HP 8992A Digital Video Power Analyzer![]()
|
The HP 8992A digital video analyzer provides complete and accurate characterization of today's complex pulsed signals. It is capable of performing seven automatic timing measurements (rise time, fall time, pulse width, PRI, PRF, duty cycle, and delay) and five automatic power measurements (peak power, average power, pulse top/base amplitude, and overshoot) with push-button ease. Front panel operation is intuitive and straightforward.
The HP 8992A features two oscilloscopechannels permitting simultaneous measurements on the modulating and detected power waveforms. Dual-timebase windowing allows you to zoom in on a specific portion of the waveform and display it with more accuracy and detail.
Frequency Specifications
Frequency Range: 20 MHz to 40 GHz, sensor dependent Power Measurement Range: -33 to + 20 dBm Rise/Fall Time Video BW Tr/Tf 0 to +20 dBm: High <10 ns -16 to 0 dBm: Low <1 µs -26 to -16 dBm: CW <100 µs (Note: Rise/fall times limited to <45 ns with HP 84815 sensor) Instrumentation Uncertainty (including noise and offset) High Bandwidth: ±(0.07 + 1/(signal power in dBm +26)) dB Low, CW Bandwidth: ±(0.07 + 1.3/(signal power in dBm + 33)) dB Using HP 84815 sensor and 50 MHz, 0 dB Reference Source: ±0.07dB, -15 to +20 dBm Maximum Pulse Repetition Rate: 100 MHz externally triggered, 1 MHz internally triggered Video Inputs (Channels 2 and 3) Bandwidth DC Coupled Repetitive: dc to 100 MHz Single Shot: dc to 1 MHz AC Coupled Repetitive: 10 Hz to 100 MHz Single Shot: 10 Hz to 1 MHz Rise Time: <10 ns Vertical Sensitivity: 100 mV/div to 500 mV/div Vertical Gain Accuracy: ±1.5% Available Offset Range: ±20 Vdc, ±10 Vac Time Base Range: 2 ns/div to 5 s/div in 1-2-5 sequence Resolution: 100 ps Accuracy: 0.005%
General Characteristics
Power Requirements Voltage: 90 to 132 or 198 to 264 Vac; 48 to 66 Hz Power: 250 VA, maximum HP-IB codes: SH1, AH1, T5, L4, SR1, RL1, PP1, DC1, DT1, CO, E2 Dimensions Size: 422 mm W x 194 mm H x 366 mm D (16.6 in x 7.65 in 14.4 in) Net Weight: 12.8 kg (28 lb) Shipping Weight: 20.1 kg (44 lb)
If you have Adobe's Acrobat Reader already available on your system, please click on the icon provided below to print or view the specifications.
If you do not have the Adobe Acrobat Reader necessary for viewing PDF documentation, download your free Acrobat Reader, and then click on the icon below.
HP 8992A Digital Video Power Analyzer Opt 001 Deletes Channel Opt 002 Rear Panel Sensor Inputs Only Opt 003 Adds 1 GHz Sensor Check Source Opt 004 Adds Precision 50 MHz Reference Opt 915 Service Manual (HP 08991-90004) Opt 916 Extra User, Getting Started Calibration and Programming Guides Opt W30 Three Additional Years of Return-to-HP Warranty HP 84815A 50 MHz to 18 GHz Peak Power Sensor HP 84812A 500 MHz to 18 GHz Peak Power Sensor HP 84813A 500 MHz to 26.5 GHz Peak Power Sensor HP 84814A 500 MHz to 40 GHz Peak Power Sensor
HP 8992A Digital Video Power Analyzer, Digital Specifications
( PDF)
If you do not have the Adobe Acrobat Reader necessary for viewing PDF documentation, download your free Acrobat Reader now. |
![]() |
Related HP 8990 Peak Power Analyzer Family product datasheets:
Related product families available online:
Page Updated: Thursday February 18 09:41:40 UTC 1999