HP 89400 Vector Signal Analyzer Family![]()
|
Complex and time-varying signals include burst, pulsed, transient, hopping, and analog- and digital-modulated signals. Signals such as these are common in RF communications, video broadcast satellite, radar, sonar and ultrasound imaging systems and must be characterized in the design, manufacture and monitoring of these systems. Vector signal analyzers are ideally suited to making the needed power, frequency, and modulation measurements on these complex signals.
The HP 89410A covers baseband frequencies from dc to 10 MHz with one or two full-bandwidth input channels. The HP 89441A covers baseband through RF frequencies of dc to 2.65 GHz with a single RF input and an (optional) dc to 10 MHz second baseband channel.
This product family includes
The following specifications cover HP 89410A and HP 89441A. See individual products for detailed specifications.Frequency Specifications
HP 89410A HP 89441A Frequency Range: dc to 10 MHz 2 MHz to 2650 MHz Frequency Span Scalar Mode: 1.0 Hz to 10 MHz 1.0 Hz to 2648 MHz Vector Mode: 1.0 Hz to 10 MHz 1.0 Hz to 7 MHz Frequency Resolution: 0.001 Hz 0.001 Hz Accuracy, Initial: ą10 ppm ą0.1 ppm Resolution Bandwidth: 312.5 ľHz to 3 MHz (1, 3, 10 sequence or arbitrary) Phase Noise Center Frequency Center Frequency at 10 MHz <= 1000 MHz <=2650 MHz ------------------------------------------------------------------------------------------------ 100 Hz offset: -106 dBc/Hz -87 dBc/Hz 1 kHz offset -110 dBc/Hz -97 dBc/Hz 10 kHz offset -120 dBc/Hz -116 dBc/Hz 100 kHz offset -120 dBc/Hz -116 dBc/Hz
Amplitude Specifications
HP 89410A HP 89441A Input Range 50 ohms: -30 dBm to +24 dBm -50 dBm to +25 dBm Accuracy Absolute Full-Scale: ą0.5 dB ą2.0 dB (ą0.5 dB typical) Input Noise Density (50 ohms): - 114 dBfs/Hz -112 dBfs/Hz Sensitivity (lowest range, 50 ohms): -144 dBm/Hz -159 dBm/Hz Spurious Responses General Spurious: <= -.75 dBfs (>=1 MHz) < -70 dBc Second Harmonic Distortion: < -75 dBc < -75 dBc Intermodulation (third order relative to two tones at -6 dBfs): <= -75 dBc <= -75 dBc Residual Responses <1 MHz: < -75 dBfs -80 dBfs >=1 MHz: -80 dBfs Analog Demodulation Maximum Bandwidth: 10 MHz (typical) 7 MHz (typical) Demodulation Accuracy AM: ą1% (typical) ą1% (typical) PM: ą3° (typical) ą3° (typical) FM: ą1% of span (typical) ą1% of span (typical) Trigger: Free run, HP-IB, internal source, external Input, Coupling: dc/ac ac Source Vector Mode: Arbitary, fixed sine, random noise, periodic chirp Arbitrary Mode: Up to 4096-complex or 8192-real points Source Level (fixed sine): -110 dBm to +23 dBm -40 dBm to +13 dBm
Measurement Characteristics
Two Channel Note: Requires second 10 MHz input channel (Option AY7) Channel Match (dc to 10 MHz): ą0.25 dB, +2.0° Real-Time Bandwidth Single-Channel Vector Mode Real-Time Bandwidth: 78.125 kHz Measurement Speed Vector Mode Maximum Display Update Rate: 60 traces/second Averaging Scalar Mode: rms (video), rms (video) exponential, peak hold Vector Mode: rms (video), rms (video) exponential, time, time exponential, peak hold Numbers of Averages: 1 to 99,999 Time-Gating Minimum Gate Length: <400 ns (dependent on span and RBW window selected) Time Capture Memory: 64 Ksample (1 Msample with Option AY9) Trace Math Operanda: Measurement data, data register, constant, other trace math functions, j(omega) Operations: +, -, x, +, cross correlation, conjugate, magnitude, phase, real, imaginary, square root, FFT, inverse FFT, natural logarithm, exponential Marker Functions: Peak signal track, frequency counter, bandpower
Vector Modulation Specifications (Option AYA)
Continuous or TDMA Formats: BPSK, QPSK, DQPSK, Pi/4 DQPSK, 8PSK, 16QAM, 32QAM, MSK, 2 and 4 level FSK Digital Video Format: 64-256QAM, 8-16VSB Data Block Lengths: Adjustable up to 4096 symbols (with Option UFG) User Selection of Filters: Raised cosine, root raised cosine, Gaussian, user-defined Adjustable Alpha (BT): 0.05 to 100 Trace and Table Formats: Constellation, vector diagram, eye diagrams, trellis diagrams, continuous I or Q vs. time, continuous error vector magnitude vs. time, modulation quality summary, detected symbol table Analysis Types: Error vector magnitude, phase error, magnitude error, amplitude droop, carrier frequency error, IQ offset, error vector spectrum, measured IQ spectrum One-Button Measurement (HP 86451A): Adjacent and alternate channel power, occupied bandwidth, modulation accuracy, power due to modulation (or full signal)
