HP 8901B Modulation Analyzer, 150 kHz to 1.3 GHz![]()
|
RF Input
Frequency Range: 150 kHz to 1300 MHz Operating Level: 12 mV rms to 7 V rms Input Impedance: 50 ohms nominal Tuning: Manual frequency entry, automatic, or track Acquisition Time: (automatic operation):1.5s
Frequency Modulation
Rates: 20 Hz to 200 kHz Deviations: To 400 kHz Accuracy: +/- 2% of reading +/- 1 digit, 20 Hz to 10 kHz rates, 250 kHz to 10 MHz; +/- 1% of reading +/- 1 digit, 50 Hz to 100 kHz rates, 10 to 1300 MHz
Phase Modulation
Carrier Frequency: 10 to 1300 MHz Rates: 200 Hz to 20 kHz; typically usable from 20 Hz to 100 kHz with degraded performance Deviation: To 400 radians Accuracy: +/- 3% of reading +/- 1 digit
Amplitude Modulation
Rates: 20 Hz to 100 kHz Depth: To 99% Accuracy: +/- 2% of reading +/- 1 digit, 50 Hz to 10 kHz rates, 150 kHz to 10 MHz +/- 1% of reading +/- 1 digit, 50 Hz to 50 kHz rates, 10 to 1300 MHz
RF Frequency Counter
Range: 150 kHz to 1300 MHz Accuracy: +/-3 counts of least significant digit +/- reference accuracy
RF Level (True RMS)
Frequency Range with HP 11722A: 100 kHz to 2.6 GHz Power Range: -20 to +30 dBm
Selective Power Measurements(Options 030 through 037)
Frequency Range: 10 MHz to 1.3 GHz Carrier Power Range: +30 to -20 dBm, 12.5, 25 and 30 kHz filters; +30 to -10 dBm, carrier noise filter Dynamic Range: 115 dB Carrier Rejection: > 90 dB for offsets >/= 1 channel spacing or 5 kHz, whichever is larger
Power Reference
Power Output: 1.00 mW, factory set to +/-0.7%, traceable to the U.S. National Institute of Standards and Technology Accuracy: +/- 1.2% worst case (+/- 0.9% rss) for one year (0 degrees to 55 degrees C)
HP 8901B Modulation Analyzer, 150 kHz to 1.3 GHz Opt 001 RF Connectors on Rear Panel Only Opt 002 High stability Internal Reference Opt 003 Connections for External Local Oscillator Opt 004 Operation from 48 to 440 Hz Power (temp. <40 degrees C) Opt 010 AM and FM Calibrators Opt 021 Add HP 11722A Sensor Module Opt 030 High Selectivity (select only 2 filter options) (Options 032 through 037 require Option 030; Option 030 includes Option 003 connections for external local oscillators.) Opt 032 12.5 kHz Filter Opt 033 20.0/25.0 kHz Filter Opt 035 Cellular Radio Filter Opt 037 Carrier Noise Filter Opt 907 Front Handle Kit (5061-9690) Opt 908 Rack Flange Kit (5061-9678) Opt 909 Front Handle & Rack Flange Kit (5061-9684) Opt 910 2 sets of Operating (08901-90113) and Service Manuals (08901-90114) Opt 915 Service Manual (08901-90114) Opt W30 Extended Repair Service Opt W32 Calibration Service
Technical Specifications
( PDF)
Selective Power Measurements
( PDF)
If you do not have the Adobe Acrobat Reader necessary for viewing PDF documentation, download your free Acrobat Reader now. |
![]() |
Related product families available online:
Page Updated: Thursday February 25 09:44:58 UTC 1999