HP 8922B GSM Base Station Test Set, 10 MHz to 1000 MHz


[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

Product Summary

The HP 8922B provides many RF measurements and tools for testing the RF sections of GSM radios and base stations. In addition to the frequency agile 0.3 GMSK RF generator, the RF analyzer has a frequency agile local oscillator, a coherent data demodulator, a pulse demodulator, a FM demodulator, a DSP based analyzer for phase and frequency error, a synthesized spectrum analyzer, and a pulsed power meter.


[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

Product Features



[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

Product Specifications

RF Generator Specifications

Frequency
	Range:
		10 MHz to 1 GHz
	Resolution:
		1 Hz
	Accuracy:
		Reference accuracy +/- 0.5 Hz
Output
RF In/Out Connector
	Level Range:
		-19 to -127 dBm
	Level Resolution:
		0.1 dB
	Level Accuracy:
	GSM bands:
		+/-1.0 dB, levels >/= -127 dBm

RF Analyzer Specifications

Frequency
	Range:
		10 MHz to 1 GHz
	Resolution:
		1 Hz
	Hop Mode:
		Resolution:
			100 kHz
		Offset Frequency:
			< or = 50 kHz
		Offset Resolution:
			1 Hz for digital data recovery and modulation
			accuracy measurements
	RF In/Out SWR:
		< 1.5:1
CW RF Frequency Measurement
	Range
		10 MHz to 1 GHz
	Level range:
	RF In/Out:
		0 to +45 dBm
	Aux RF IN:
		-36 to +20 dBm
CW RF Power Measurements
	RF In/Out Only
	Range:
		+10 to +45 dBm
	Input Frequency Setting Error:
		+/- 500 kHz
	Accuracy:
		+ 5 dB + noise effects (0.75 mW)
Peak Transmitter Carrier Power Measurement	
	RF In/Out Only. After one timeslot, 577 us,
	from an isolated Receiver Hop Trigger in the GSM bands.
	Range:
		+10 to +45 dBm
	Input Frequency Setting Error:
		+/- 10 kHz
	Input Level Setting Error:
		+/- 3 dB
	Accuracy:	
		+/- 0.6 dB +/- noise effects (0.75 mW)



[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

Product Ordering Information

HP 8922B	GSM Base Station Test Set, 10 MHz to 1000 MHz
   Opt 001      High Stability Timebase Reference
   Opt 002      Transit Protection
   Opt 910      Extra Manuals
   Opt 913      Rack Flange Kit
   Opt 915      Adds one Assembly Level Repair Manual
   Opt W30      3 Year Return Repair Service
   Opt W32      3 Year Customer Return Calibration Service
   Opt W34      3 Year Customer Return Standards Compliant Calibration Service
   Opt W50      5 Year Return Repair Service
   Opt W52      5 Year Customer Return Calibration Service
   Opt W54      5 Year Customer Return Standards Compliant Calibration Service

Techniques for Programming the HP 892X Family of Instruments

If you do not have the Adobe Acrobat Reader necessary for viewing this documentation, download your Free Acrobat Reader now. Get Acrobat



[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

To help you make the right instrument choices



Contact HP Test & Measurement. (c) Copyright 1994, 1995, 1996, 1997 Hewlett-Packard Company.