HP E1361A VXI 4x4 Relay Matrix |
The HP E1361A relay matrix is a B-size, 1-slot, register-based VXI
module that consists of 16 latching relays arranged as a 4x4 matrix.
The 4x4 matrix can be reconfigured as a dual 2x4 matrix by removing
factory-installed jumpers. Capable of switching voltages up to 250 Vdc
and 354 Vpk, the HP E1361A relay matrix switch has the highest voltage
rating of any HP matrix module. All relays remain in their programmed
state during power-down and are reset at power-on. After reset, all
relays are open. The HP E1361A can be reconfigured as an 8:1, 2-wire
multiplexer. This product is adaptable to C-size using the HP E1403B
adapter.
Note: Configuration software that allows you to easily select
compatible interfaces, controller platforms, and VXI test system
components is available on CD ROM. Order the
"Test System
and VXI Products Catalog and CD ROM" from
the Test & Measurement Catalogs available on this web site.
Input (HP E1361A):
Maximum voltage: 250 V DC (any terminal to any other terminal or chassis) 250 V AC rms 354 V Peak Maximum power: Per channel: 40 W, VA Per module: 320W, 960 VA
DC (HP E1361A):
Maximum thermal offset: 14 µV (per channel, differential Hi-Lo) Closed channel resistance (per channel): <1.5 ohm initial (typical) <3.5 end of life Insulation resistance (between any two points): >10E7 ohm <=40°C, <=65% RH >10E8 ohm <=25°C, <=40% RH Minimum bandwidth (-3 dB, ZL=ZX=50 ohm): 10 MHz
AC (HP E1361A):
Crosstalk (typical channel-to-channel): <100 kHz: <-80 dB <10 MHz: <-30 dB Closed channel capacitance: Hi to Lo: <150 pF* Hi to Chassis: n/a Lo to Chassis: <150 pF *All contacts closed Insertion loss: <0.1 dB @ <=100 kHz <3 dB @ <=10 MHz Minimum bandwidth (-3 dB, 50 Ohm): 10 MHz
General (HP E1361A):
Minimum relay life: 10E6 operations (no load) Screw terminal wire Size: 16 AWG (15 mm) maximum
VXI Characteristics
VXI device type: Register-based, A16, slave only Size: B Slots: 1 Connectors: P1 Shared memory: None VXI busses: None C-size compatibility: Requires E1403B
Instrument Drivers
Click here for driver availability and downloading.Command module firmware: ROM Command module firmware rev: A.01 I-SCPI Win 3.1: Yes I-SCPI Series 700: Yes C-SCPI LynxOS: Yes C-SCPI Series 700: Yes HP VEE Drivers: Yes VXIplug&play Win Framework: Yes VXIplug&play Win95/NT Framework: Yes VXIplug&play HP-UX Framework: No (not available at time of publication)
Module Current
IPM IDM +5 V: 0.1 0.01 +12 V: 0.24 0.01
Cooling/Slot
Watt/slot: 1.00 deltaP mm H2O: 0.02 Air flow liter/s: 0.10
HP E1361A VXI 4x4 Relay Matrix Opt 0B3 Service manual Opt ABJ Japan - Japanese localization Opt W01 3-yr. return to HP to 1-yr. on-site warranty HP E1361-80001 Extra terminal block for the E1361AHP E1361A Technical Specifications Datasheet
If you do not have the Adobe Acrobat Reader necessary for viewing this documentation, download your Free Acrobat Reader now. |
![]() |
Page Updated: Thursday June 19 16:15:53 UTC 1997