General Specifications
Environmental Calibration Interval: 1 year Warm-up Time: 30 minutes Temperature Operating: 0 to 55°C Not Operating: -20 to 65°C Power Requirements 115 Vac Operation: 90 to 140 V rms, 47 to 440 Hz HP 89440A/89441A: 47 to 63 Hz 230 Vac Operation: 198 to 264 V rms, 47 to 63 Hz Maximum Power Dissipation HP 89410A: 750 VA HP 89441A: 1025 VA Audible Noise (typical at 25°C) HP 89410A: LpA<=50 dB HP 89441A: <=55 dB Dimensions: HP 89410A Weight: 25 kg (55 lb) Size: 426 mm W x 230 mm H x 530 mm D (16.7 in x 9.1 in x 20.9 in) HP 89441A IF Section Weight: 25 kg (55 lb) Size: 426 mm W x 230 mm H x 530 mm D (16.7 in x 9.1 in x 20.9 in) HP 89441A RF Section Weight: 25 kg (55 lb) Size: 419 mm W x 173 mm H x 495 mm D (16.5 in x 6.8 in x 19.5 in)
HP 89411A 21.4 MHz Downconverter Specifications
Bandwidth With HP 8566A/B: Approx. 3 MHz With HP 70902A/70903A: Approx. 8 MHz Power Requirements 115 Vac Operation: 90 to 132 V rms, 47 to 440 Hz 230 Vac Operation: 198 to 264 V rms, 47 to 63 Hz Maximum Power Dissipation: 100 VA Dimensions Weight: 11.1 kg (25 lb) Size: 426 mm W x 90 mm H x 520 mm D (16.7 in x 3.5 in x 20.5 in)
HP 89400 Series Vector Signal Analyzers, Configuration Guide
( PDF)
HP 89400 Series W-CDMA Code-Domain Power Measurment Option for BTS Transmitter Test, Product Overview
( PDF)
If you do not have the Adobe Acrobat Reader necessary for viewing PDF documentation, download your free Acrobat Reader now. |
![]() |
HP 89410A Vector Signal Analyzer, DC to 10 MHz HP 89441A Vector Signal Analyzer, DC to 2.65 GHz HP 89441V VSB/QAM Signal Analyzer Opt AY4 Delete High Precision Frequency Reference (HP 89440/41A only) Opt AY5 Precision Frequency Reference (HP 89410A only) Opt AY7 Second 10 MHz Input Channel Opt AY8 Internal RF Source (HP 89440A/41A only) Opt AY9 Extend Time Capture to 1 Msample Opt AYA Vector Modulation Analysis Opt AYB Waterfall and Spectrogram Opt AYH Digital Video Modulation Analysis Opt AYJ Add Adaptive Equalization to Modulation Analysis Opt UFG 4-MB Extended RAM and Additional I/O Opt UG7 Advanced LAN Support Opt IC2 HP Instrument BASIC Opt ID7 50 to 75 Ohm Minimum Loss Pad (HP 89440A/41A only) Opt IFO PC-Style Keyboard - U.S. Version (other options available for German, Spanish, French, U.K., Italian, and Swedish) HP 89450A DMCS Radio Test Personality HP 89451A Radio Test Personality HP 89411A 21.4 MHz Downconverter Upgrade Kits HP 89410U HP 89410A Vector Signal Analyzer Optional Upgrade Kits HP 89441U HP 89410A Vector Signal Analyzer Optional Upgrade Kits Opt UE2 Firmware Upgrade to the Latest Revision HP 89430A 1.8 GHz RF Section (converts HP 89410A to HP 89440A) HP 89431A 2.65 GHz RF Section (converts HP 89410A/40A to HP 89441A) HP 89400A +24C User Training (scheduled) HP 89400A +24Y User Training (dedicated)
Related Information:
HP 89441A RF Section Firmware Update, Revision A.00.04 This firmware updates the HP 89441A RF Section (HP 89431A).
10 Steps to a Perfect Digital Demodulation Measurement
CDMA Measurements with the HP 89400 Series Vector Signal Analyzers
Characterization of Digital Communications Channels with the HP 89410A and 89440A
Downconverted Measurements Using the HP 89410A and HP 89441A
Dynamic Range Benefits of Large-Scale Dithered Analog-to-Digital Conversion in the HP 89410A VSAs
Frequency and Time-Selective Power Measurements With the HP 89410A and 89440A
Measuring Transmitter Transients with the HP 89400 Series Vector Signal Analyzers
Measuring Phase Noise with the HP 89400 Series Vector Signal Analyzers
Phase Noise Performance of the HP 89400 Series Vector Signal Analyzers
Time-Capture Capabilities of the HP 89400 Series Vector Signal Analyzers
Translated Frequency Response Measurements Using the HP 89400 Series Vector Signal Analyzers
Understanding Time and Frequency Domain Interactions in the HP 89400 Series Vector Signal Analyzers
Using EVM Measurements to Analyze and Troubleshoot Vector-Modulated Signals
Using Vector Modulation Analysis in ... Digital RF Communications Systems
Related product groups available online: