HP E1306A HP-IB Command Module for B-Size VXI

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]
The HP E1306A HP-IB Command Module is a B-size, 2 - slot VXI command module designed for use with the 20-slot HP E1302A mainframe. The HP E1306A offers an HP-IB interface for your VXI mainframe, and manages the VXI backplane communications.

The command module, which parses the measurement control commands to register-based VXI devices, allows you to send SCPI commands to instruments in a consistent, readable form, making programming easier. The HP E1306A does not support Compiled SCPI (C - SCPI). The built-in RS-232 is intended for an RS-232 terminal connection that provides debugging capabilities for the user.

You can add up to 2 MB nonvolatile RAM memory in 1 MB increments to this command module. Additional memory of 1 MB is recommended for all configurations. For some configurations (i.e., HP E1313A 32-Channel Scanning A/D), you may need to add 2 MB RAM. The additional RAM is needed for instrument reading storage and SCPI downloadable drivers for register-based modules.

Refer to the Related Links Section below for instrument driver availability and downloading instructions, as well as for recent product updates, if applicable.


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

VXI Characteristics

VXI device type: Register-based commander Size: B Slots: 2 Connectors: P1 Shared memory: n/a VXI busses: n/a C-size compatibility: Not adaptable to C-size

Instrument Drivers

See the HP Website (http://www.hp.com/go/inst_drivers) for driver availability and downloading. Command module firmware: n/a Command module firmware rev: n/a I-SCPI Windows® 3.1: n/a I-SCPI Series 700: n/a C-SCPI LynxOS: n/a C-SCPI Series 700: n/a HP Panel Drivers: Yes VXIplug&play Windows Framework: Yes VXIplug&play Windows 95/NT® Framework: Yes VXIplug&play HP-UX Framework: No

Module Current

IPM IDM +5 V: 0.95 0.01 +12 V: 0.02 0.01 –12 V: 0.02 0.01 +24 V: 0 0 –24 V: 0 0 –5.2 V: 0 0 –2 V: 0 0

Cooling/Slot

Watts/slot: 5.50 DeltaP mm H2O: 0.10 Air Flow liter/s: 0.44


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

HP E1306A    HP-IB Command Module for B-size VXI
   Opt  010  1 Mbyte Nonvolatile Memory Kit*
   Opt  AB0  Taiwan-Chinese Localization
   Opt  ABD  Germany-German Localization
   Opt  ABE  Spain-Spanish Localization
   Opt  ABF  France-French Localization
   Opt  ABJ  Japan- Japanese Localization
   Opt  ABZ  Italy-Italian Localization
   Opt  W50  5 Years of Customer Return Repair Service

*1 MB of expanded memory is required if HP E1313A is being used; however 2 MB is recommended.

Trademarks:


[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

HP E1306A Technical Specifications Datasheet

If you do not have the Adobe Acrobat Reader necessary for viewing PDF documentation, download your free Acrobat Reader now. Get Acrobat

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

To help you make the right instrument choices

[ Summary ][ Features ][ Specifications ][ Key Literature ][ Ordering ][ Assistance ][ Related ]

Related Information:


Related VXI Command Modules product datasheets:


Related VXI Direct Access Interfaces product datasheets:

Related product families available online:



Privacy Statement    Use of this site indicates that you accept the Terms of Use
Contact HP Test & Measurement. (c) Copyright 1994 - 1998 Hewlett-Packard Company.

Page Updated: Monday October 05 19:31:52 UTC 1998