- Multi-function measurement capability: DC Voltage, Temperature, Resistance, Strain
- Sample/Hold, Transient Strain
- On-board averaging and hi/lo limit checking
- Optional Signal Conditioning Plug-ons (SCPs)
- Auto-sequenced scan lists for easier multirate scanning
- Engineering units conversion at full speed
- Built-in self-test *TST
- VXIplug&play-compatible with WIN / WIN95 / HP-UX drivers (See HP E2091D / E2094D product pages)
Refer to the 1996 HP VXI Source Book for more product and selection information. This catalog
is also available on CD ROM and includes VXI system design software (requires Windows 3.1 or later,
486/50MHz, 12 MB RAM, 2X CD_ROM drive). Either media can be requested from this web site
![[ Summary ]](/tmo/datasheets/Graphics/tab-summary-reg.gif) ![[ Features ]](/tmo/datasheets/Graphics/tab-features-reg.gif) ![[ Specifications ]](/tmo/datasheets/Graphics/tab-specs-high.gif) ![[ Key Literature ]](/tmo/datasheets/Graphics/tab-keylit-reg.gif) ![[ Ordering ]](/tmo/datasheets/Graphics/tab-ordering-reg.gif) ![[ Assistance ]](/tmo/datasheets/Graphics/tab-assistance-reg.gif)
Specifications include the SCP and scanning A/D
performance together as a unit. Accuracy is stated for a single
sample (averaging multiple samples will improve accuracy by reducing
noise).
Voltage Measurements
E1413B Option Number Opt. 011 Opt. 012 Opt. 013 Opt. 018/019 Opt. 020
Direct Input SCP 10 Hz Low Pass Programmable Fixed Sample and Hold
8 Channels 8 Channels Filter Input SCP Gain/Filter SCP 4 Channels
8 Channels 8 Channels
Signal Range
---------------------- ------------------ ------------------ ------------------ ------------------ ------------------
x0.062 SCP Gain n/a n/a n/a n/a n/a
x0.5 SCP Gain n/a n/a n/a n/a ±125 mV FS to ±8 V FS
x1 SCP Gain ±62.5 mV FS to ±62.5 mV FS to ±62.5 mV FS to n/a n/a
±16 V FS ±16 V FS ±16 V FS
x8 SCP Gain n/a n/a ±7.8 mV FS to n/a ±7.8 mV FS to
±2.0 V FS ±0.5 V FS
x16 SCP Gain n/a n/a n/a ±3.9 mV FS to n/a
±1 V FS (E1508A)
x64 SCP Gain n/a n/a ±3.9 mV FS to ±3.9 mV FS to ±3.9 mV FS to
±250 mV FS ±250 mV FS (E1509A) ±62.5 mV FS
x512 SCP Gain n/a n/a n/a n/a ±3.9 mV FS to
±7.8 mV FS
Accuracy
----------------------- ------------------ ------------------ ------------------ ------------------ ------------------
Gain Error % of Reading 0.01% 0.01% 0.01% 0.01% 0.02% to 0.4%
Offset Error 5.3 µV to 488 µV 7.2 µV to 488 µV 2.1 µV to 488 µV 2.3 µV to 31 µV 15 µV to 488 µV
Noise-3 Sigma 18 µV to 1.8 mV 34 µV to 1.8 mV 1.6 µV to 1.8 mV 1.7 µV to 113 µV 2 µV to 1.5 mV
Noise-3 Sigma 8 µV to 15 mV 15 µV to 1.5 mV 1.3 µV to 1.5 mV 1.4 µV to 94 µV 2 µV to 1.5 mV
(A/D Filter on)
Temperature Coefficient
Gain 10ppm/°C 10ppm/°C 15ppm/°C(after CAL) 15ppm/°C 15ppm/°C
Offset 0-30 °C no increase 0.16 µV/°C 0.16 µV/°C
30-40 °C 0.14 µV/°C 0.1 µV/°C 0.18 µV/°C+0.1 µV 0.18 µV/°C 0 to .14 µV/°C
40-55 °C .038 µV/°C+.8 µV 0.27 µV/°C+2.4 µV 0.39 µV/°C+0.6 µV 0.39 µV/°C+0.3 µV 0.14 µV/°C to 0.75 µV/°C
Normal Mode Rejection
----------------------- ------------------ ------------------ ------------------ ------------------ ------------------
Filter Frequency n/a 10 Hz 2 Hz 10 Hz 100 Hz 10 Hz 15 Hz, 100 Hz, 250 Hz
500 Hz, 1 kHz
NMR at Cutoff Frequency 0 dB 6 dB 3 dB 3 dB 3 dB 3 dB 3 dB
NMR at 50 Hz 0 dB 23 dB 44 dB 18 dB 0 dB 23 dB 6th order bessel
NMR at 60 Hz 0 dB 25 dB 45 dB 20 dB 0 dB 25 dB 6th order bessel
NMR at 400 Hz 0 dB 56 dB 80 dB 50 dB 15 dB 56 dB 6th order bessel
Input
Range: ±16 V
Gains: x1, x4, x16, x64
Maximum reading rate:
100 kSa/s divided by the number of channels in the scan
(minimum of 3 channels), subject to the bandwidth of the SCP
being used. For example:
100 K/64= 1.56 kSa/s per channel
100 K/16= 6.25 kSa/s per channel
Maximum input voltage, normal mode plus common mode:
Operating: <±16 Vpk
Damage level: >±42 Vpk
Maximum common mode voltage:
Operating: <±16 Vpk
Damage level: >±42 Vpk
Maximum tare cal offset:
62.5 mV range: ±50% of full scale
Other ranges: ±25% of full scale
SCP input impedance:
>100 MOhm (differential)
Common mode rejection:
100 to 120 dB (DC-60 Hz, depends on SCP in use)
Measurement resolution:
16 bits (including sign)
Measurement accuracy:
90 days, 23 ±1 ºC (with CAL done after 1 hr warm-up and CALZERO done within 5 min)
Filters:
Filters are 2-pole and are buffered
Temperature Measurements
Any SCP can be used for temperature measurements with thermocouples, thermistors,
and RTDs, but the HP E1503A, E1508A, and E1509A provide better accuracy.
