HP 8757E Scalar Network Analyzer
| |||
![[ Summary ]](/tmo/datasheets/Graphics/tab-summary-high.gif)
![[ Features ]](/tmo/datasheets/Graphics/tab-features-reg.gif)
![[ Specifications ]](/tmo/datasheets/Graphics/tab-specs-reg.gif)
![[ Key Literature ]](/tmo/datasheets/Graphics/tab-keylit-reg.gif)
![[ Ordering ]](/tmo/datasheets/Graphics/tab-ordering-reg.gif)
![[ Assistance ]](/tmo/datasheets/Graphics/tab-assistance-reg.gif)
System Accuracy
Insertion Loss or Gain Measurement Accuracy
Transmission loss or gain measurements are made relative to a
0 dB reference point established at calibration.
Transmission Measurement Uncertainty
= dynamic power accuracy + mismatch uncertainty
Dynamic power accuracy is the measurement uncertainty due to the
change in power level between calibration and the measurement.
Mismatch uncertainty is the uncertainty due to reflections in the
measurement setup. The frequency response errors of the source,
detectors, bridge, and power splitter are removed via calibration.
Transmission Measurement Uncertainty Examples
Assumptions:
* Measurement frequency = 10 GHz
* DUT input/output SWR = 1.5
* Change in power after calibration <30 dB (+0 to -30 dBm range)
Uncertainty component HP 85037B HP 85025E HP 11664E
precision detector detector detector
----------------------------------------------------------------------------------
Dynamic Accuracy (±dB) 0.11 0.40 0.30
Mismatch (±dB) 0.45 0.33 0.61
----------------------------------------------------------------------------------
Uncertainty Total (±dB) 0.56 0.73 0.91
Absolute Power Measurement Accuracy
This specification is useful for determining the accuracy
of power measurements in dBm when using the HP 85037 series
precision detectors or the HP 85025 series precision detectors
in dc mode.
Absolute power uncertainty = absolute power accuracy at 50 MHz
+ frequency response + mismatch uncertainty
Absolute Power Measurement Uncertainty Examples
Assumptions:
* Measurement frequency = 10 GHz
* DUT input/output SWR = 1.5
* Measured power = 0 dBm
Uncertainty component HP 85037B HP 85025E
detector detector
-------------------------------------------------------------
Absolute Power
accuracy at 50 MHz (±dB) 0.11 0.40
Frequency response (±dB) 0.18 0.50
Mismatch (±dB) 0.18 0.10
-------------------------------------------------------------
Uncertainty Total (±dB) 0.47 1.00
Reflection Measurement Accuracy
Uncertainties due to calibration error and the frequency
response of the source, detectors, and bridges are removed
via open/short averaging. The remaining uncertainties are
primarily the sum of directivity uncertainty, effective source
match uncertainty, and dynamic power accuracy. See technical
data sheet for further information.
If you have Adobe's Acrobat Reader already available on your system, please click on the icon provided below to print or view the specifications.
If you do not have the Adobe Acrobat Reader necessary for viewing PDF documentation, download your free Acrobat Reader, and then click on the icon below.
HP 8757E Scalar Network Analyzer
Opt 1BP Adds certificate of calibration in full compliance with MIL-STD 45662A.
Data Included
Opt 908 Adds rack flange kit for instruments without handles
Opt 910 Adds extra manual set
Opt 913 Adds rack flange kit for instruments with handles
HP 8757XC 20 GHz coaxial synthesized scalar system
Includes: HP 8757D scalar network analyzer
HP 83752A synthesized sweeper¹
HP 85022A cable kit
Opt 001 Adds a fourth detector input to analyzer
Opt 002 Adds 50-MHz calibrator to analyzer
Opt 1E1 Adds 70-dB step attenuator to source
Opt 1E5 Adds high-stability timebase to source
Opt 1ED Adds type-N connector to source
Opt 802 Adds an HP 9122C dual 3.5 in. disk drive and an HP 10833A HP-IB cable
Opt 57E Substitutes HP 8757E analyzer
Opt 51A Substitutes HP 83751A source (2 to 20 GHz)
Opt 51B Substitutes HP 83751B source (2 to 20 GHz, high power)
Opt 52B Substitutes HP 83752B source (0.01 to 20 GHz, high power)
Opt 910 Adds extra manual set
Opt 1BN Adds certificate of calibration in full compliance with MIL-STD 45662A
Opt 1BP Adds certificate of calibration in full compliance with MIL-STD 45662A
Data included.
