Signal Generator,
                252 kHz to 1030 MHz

   Opt 001	High-Stability Timebase
   Opt 002	2-GHz Doubled Output
   Opt 003	Rear-Panel Input/Output
   Opt 005	Electronic Attenuator1
   Opt 009	Specified VOR/ILS1
   Opt 010	Reduced-Leakage Configuration
   Opt 011	2-GHz Internal Frequency Counter
   Opt 907	Front-Handle Kit
   Opt 908	Rack Flange Kit
   Opt W30	Add 3 Years to Return Warranty
   Service Kit  HP 08645-61116


1Not compatible with Option 002


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

No literature is available online at this time.

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

To help you make the right instrument choices

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

Related product families available online:



Privacy Statement    Use of this site indicates that you accept the Terms of Use
Contact HP Test & Measurement. (c) Copyright 1994 - 1998 Hewlett-Packard Company.

Page Updated: Monday October 05 19:59:32 UTC 1998