Temperature accuracy:
All specifications for the following were measured with the A/D filter off.
Reference
One of the 64 channels can be used to measure the reference, or a known
reference temperature can be downloaded to the A/D converter.
Thermistor and RTDs
Require an input SCP and the HP E1505A Current Source SCP for excitation.
Engineering units:
Conversion to decrees C are made on card at full speed.
Transducer types:
Thermocouples: E, E ext, J, K, N, R, S, T (linearized to ITS90 per NIST Monograph 175)
Thermistors: 2252, 5 kOhm, and 10 kOhm
RTDs: 100 Ohm type 85 and 92
Measurement Range:
-200 to +1500 ºC (dependent on transducer type)
Accuracy:
0.003 ºC to 3 ºC total error (not including transducer type)
![[ Summary ]](/tmo/datasheets/Graphics/tab-summary-reg.gif) ![[ Features ]](/tmo/datasheets/Graphics/tab-features-reg.gif) ![[ Specifications ]](/tmo/datasheets/Graphics/tab-specs-reg.gif) ![[ Key Literature ]](/tmo/datasheets/Graphics/tab-keylit-reg.gif) ![[ Ordering ]](/tmo/datasheets/Graphics/tab-ordering-high.gif) ![[ Assistance ]](/tmo/datasheets/Graphics/tab-assistance-reg.gif)
Order the HP E1413B only to add to an existing system that uses other
HP E1413Bs. Otherwise, order the HP E1413C instead. Note: all SCPs
are 8 channels unless otherwise noted.
HP E1413B 64-Channel Scanning A/D Converter
SCP options:
Opt 011 Direct input
Opt 012 Low-pass filter
Opt 013 Gain/filter
Opt 014 Breadboard
Opt 015 Current source
Opt 016 120 ohm strain gage
Opt 017 350 ohm strain gage
Opt 018 Fixed gain/filter
Opt 019 Fixed gain/filter (64x)
Opt 020 Sample and hold (4 channels)
Opt 021 Transient strain (4 channels, double wide SCP)
Other options:
Opt 0B3 Service manual
Opt 0R3 3 yr. uptime support loaner
Opt ABD Germany - German localization
Opt W01 3-yr. return to HP to 1-yr. on-site warranty
HP E1413-80001 Extra Terminal Block
HP E1413G E1413A upgrade kit
![[ Summary ]](/tmo/datasheets/Graphics/tab-summary-reg.gif) ![[ Features ]](/tmo/datasheets/Graphics/tab-features-reg.gif) ![[ Specifications ]](/tmo/datasheets/Graphics/tab-specs-reg.gif) ![[ Key Literature ]](/tmo/datasheets/Graphics/tab-keylit-high.gif) ![[ Ordering ]](/tmo/datasheets/Graphics/tab-ordering-reg.gif) ![[ Assistance ]](/tmo/datasheets/Graphics/tab-assistance-reg.gif)
No literature is available online at this time.
![[ Summary ]](/tmo/datasheets/Graphics/tab-summary-reg.gif) ![[ Features ]](/tmo/datasheets/Graphics/tab-features-reg.gif) ![[ Specifications ]](/tmo/datasheets/Graphics/tab-specs-reg.gif) ![[ Key Literature ]](/tmo/datasheets/Graphics/tab-keylit-reg.gif) ![[ Ordering ]](/tmo/datasheets/Graphics/tab-ordering-reg.gif) ![[ Assistance ]](/tmo/datasheets/Graphics/tab-assistance-high.gif)
To help you make the right product choices
Your Field Sales Engineer is ready to answer your questions about
specifications, applications and ordering. Find your Field Sales Engineer at
your
Local Sales Office.
![[ Summary ]](/tmo/datasheets/Graphics/tab-summary-reg.gif) ![[ Features ]](/tmo/datasheets/Graphics/tab-features-reg.gif) ![[ Specifications ]](/tmo/datasheets/Graphics/tab-specs-reg.gif) ![[ Key Literature ]](/tmo/datasheets/Graphics/tab-keylit-reg.gif) ![[ Ordering ]](/tmo/datasheets/Graphics/tab-ordering-reg.gif) ![[ Assistance ]](/tmo/datasheets/Graphics/tab-assistance-reg.gif)