HP E1419A Multifunction Plus Measurement and Control![]()
|
The flexibility in configuring with multiple SCPs allows for multiple test setups of mixed signals, both input and output, without adding extra VXI measurement modules. The integrated signal conditioning provides for more accurate and repeatable calibration and eliminates the need for separate signal conditioning carriers. The intelligent measurement and control allows for scaleable configurations, on-board Engineering Unit (EU) conversion, and real-time decision making.
HP VEE graphical programming provides easy-to-use supervisory and analytical control. Example templates for checking channel setup, calibration, loading custom EU conversions, and others are included. Drivers are available for use with National Instruments' LabVIEW.
Refer to the Related Links Section below for instrument driver availability and downloading instructions, as well as for recent product updates, if applicable.
VXI Characteristics
VXI device type: A16, slave only, register based Data transfer bus: n/a Size: C Slots: 1 Connectors: P1/2 Shared memory: n/a VXI busses: TTL Trigger bus (T) C-size compatibility: n/a
Instrument Drivers
See the HP Website (http://www.hp.com/go/inst_drivers) for driver availability and downloading. Command module firmware: Downloadable Command module firmware rev: A.08 I-SCPI Windows 3.1: No I-SCPI Series 700: Yes C-SCPI LynxOS: Yes C-SCPI Series 700: Yes HP Panel Drivers: No VXIplug&play Windows Framework: No VXIplug&play Windows 95/NT Framework: Yes VXIplug&play HP-UX Framework: No * The HP VEE application can use VXIplug&play drivers or HP panel drivers.
Module Current (with no SCPs installed)
IPM IDM +5 V: 1.0 0.02 +12 V: 0.06 0.01 –12 V: 0 0 +24 V: 0.1 0.01 –24 V: 0.1 0.01 –5.2 V: 0.15 0.01 –2 V: 0 0
Cooling/Slot
Watts/slot: 14.00 DeltaP mm H2O: 0.08 Air flow liter/s: 0.8
Family Specifications
Model HP E1419A Multifunction Plus Measurement and Control VXI Characteristics: Size: C Slots: 1 VXI Device Type: Register based Instrument Drivers: See the HP Website (http://www.hp.com/go/inst_drivers) for driver availability and downloading. VXIplug&play Windows® Framework: No VXIplug&play Windows95/NT® Framework: Yes VXIplug&play HP-UX Framework: No Specifications Measurement resolution: 16 bit Memory: 64 kSa Number of channels: 64 (depends on configuration) Maximum reading rate: 56 kSamples/s Measurement accuracy: Typically ±0.01% of input level, varies with the SCP used
E1419A Multifunction Plus Measurement and Control Opt 001 Delete 4 Direct Input SCP's Opt 011 Screw Terminal Block E1419A Opt 013 Spring-Clamp Terminal Block Opt A3F Interface to Rack Mount Panel Opt W01 Convert 3 year Return to 1 year OnSite E1501A 8-Channel Direct Input SCP E1502A 8-Channel 7 Hz Low-pass Filter SCP E1503A 8-Channel Programmable Filter and Gain SCP E1504A Breadboard SCP E1505A 8-Channel Current Source SCP E1506A 8-Channel 120 Ohm Strain Completion & Excitation SCP E1507A 8-Channel 350 Ohm Strain Completion & Excitation SCP E1508A 8-Channel x16 Gain & 7 Hz fixed Filter SCP E1509A 8-Channel x64 Gain & 7 Hz fixed Filter SCP E1510A 4-Channel Sample & Hold Input SCP E1511A 4-Channel Transient Strain SCP E1512A 8-Channel 25 Hz Fixed Filter SCP E1513A 8-Channel Divide-by-16 Fixed Attenuator & 7 Hz Low-pass Filter SCP E1514A 4-Channel Isolated x1 Gain & 10 Hz Fixed Filter SCP E1515A 4-Channel Isolated x1 Gain & 100 Hz Fixed Filter SCP E1516A 4-Channel Isolated x64 Gain & 10 Hz Fixed Filter SCP E1517A 4-Channel Isolated x64 Gain & 100 Hz Fixed Filter SCP E1518A 4-Wire Resistance Measurement SCP E1531A 8-Channel Voltage Output SCP E1532A 8-Channel Current Output SCP E1533A 16-Bit Digital I/O SCP E1534A 8-Bit Frequency/Totalize/PWM SCP E1535A Watchdog Timer SCP E1536A 8-Bit Isolated Digital I/O SCP E1537A 4-Channel Voltage Output SCP E1538A Enhanced Frequency/Totalize/PWM SCP **Note: The HP E1419A ships with four HP E1501A 8-Channel Direct Input SCPs mounted in the first four (of eight) SCP positions. Only those SCPs marked (**) will operate in the first four SCP positions and any of those may be substituted for the HP E1501A using the delete option 001 and purchasing them separately. The HP E1419A ships with no SCPs specified for the final four SCP positions. Any of the SCPs listed in the ordering information may be used in these four SCP positions. 1Note: No terminal block is included with the HP E1419A. You must specify a terminal block option when ordering.
HP E1419A Multifunction Plus Measurement and Control Technical Specifications
( PDF)
If you do not have the Adobe Acrobat Reader necessary for viewing PDF documentation, download your free Acrobat Reader now. |
![]() |
Related Information:
Analog Signal Conditioning Plug-on (SCP) Table measurement selection table designed to work with the following VXI cards: HP E1313A, E1413C, E1415A, and E1419A
Digital Signal Conditioning Plug-on (SCP) Table measurement selection table designed to work with the following VXI cards: HP E1415A and E1419A
Related VXI Scanning A/D Converters product datasheets: