HP E1347A VXI 16-Channel Thermocouple Compensation, Low-Offset Relay Multiplexer |
The HP E1347A general-purpose reed relay multiplexer is a B-size, 1-slot, register-based VXI module with thermocouple compensation. It switches 16 channels of high, low, and guard each. The module has low thermal offset performance. The multiplexer module consists of a B-size component card (labeled E1345-66201) and a screw terminal block that plugs onto the component card. The HP E1347A is functionally similar to the HP E1345A.
Using the HP E1326B or HP E1411B DMMs, the HP E1347A performs channel scanning with automatic conversions for many thermocouple types. Temperature measurements are made with automatic cold junction compensation. The card, in conjunction with HP VXI DMMs, also measure voltage, current, and 2- and 4-wire ohms.
C-size Adapter
For installing the HP E1347A in a C-size mainframe, the HP E1403B
active adapter is recommended.
Note: Configuration software that allows you to easily select
compatible interfaces, controller platforms, and VXI test system
components is available on CD ROM. Order the
"Test System
and VXI Products Catalog and CD ROM" from
the Test & Measurement Catalogs available on this web site.
Input
DC: Maximum voltage (any terminal to any other terminal or chassis): 120 Vdc AC rms: Maximum voltage (any terminal to any other terminal or chassis): 170 Vrms Maximum current (per channel common, non-inductive): 50 mA Maximum power per channel: 1 VA
DC
Maximum thermal offset per channel, differential Hi-Lo: 4 µV Closed channel resistance: 100 Ohm ±10% Insulation resistance (between any two points): 10E9 Ohm Insultation resistance (Hi to Lo, power off): n/a
AC
Minimum bandwidth (-3 dB, 50 Ohm source/ load): 10 MHz (protection resistors shorted) Crosstalk (channel-to-channel): 100 kHz: -80 dB 10 MHz: -40 dB Both: n/a Closed channel capacitance: <150 pF Hi-Lo, <150 pF Lo-Guard, <2000 pF Guard-Chassis
General Characteristics
Relays: Reed relays Break-before-make Relays open on power down Relays open on power up Minimum relay life: No load: 10E8 operations Rated load: 10E7 operations Reference junction measurement accuracy (18 to 28 °C operating): 0.3°C Strain gage excitation: n/a Screw terminal wire size: 16 to 26 AWG (1.5, 1.2, 0.9, 0.75, 0.5 mm) Scanning rate: 600 channels/s typ.
VXI Characteristics
VXI device type: Register-based, A16, slave only Size: B Slots: 1 Connectors: P1 Shared memory: None VXI busses: None C-size compatibility: Requires E1403B
Instrument Drivers
Click here for driver availability and downloading.Command module firmware: ROM Command module firmware rev: A.06 I-SCPI Win 3.1: Yes I-SCPI Series 700: Yes C-SCPI LynxOS: Yes C-SCPI Series 700: Yes HP VEE Drivers: Yes VXIplug&play Win Framework: Yes VXIplug&play Win95/NT Framework: Yes VXIplug&play HP-UX Framework: No (not available at time of publication)
Module Current
IPM IDM +5 V: 0.2 0.01 +12 V: 0.13 0.01
Cooling/Slot
Watt/slot: 1.00 deltaP mm H2O: 0.02 Air flow liter/s: 0.10
HP E1347A VXI 16-Channel Thermocouple Compensation, Low-Offset Relay Multiplexer Opt 0B3 Service manual Opt ABJ Japan - Japanese localization Opt W01 3-yr. return to HP to 1-yr. on-site warranty HP E1347-80001 Extra screw terminal blockHPE1347A Technical Specifications Datasheet
If you do not have the Adobe Acrobat Reader necessary for viewing this documentation, download your Free Acrobat Reader now. |
![]() |
Page Updated: Thursday June 19 16:15:23 UTC 1997