HP 8920D Dual-Mode Cellular Mobile Test System |
The HP 8920D dual-mode cellular mobile test system builds on the
flexibility of the HP 8920A and adds NADC-TDMA mobile test capability.
This allows you to test existing analog cellular phones and North
American dual-mode cellular phones complying with EIA/TIA IS-54/55
specifications. The HP 8920D consists of an HP 8920A with Options
103,004,013,050, and the HP 83201A dual-mode cellular adapter. With the
HP 8920D, the HP 83201A dual-mode cellular adapter is bolted on top of
the HP 8920A; all necessary interconnects are made, and the combination
is shipped as a single unit.
Signal Generator Specifications
Frequency Range: 250 kHz to 1 GHz Output Level Range: -137 to -19 dBm into 50 ohms Output Level Accuracy RF In/Out Connector: +/-1.8 dB (level less than or equal to -127 dBm) Typically +/-1.0 dB for all levels Output Level Accuracy DUPLEX Out connector: +/- 1.5 dB (-127 to +7 dBm into 50 ohms) Typically +/- 1.0 dB for all levels Harmonics: <-30 dBc (-127 to +7 dBm into 50 ohms Non-Harmonic Spurious: <-60 dBc (at >5kHz offset from carrier)
TDMA Signal Generator Specifications
Frequency Range: 824 MHz to 894 MHz Output Level Range: RF In/Out: -22 dBm to -127 dBm Duplex Out: +4 dBm to -127 dBm Residual Error Vector Magnitude: <3.0% Residual Phase Error: <2.6 degrees Residual Magnitude Error: <2.6% IQ Origin Offset: <-30 dBc within 15 degrees C of last calibration Frequency Error: +/- 4 Hz plus reference error
RF Analyzer Specifications
Frequency Measurement Range: 400 kHz to 1 GHz Level Measurement Range RF In/Out: 1 mW to 60W continuous 100W for 10 seconds/minute Level Measurement Range ANT IN: -36 dBm to +20 dBm Frequency Measurement Accuracy: +/- 1Hz plus timebase accuracy Power Measurement Accuracy: +/- 10% of reading +/- 1 mW (for inputs >/= 200mW)
TDMA Analyzer Specifications
Frequency Range: 824 MHz to 894 MHz Input Level Range: RF In/Out: 1 mW to 60 W (0 to 47.8 dBm) Antenna In: -36 to +16 dBm Input Frequency Setting Error: 1 kHz Frequency Error Accuracy: +/- 2.5 Hz plus reference accuracy Residual Error Vector Magnitude: <1.3% Error Vector Magnitude Measurement Accuracy: +/- 0.4% plus 2% of reading Residual Phase Error: <1.0 degree ResidualMagnitude Error: <0.9% I/Q Origin Offset Accuracy: +/- 0.5 dB for values to -40 dBc
AF Analyzer Specifications
Frequency Measurement Range: 20 Hz to 400 kHz Frequency Accuracy: +/-0.02% plus resolution plus timebase accuracy AC Voltage Measurement Range: 0 to 30 Vrms AC Voltage Accuracy: (20 Hz to 15kHz, inputs >/= 1 mV):+/-3% of reading DC Voltage Measurement Range: 100 mV to 42V DC Voltage Accuracy: +/-1.0% of reading plus DC offset
Oscilloscope Specifications
Frequency Range: (3 dB): 2 Hz to 50 kHz Scale/Division: 10 mV to 10V Amplitude Accuracy: (20 Hz to 10kHz):+/- 1.5% of reading +/- 0.1 div. Time/Division: 10 us to 100 ms
Spectrum Analyzer Specifications (Option 102)
Frequency Range: 400 kHz to 1 GHz Frequency Span/Resolution Bandwidth (coupled): Span Bandwidth <50 kHz 300 Hz <200 kHz 1 kHz <1.5 MHz 3 kHz <18 MHz 30 kHz >/=18 MHz 300 kHz, Plus full span capability Display: Log 1,2,and 10 dB/div Display Range: 80 dB Reference Level Range: +50 to -50 dBm
HP 8920D Dual-Mode Cellular Mobile Test Systems Opt 102 Spectrum Analyzer with Tracking Generator Opt 008 Cellular Mobile RF Power Measurement Range Opt 010 400 Hz High Pass Filter Opt 011 CCITT Weighting Filter Opt 012 4 kHz Bandpass Filter Opt 014 6 kHz Bandpass Filter Opt 019 Variable Frequency Notch Filter Opt 020 Radio Interface Card HP 83201A Dual Mode Cellular AdapterProduct Note