HP 8902A Measuring Receiver |
The HP 8902A measuring receiver combines 6 precise measurement functions
into one fully automatic, HP-IB programmable instrument. It accurately
measures RF power, tuned RF level, carrier noise/adjacent channel power,
modulation, and RF Frequency; and it characterizes audio signals. For
precise signal analysis, the HP 8902A measuring receiver provides the
performance you need.
RF Power (with HP 11722A Sensor Module)
Range: +30 dBm (1W) to -20 dBm (10 uW) Frequency Range: 0.1 MHz to 2.6 GHz Linearity: +/- 0.02 dB (within range) +/- 0.02 dB per range change from reference range +/- 1 count LSD Input SWR: <1.15
Tuned RF Level
Range 0 to -127 dBm Frequency Range: 2.5 to 1300 MHz Relative Accuracy: +/- 0.02 dB +/- 0.02 dB per IF range change +/- 0.04 dB per RF range change +/- 1 digit Worst-Case Cumulative Relative Power Accuracy(with Opt 050) +/- 0.005 dB/10 dB step (0 to -100 dBm) +/- 0.050 dB/10 dB step (-100 to -120 dBm) +/- 0.015 dB +/- 1 digit
Selective Power Measurements (Carrier Noise, Options 030 to 037)
Frequency Range: 10 to 1300 MHz Carrier Power Range: +30 dBm to -20 dBm: 12.5 kHz, 25 kHz and 30 kHz filters +30 dBm to -10 dBm: carrier noise filter Relative Measurement Accuracy: +/- 0.5 dB;levels > -95 dBc; 12.5 kHz, 25 kHz and 30 kHz filters +/- 0.5 dB;levels > -129 dBc/Hz carrier noise filter Filter Bandwidths: 2.5 kHz, carrier noise filter; 8.0 kHz, 12.5 kHz filter; 16.0 kHz, 25 kHz filter; 30.0 kHz, cellular radio filter
RF Frequency
Range: 150 kHz to 1300 MHz Maximum Resolution: 1 Hz
Amplitude Modulation
Rates: 20 Hz to 100 kHz Depths: To 99% Accuracy: +/- 1% of reading +/- 1 digit, for rates 50 Hz to 50 kHz and depths >/= 5%
Frequency Modulation:
Rates: 20 Hz to 200 kHz Deviations: To 400 kHz Accuracy: +/- 1% of reading +/- 1 digit, for rates 50 Hz to 100 kHz
Phase Modulation
Rates: 200 Hz to 20 kHz Deviations: To 400 radians Accuracy: +/- 3% of reading +/- 1 digit
HP 8902A Measuring Receiver Opt 050 Increased Power Measurement Accuracy Opt 002 1x10E -9/Day Internal Reference Oscillator Opt 021 Add HP 11722A Sensor Module Opt 001 Rear Panel Instead of Front Panel Connectors for Input, Modulation Ouput, and Calibrators Opt 003 Rear Panel External LO Connectors Opt 004 Operation from 48 to 440 Hz Power Line (temp. <40 degrees C) Opt 030 High Selectivity (select only two filter options) (Options 032 to 037 require Option 030. Opt 030 includes Option 003 connections for external local oscillator.) Opt 032 12.5 kHz Filter Opt 033 25.0 kHz Filter Opt 035 Cellular Radio Filter Opt 037 Carrier Noise Filter Opt 907 Front Handle Kit (5061-9690) Opt 908 Rack Flange Kit (5061-9678) Opt 909 Rack Flange Kit (5061-9684) with Front Handles Opt 910 Additional Operation and Calibration Manual (08902-90029) and 2 Service Manuals (08902-90031) Opt 915 Add Service Manual (08902-90031) Opt W30 Extended Repair Service Opt W32 Calibration Service