¹Several options are available for replacing the HP 83752A synthesizer, with
HP 8360-series or HP 8370-series synthesized sources.
Sweepers
Choose the HP 83752A/B Synthesized Sweepers for applications from 10 MHz to 20 GHz;
the HP 8360 Series Synthesized Sweepers for applications up to 50 GHz in coax or 110 GHz.
in waveguide.
Precision Detectors
HP 85037A 0.01 to 18 GHz, 7-mm, 50 Ohms
Opt 001 7-mm Connector
HP 85037B 0.01 to 26.5 GHz, 3.5 mm (m)
Detectors
HP 85025A 0.01 to 18 GHz, Type-N (m)
Opt 001 7-mm Connector
HP 85025B 0.01 to 26.5 GHz, 3.5 mm (m)
HP 85025D 0.01 to 50 GHz, 2.4 mm (m)
HP 85025E 0.01 to 26.5 GHz, mm (m) with improved return loss
Directional Bridges²
HP 85027A 0.01 to 18 GHz, 7mm, 50 ohms
HP 85027B 0.01 to 26.5 GHz, 3.5mm (f), 50 ohms
HP 85027C 0.01 to 18 GHz Type-N (f), 50 ohms
HP 85027D 0.01 to 50 GHz, 2.4mm (m), 50 ohms
HP 85027E 0.01 to 26.5 GHz 3.5 mm (m), 50 ohms
Optional Accessories
(For ratio and/or modulation measurements)
HP 11636A Power Divider DC to 18 GHz
HP 11636B Power Divider DC to 26.5 GHz
HP 11667A Power Splitter DC to 18 GHz
Opt 001 N-male on Input Port; N-female on Output Ports
Opt 002 N-female on Input Port; 7 mm on Output Ports
HP 11667B Power Splitter DC to 26.5 GHz
HP 11667C Power Splitter DC to 50 GHz
HP 11679A Detector Extension Cable, 7.6 m (25 ft)
HP 11679B Detector Extension Cable, 61 m (200 ft)
HP 11852B 50 to 75 ohm Minimum Loss Pad
Service and Support Products
HP 11613B Calibrator
Upgrade Kits
HP 86383C Upgrade Kit for HP 8757D
Opt 001 Adds Fourth Detector Input
Opt 002 Adds Internal Power Calibrator
²Connector types refer to the test port (or output) connector unless
otherwise specified.
HP 8757D/E Scalar Network Analyzers, Technical Specifications
(
PDF)
HP 8757D/E Scalar Network Analyzers, Configuration Guide
(
PDF)
| If you do not have the Adobe Acrobat Reader necessary for viewing PDF documentation, download your free Acrobat Reader now. |
|
![[ Summary ]](/tmo/datasheets/Graphics/tab-summary-reg.gif)
![[ Features ]](/tmo/datasheets/Graphics/tab-features-reg.gif)
![[ Specifications ]](/tmo/datasheets/Graphics/tab-specs-reg.gif)
![[ Key Literature ]](/tmo/datasheets/Graphics/tab-keylit-reg.gif)
![[ Ordering ]](/tmo/datasheets/Graphics/tab-ordering-reg.gif)
![[ Assistance ]](/tmo/datasheets/Graphics/tab-assistance-high.gif)
Additional help is a click away!
Want to talk to a knowledgeable engineer about this product at our HP Test and Measurement Call Center? Do you have a question you want to send in by e-mail? A measurement set-up you need to fax in? Follow the link above and you will be given a choice to contact an engineer by telephone, e-mail or fax.
This service is available only in the United States. For assistance outside the U.S., contact your local HP DIRECT or HP Sales Office.