HP 8657A Option H46 0.5 GMSK Modulated Synthesized Signal Generator |
The HP 8657A Option H46 signal generator provides 0.5 Gaussian Minimum Shift Keying (0.5 GMSK) modulation for testing Cellular Digital Packet Data (CDPD) systems. This option contains the necessary premodulation filtering and signal-processing circuitry to generate precise 0.5 GMSK modulation over the entire 520 to 1040 MHz and 10 to 130 MHz frequency bands. The versatility of the signal generator provides the CDPD system designer with source signals which simplify CDPD system performance verification. The AM features of the instrument can be used to stimulate amplitude fluctuations, while dcFM can be used to add doppler shifts to test signals.
The HP 8657A Option H64 is also a versatile signal generator offering
a variety of features, such as simultaneous and mixed modulation
combinations of AM and FM, as well as a carrier phase adjust
capability that allows you to characterize phase-sensitive devices.
The low residual FM and excellent phase noise performance of this
signal generator makes it ideal for m any in-channel and even
out-of-channel receiver tests. It offers a wide output range, +13 to
-143.5 dBm with superior level accuracy of 1.0 dB. Built for
reliability, this signal generator uses an all-electronic attenuator,
which is backed with a five-year warranty to ensure that you will make
consistent, repeatable measurements every time.
Frequency Specifications
Range: 100 kHz to 1040 MHz Resolution: 10 Hz Timebase aging rate: ±2 ppm/year (typ) Option 001: 1.5 x 10e-6 parts/day after 10 days (typ.) Switching speed (w/i 100 Hz): <35 ms (30 ms typ. at 25°C)
Spectral Purity Specifications
SSB phase noise (fc = 500MHz, 20 kHz offset): <-130 dBc/Hz Nonharmonics (>5 kHz offset, CW): <-60 dBc Harmonics (at levels <=+7 dBm): <-30 dBc Subharmonics: None Residual FM (0.3 to 3 kHz, CW): <4 Hz rms
Output Specifications
Range: +13 to -143.5 dBm into 50 ohms, +10 dBm, 100 kHz to 1 MHz Resolution: 0.1 dB Absolute Accuracy: < ±1 dB, >=-127 dBm < ±1.5 dB, >+7 dBm Reverse Power Protection: 50 W
0.5 GMSK Digital Modulation Mode Specifications
Frequency Range Date Normal: 520 to 1040 MHz Date Inverted: 10 to 130 MHz Harmonics (fc <1030 MHz): <-30 dBc Non-Harmonic Spurious (>150 kHz offset): <-50 dBc Noise Floor: <-100 dBc/Hz Modulation Format: 0.5 GMSK Pre-modulation Filter: Gaussian LPF Data Rate: 19.2 kHz (±1 kHz) Modulation Phase Error: <3° rms (890-960 MHz, 10-130 MHz): <8° peak Frequency Error (fc <1030 MHz): 10 Hz Inputs: Serial data and clock
Frequency Modulation Specifications
Maximum Peak Deviation: 99 kHz, fc < 130 and > 260 50 kHz, fc 130 to 260 MHz Resolution: 100 Hz, deviations < 10 kHz, 1 kHz, deviations > 10 kHz Bandwidth (1 dB): dc to 50 kHz Indicator Accuracy: <±5% of setting Distortion (internal rates, >=3 kHz peak deviation): <0.5 % THD + noise
Amplitude Modulation Specifications
Depth (output at <=+7 dBm): 0 to 99% fc >= 400 kHz Resolution: 1% Bandwidth (1 dB): 20 Hz to 40 kHz Accuracy (internal rates): <±(2% + 6% of setting) (<90% AM < + 7 dBm) Distortion: <1.5% (0 to 30% AM, internal rates, level < + 7 dBm)
0.5 GMSK Specifications
Frequency Range: 10 to 130 MHz 520 to 1040 MHz Modulation Phase Error: <3° rms.
General Specifications
Remote Programming Interface: HP-IB (HP's implementation of IEEE-488) IEEE-488 functions: SH0, AH1, T0, L2, SR0, RL1, PP0, DC1, DT0, C0, E1 Save/recall/sequence registers: 100 nonvolatile registers to save front-panel settings Operating temperature: 0°C to 55°C 0.5 GMSK Modulation: +15°C to 35°C Power requirements: 100 V, 120 V, 220 V, or 240 V (+5, -10%) from 48 to 440 Hz, 175 VA maximum 0.5 GMSK Modulation: 90 TO 264 V from 48 to 66 Hz; 75 VA maximum Size: 133 mm H x 425 mm W x 520 mm D (5.25 in x 16.75 in x 20.5 in) 0.5 GMSK Modulation: 44 mm H x 425 mm W x 422 mm D (1.75 in x 16.75 in x 17 in) Weight: 18.2 kg (40 lb) 0.5 GMSK Modulation: 4 kg (9 lb)
HP 8657A Opt H46 0.5 GMSK Modulated Synthesized Signal Generator Opt 001 High-Stability Timebase Opt 907 Front Handle Kit (HP 5061-9689) Opt 908 Rack Flange Kit (HP 5061-9677) Opt 910 Extra Operation/Calibration and Two Service Manuals Opt 915 Add Service Manual Opt +W30 3-year Repair Service Opt +W32 3-year Calibration Service