HP 8923B DECT Test Set


[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

Product Summary

The HP 8923B DECT test set is an integrated, one-box test solution for Digital Enhanced Cordless Telecommunications. The test set provides the functionality and the measurement speed required to meet the needs of high throughput testing, essential in DECT production. In R&D applications it provides the measurement accuracy and traceability required to ensure a very high level of confidence during portable parts or fixed parts design.


[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

Product Features



[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

Product Specifications

 

DECT Source Specifications

Frequency Range: 1880 to 1990 MHz at DECT channels Frequency drift across the burst: Negligible (I/Q modulation technique used) Frequency Accuracy: +/- 5 kHz Output Range: -100 dBm to -10 dBm Output Resolution: 0.1 dB Output Accuracy: +/- 1.0 dB

Transmitter Test Specifications

Normal Transmitted Power (NTP) Measurement Range: -10 dBm to +30 dBm Accuracy: +/- 0.6 dB +/- noise effects (0.015 mW) Power Versus Time Template Measurement Measurement Range: -10 dBm to +30 dBm Dynamic Range: 40 dB GFSK Measurement: Level: -10 dBm to +30 dBm Peak Frequency Deviation Error: Input Amplitude Frequency Deviation Setting Uncertainty Measurement Uncertainty +/- 1 dB 10 kHz +/- 3 dB 12 kHz +/- 6 dB 16 kHz

Receiver Test Specifications

Residual Bit Error Ratio: 10E -6 for PRBS 29E-1 (CCITT 0.153)

Audio Source Specifications

Frequency Range: 20 Hz to 21 kHz Frequency Accuracy: +/- 0.03% of setting Level Range: 0 V to 2 V Level Accuracy: +/- (4% of setting + resolution)

Audio Analyzer Specifications:

DC Voltmeter Accuracy: (1% of reading + dc offset) DC Offset: 45 mV Audio Frequency Counter Range: 30 Hz to 400 kHz Accuracy: (0.05% + resolution + reference accuracy) for input > 30 mV AC Voltmeter Frequency Range: 50 Hz to 50 kHz Input Range: 10 mV to 5 V Digital Oscilloscope Accuracy: +/- 3% (inputs >10 mV) Frequency Range: dc to 50 kHz

Reference Specifications

Standard Frequency Reference Stability: 1 ppm (0 to 55 degrees C) Aging: 2 ppm/year Precision Frequency Reference (Option 1D5) Stability: <2.5 x 10E-3 ppm/degrees C (0 to 55 degrees C) Aging: <0.1 ppm/year for continuous operation



[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

Product Ordering Information

HP 8923B	DECT Test Set
   Opt 1D5	High Stability Frequency Reference

   Opt 0BF	Programming Reference Manual
   Opt AV4	Users Manual
   Opt 0B3	Service Manual

   Opt 1CP	Rack Mount and Handle Kit

Product Note
Brochure

If you do not have the Adobe Acrobat Reader necessary for viewing this documentation, download your Free Acrobat Reader now. Get Acrobat



[ Summary ][ Features ][ Specifications ][ Ordering ][ Assistance ]

To help you make the right instrument choices



Contact HP Test & Measurement. (c) Copyright 1994, 1995, 1996, 1997 Hewlett-Packard Company.

Page Updated: Thursday June 19 16:16:39 UTC 1997