HP 8901A Modulation Analyzer, 150 kHz to 1300 MHz |
The HP 8901A modulation analyzer combines the capabilities of several RF
instruments to give complete, accurate characterization of modulated
signals in the 150 kHz to 1300 MHz frequency range. It very accurately
measures modulation and recovers the modulation signal, determines RF
frequency and measures RF power.
RF Input
Frequency Range: 150 kHz to 1300 MHz Operating Level: 12 mV rms to 7 V rms Input Impedance: 50 ohms nominal Tuning: Manual frequency entry, automatic, or track Acquisition Time: (automatic operation):1.5s
Frequency Modulation
Rates: 20 Hz to 200 kHz Deviations: To 400 kHz Accuarcy: +/- 2% of reading +/- 1 digit, 20 Hz to 10 kHz rates, 250 kHz to 10 MHz; +/- 1% of reading +/- 1 digit, 50 Hz to 100 kHz rates, 10 to 1300 MHz
Phase Modulation
Carrier Frequency: 10 to 1300 MHz Rates: 200 Hz to 20 kHz; typically usable from 20 Hz to 100 kHz with degraded performance Deviation: To 400 radians Accuracy: +/- 3% of reading +/- 1 digit
Amplitude Modulation
Rates: 20 Hz to 100 kHz Depth: To 99% Accuracy: +/- 2% of reading +/- 1 digit, 50 Hz to 10 kHz rates, 150 kHz to 10 MHz +/- 1% of reading +/- 1 digit, 50 Hz to 50 kHz rates, 10 to 1300 MHz
Frequency Counter
Range: 150 kHz to 1300 MHz Accuracy: +/-3 counts of least significant digit +/- reference accuracy
RF Level (Peak Voltage Responding, RMS Sine Wave Power Calibrated)
Range: 1 mW to 1 W Instrumentation Accuracy: +/- 1.5 dB
Power Reference
Power Output: 1.00 mW, factory set to +/- 0.7%, traceable to the U.S. National Institute of Standards and Technology Accuracy: +/- 1.2% worst case (+/-0.9% rss) for one year (0 degree to 55 degrees C)
HP 8901A Modulation Analyzer, 150 kHz to 1300 MHz Opt 001 Rear panel connections instead of front Opt 002 High stability timebase, internal Opt 003 Rear panel external LO connectors Opt 004 48-440 Hz line power Opt 010 AM/FM calibrator; accuracy: +/- 0.1% Opt 0Q8 Factory delivered service training Opt 907 Front handle kit Opt 908 Rack flange kit Opt 909 Front handle and rack flange kit Opt 910 Additional manuals Opt 915 Service manual Opt UK6 Commercial cal. certificate w/ test data Opt W30 3 Year Return Repair Service Opt W32 3 Year Calibration Customer Return Service Opt W34 3 Year Standards Compliant Calibration Service Opt W50 5 Year Return Repair Service Opt W52 5 Year Customer Return Calibration Service Opt W54 5 Year Standards Compliant Calibration Service