HP E1430A VXI 10 MSa/s 23 bit Digitizer with DSP and Local Bus![]()
|
The digitizer used in this module is a two-pass (or sub-ranging) type that provides 23 bits of raw resolution. The digitizer in the E1430A has low distortion with a noise density of -136 dBfs/Hz and a noise figure of 15 dB.
The module comes with multistage digital filters built in with a maximum information bandwidth of 4 MHz. A high-speed 8 MB memory captures transients or acts as a FIFO for continuous digitizing during the block-mode data transfers used by some DSP algorithms. Data transfer is over the VXI data path and the high speed Local Bus (40 Mbytes/sec).
Note: Configuration software that allows you to easily select compatible interfaces, controller platforms, and VXI test system components is available on CD ROM. Order the "Test System and VXI Products Catalog and CD ROM" from the Test & Measurement Catalogs available at this web site.
Analog Input
Ranges (in 6 dB steps): +28 dBm to -32 dBm Input impedance: 50 ohms, >40 dB return loss, 4 MHz, DC coupled.
Accuracy
Resolution: 23 bits Absolute voltage measurements: ±0.03 dB (DC coupled, f<100 kHz, decimation filters off, ±1V range, 25°C, analog anti-alias filter on) Range accuracy: ±0.03 dB (relative to ±1 V range for all ranges, f <100 kHz) Temperature drift: <0.001 dB per °C of deviation from 25°C
Dynamic Range
Harmonic distortion: <-110 dBfs or -80 dBc, whichever is greater (with additional signal applied >-20 dBfs) Spurious signals: <-110 dBfs, using internal clock (includes alias products)
Clock
ADC clock sources: internal, external BNC input compatible with TTL, ECL, and >-6 dBm sinewaves ADC clock accuracy: ±70 Hz (Internal 10 MHz; 0°C to 40°C)
Trigger
Sources: external TTL with ±slope select, ADC output, signal magnitude, software initiated, immediate Trigger delay: -1,048,575 to 8,388,607 samples
Filtering
Anti-alias filter: 4 MHz Filtering and decimation: 25 baseband spans, octave steps Zoom: digital LO tunes the center frequency of any filter span with better than 10 µHz resolution
Programming
Output data formats: 16-bit and 32-bit real, 16-bit and 32-bit complex Data output modes: VME data transfers, high-speed Local Bus Measurement modes: block and continuous Read registers: mfr./model, logical address, status bits, measurement loop status Data interrupt: two independent priority interrupts initiated by masked status bits Memory: 8 MB (4 MSa, 16 bits), FIFO
Backplane Connector Shielding
To ensure compliance with RFI levelsspecified in standards EN55001 and CISPR11, this product requires the backplane connector shields installed in an HP VXI C-size mainframe. Option 918 is available with the purchase of a new mainframe; accessories, HP P/Ns E1400-80920 and E1421-80920 (one kit per mainframe), are available for retrofitting existing HP mainframes HP E1401A/B and HP E1421A, respectively.
HP E1430A VXI 10 MSa/s 23 bit Digitizer with DSP and Local Bus Opt 0B0 Delete manual set Opt 0B1 Add manual set Opt AYD 10.24 MHz time base Opt W01 3-yr. return to HP to 1-yr. on-site warranty E1400-80920 Shield, backplane connector E1421-80920 Shield, backplane connector kit
HP E1430A 10 MSample/sec ADC, with Filtering and Memory
Precision Time-Domain Measurements using the HP E1430A
A 10-Megasample-per Second Analog-to-Digital Convertr with Filter and Memory
If you do not have the Adobe Acrobat Reader necessary for viewing PDF documentation, download your free Acrobat Reader now. |
![]() |
Related HP E3238S Signals Development System product datasheets:
Related product families available online: