HP 8900D Peak Power Meter


[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

Product Summary

The HP 8900D power meter directly displays the peak power of RF pulses over a 100 MHz to 18 GHz frequency range. Measurements can be made on pulses with widths from 1 µs (100 ns in compare mode) to CW, and repetition rates from 100 Hz (0 Hz in compare mode) to 100 kHz. The HP 8900D has a high-resolution 3 and 1/2 digit digital display calibrated in watts. The direct reading display and range annunciator make the digital version a good choice for productions and field applications where unambiguous or frequent readings are required.


[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

Product Features



[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

Product Specifications

General Specifications

Frequency Range: 100 MHz to 18 GHz Dynamic Range (2 ranges of 10 and 100 mW full scale): 20 dB (0 to +20 dBm) Pulse Response Direct Mode Pulse Width: 1 µs to CW Repetition Rate: 100 Hz to 100 kHz Compare Mode Pulse Width: 100 ns (typical) limited by rise-time specification Repetition Rate: 0 to 100 kHz Rise-Time: 75 ns Fall-Time: 125 ns (as measured on video output) Power Consumtption: 100 and 120 Vac +5, -10%, 48 to 66 Hz and 360 to 440 Hz; 220 and 240 Vac +5, -10%, 48 to 66 Hz

Meter Accuracy

CW Direct: ±0.2 dB Compare: ±0.2 dB Pulse Direct: ±0.35 dB Compare: ±0.25 dB Transfer Accuracy CW to Pulse Direct: ±0.2 dB Compare: ±0.2 dB



[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

Product Ordering Information

HP 8900D	Digital Peak Power Meter
   Opt W30	Extended Repair Service
   Opt W32	Calibration Service



[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

To help you make the right instrument choices



Contact HP Test & Measurement. (c) Copyright 1994, 1995, 1996, 1997 Hewlett-Packard Company.

Page Updated: Thursday June 19 16:13:23 UTC 1997