HP 8703A Lightwave Component Analyzer, .13-20 GHz![]()
|
Measurements can be made on connectors, splitters, couplers, and lenses, as well as fiber. This yields insertion loss, length, and optical return loss. In the distance-time domain, reflections can be located without the dead zone typical of OTDR type measurements. Transmission measurements can also be displayed in the distance-time domain to view the impulse or step response of the component. Delay and dispersion are easily viewed in this manner.
When used to measure linear electrical components, such as amplifiers, filters, and transmission lines, the lightwave component analyzer has the full measurement capability of a microwave network analyzer. Typical measurements are bandwidth, insertion loss/gain, phase, impedance, match, and group delay.
Electrical Source Specifications
Frequency Range: 130 MHz to 20 GHz Resolution: 1 Hz Output Power Range: +5 to -50 dBm in 5 dB steps at Port 1; -15 to -70 dBm in 5 dB steps at Port 2. Power Level: +5 dBm ±3 dB at Port 1; -15 dBm plus coupler roll-off at Port 2. Harmonics: <-20 dBc at +5 dBm (typical) Test Ports Connector Type: 3.5 mm (male) Impedance: 50 ohms nominal Maximum Input Level: +20 dBm DC Bias: 500 mA, 40 Vdc max
Dynamic Range (Typical)
Frequency Range: 0.13 to 20 GHz Lightwave Forward Transmission Test: Optical-to-Optical >40 dB Optical-to-Electrical >100 dB Electrical-to-Optical >70 dB Lightwave Reflection Test Optical-to-Optical >30 dB Electrical Forward Transmission (S21) >90 dB Reverse Transmission (S12) 0.13 to 0.5 GHz >40 dB 0.5to 2 GHz >60 dB 2 to 20 GHz >70 dBLightwave Source Characteristics
Wavelength: 1308 or 1550 nm, +/-10 nm Spectral width: 50 MHz Average optical output power: 70 to 600 µW Modulated optical output power (p-p): 90 to 130 µW Modulation index: 25% Optical return loss: 15 dBoLightwave Receiver Characteristics
Wavelength: 1298 to 1560 nm Input modulation bandwidth: 130 MHz to 20 GHz Max average input power op. level: 5 mW System sensitivity: 20 nW Input port return loss: 20 dBo
General Specifications
Weight: 50 kg (110 lb) Dimensions: (H x W x D) 370 x 425 x 502 mm 14.57 x 16.73 x 19.76 in. Allow 50 mm (2.0 in.) additional depth for front panel connectors.
HP 8703A Lightwave Component Analyzer, .13-20 GHz Opt 01X Select Optical Connector Opt 100 External Lightwave Source Input Opt 210 1550 nm DFB Laser Opt 220 1300 nm DFB Laser Opt 300 Additional Lightwave Receiver Opt 804 Delete HP 9122 disk drive & HP-IB cable Opt 830 3.5mm economy cal. kit & test port cable Opt 908 Rackmount kit Opt 910 Extra operating and service manual setStandard configuration includes a laser and one 1300/1550 nm receiver. One laser option must be specified when ordering. The external lightwave source input (Option 100) is used with lasers of the following wavelength: 1530 to 1570 nm (with Option 210) 1290 to 1330 nm (with Option 220) The HP 83424A and 83425A are lightwave CW sources recommended for use with Option 100.
HP 8703A Lightwave Component Analyzer
HP 8703A Lightwave Component Analyzer
If you do not have the Adobe Acrobat Reader necessary for viewing PDF documentation, download your free Acrobat Reader now. |
![]() |
Related product families available online:
Page Updated: Monday October 05 19:40:03 UTC 1998