HP 8920B Option 800 AMPS/NAMPS/DAMPS/DCCH RF Communications Test Set


[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

Product Summary

Option 800 adds the HP 83206A TDMA Cellular Adapter to the HP 8920B, providing TDMA DAMPS/DCCH mobile measurement capabilities. The HP 83206A TDMA Cellular Adapter provides a complete Pi/4 DQPSK signal generator, Pi/4 DQPSK modulation analyzer, data source, and BER analyzer for TDMA measurements on DAMPS (IS-54) and DCCH (IS-136) phones. The HP 83206A replaces the HP 83201B TDMA Cellular Adapter (HP8920B Option 500) for making measurements on TDMA dual-mode DAMPS phones, adding digital control channel test features to fully characterize DCCH phones.


[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

Product Features



[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

Product Specifications

The following specifications apply when the HP 83206A is operating with an HP 8920B.

TDMA Signal Generator

Frequency Range 824 MHz to 894 MHz Output Level Range: RF In/Out: -22 dBm to -127 dBm Duplex Out: +4 dBm to -127 dBm Residual Error Vector Magnitude: <3.0% Residual Phase Error: <2.6 degrees Residual Magnitude Error: <2.6% IQ Origin Offset: <-30 dBc within 15 degrees C of last calibration Frequency Error: +/-4 Hz plus reference error

TDMA Analyzer

Frequency Range: 824 MHz to 894 MHz Input Level Range: RF In/Out: 1 mW to 60 W (0 to +47.8 dBm) Antenna In: -36 dBm to +16 dBm Input Frequency Setting Error: 1 kHz Frequency Error Accuracy: +/- 2.5 Hz plus reference accuracy Residual Error Vector Magnitude: <1.3% Error Vector Magnitude Measurement Accuracy: +/- 0.4% + 2% of reading Residual Phase Error: <1.0 degree Residual Magnitude Error: <0.9% I/Q Origin Offset Accuracy: +/-0.5 dB for values to -40 dBc



[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

Product Ordering Information

HP 8920B       RF Communications Test Set
   Opt 800     HP 83206A TDMA Cellular Adapter
   Opt W30     Three-Year Return Repair Service
   Opt W32     Three-Year Customer Return Calibration Service
   Opt W50     Five-Year Return Repair Service
   Opt W52     Five-Year Customer Return Calibration Service

HP 83206A  TDMA Cellular Adapter

Brochure
Configuration Guide
Product Note
Product Overview
Brochure
Product Note 8920-1

If you do not have the Adobe Acrobat Reader necessary for viewing this documentation, download your Free Acrobat Reader now. Get Acrobat



[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

To help you make the right instrument choices



Contact HP Test & Measurement. (c) Copyright 1994, 1995, 1996, 1997 Hewlett-Packard Company.