HP E1440A VXI 21 MHz Synthesized Function/Sweep Generator![]()
|
With this module, you can use the modulation source as an arbitrary function generator via HP-IB to provide user-defined waveforms, or use the save-recall memory that includes nonvolative memory locations for simple and rapid access to frequently used test setups. Additionally, you can produce five different waveforms including: sine, square, triangle, negative, and positive ramps.
Note: Configuration software that allows you to easily select compatible interfaces, controller platforms, and VXI test system components is available on CD ROM. Order the "Test System and VXI Products Catalog and CD ROM" from the Test & Measurement Catalogs available on this web site.
Those specifications indicated as typical describe the instruments typical performance; all others describe the instruments warranted performance.
Waveforms
sine, square, triangle, negative and positive Ramps, dc, TTL clock
Frequency Range
Sine: 1 µHz - 21 MHz Square: 1 µHz - 11 MHz Triangle: 1 µHz - 11 kHz Ramps: 1 µHz - 11 kHz TTL clock: 1 µHz - 60 MHz Resolution: 11 digits Accuracy: ±5 ppm of selected value, 20 to 30°C, at time of calibration with standard frequency reference Stability: ±5 ppm/year, 20 to 30 °C Modulation: AM, PM
Main Signal Output (Typical)
Impedance: 50 Ohm ±1 Ohm Maximum voltage: 42 V pk (AC+DC) (chassis ground to circuit ground, 0 - 10 kHz) Maximum voltage: ±10 V (floating ground to signal output) Connector: BNC
Amplitude
(50 ohm all waveforms except TTL clock, without DC offset) Range: 1 mV to 10 Vp-p in 8 amplitude ranges, 1-3-10 sequence, amplitude can also be set up in Vrms and dBm Resolution: 4 digits (0.03% of full range) Amplitude Accuracy: ±2% FS
Sinewave Spectral Purity
Phase noise: -55 dB for a 30 kHz band centered on a 20 MHz carrier (excluding ±1 Hz about the carrier) Spurious: -60 dBc or -85 dBm, whichever is greater Sinewave harmonic distortion: Frequency Range Harmonic Level 0.1 Hz - 199 kHz -60 dBc 200 kHz - 1.99 MHz -40 dBc 2 MHz - 14.9 MHz -30 dBc 15 MHz - 20 MHz -25 dBc
Squarewave Characteristics (Typical)
Rise/fall time: <=20 ns (10% to 90% of p-p output voltage) Overshoot: 5% of p-p amplitude at full output Triangle/ramp linearity: ±0.05% of full p-p output voltage for each range (10% - 90%, 10 kHz)
DC Offset Range (50 ohm)
DC only: 0 to ±5 V DC + AC: max. ±4.5 V Resolution: 4 digits
Phase Offset
(related to other HP E1440A or equivalent) Range: ±719.9° Resolution: 0.1° Increment accuracy: ±0.5° Stability: ±1° of phase / °C
Sinewave Amplitude Modulation (Typical)
Modulation depth: 0 - 98% Modulation frequency range: DC to 350 kHz (1 µHz - 21 MHz carrier frequency) Envelope distortion: -30 dB for modulation to 80% at 1 kHz (0 Vdc Offset) Sensitivity: ±5 V peak for maximum modulation
Phase Modulation (Typical)
Sinewave range: ±900°, ±5 V input Sinewave linearity: ±0.5%, best fit straight line up to ±720° of modulation range Squarewave range: ±450° Triangle range: ±45° Positive and negative ramps range: ±90° Modulation frequency range: DC to 5 kHz
Frequency Sweep
Sweep sequence modes: single, continuous Sweep function modes: Multi-Interval: (up to 50 different intervals can be sequenced and repeated in any order in a sequence that can contain up to 100 intervals) Linear or logarithmic sweep: (can be set for each interval) Sweep time: Linear: 0.01 s to 105 s Logarithmic: 0.1 s to 105 s Minimum sweep width: Linear: 0 Hz Logarithmic: 1 decade Maximum sweep width: full frequency range Minimum sweep rate: Linear: 0.2 Hz/s Phase continuity: sweep is phase continuous over the full frequency range of the main output for all sweep modes Multimarker: Linear sweep only: up to nine markers can be set in this one dedicated interval Sweep time: 0.01 s to 105 s Sweep width: from 0 Hz to full frequency range
Auxiliary Outputs (Typical)
SYNC-OUT TTL: Signal: Phase synchronous squarewave with same frequency as the main signal output, or 1 µHz to 60 MHz TTL clock (main signal output switched off) Output impedance: 50 ohm Connector: BNC and TTL trigger bus X-Drive 0 to 10 V: Signal: 0 - 100 s sweeps only (proportional ramp to the entire sweep time) Output impedance: 650 ohm Output level: 0 to + 10 V (into open circuit) Connector: BNC Pen lift: Signal: TTL-compatible voltage levels capable of sinking current from a positive source. Current 200 mA, voltage 45 V Connector: BNC Marker TTL: Signal: High-to-low transitions at selected marker frequencies. TTL- and CMOS-compatible output levels Pulsewidth in multimarker mode: 1 ms Connector: BNC and TTL trigger bus Fan out: 4 REF out 10 MHz Signal: 10 MHz squarewave for phase-locking additional instruments to the HP E1440A. Output impedance: 50 ohm Output levels (into 50 ohm): high level >2 V, low level <0.2 V AC-coupled output levels: 10 dBm Connector: BNC
Auxiliary Inputs (Typical)
External REF in 1/10 MHz: (For phase-locking the HP E1440A to an external frequency reference) Signal: from 0 dBm to 20 dBm into 50 ohm (reference signal must be a subharmonic of 10 MHz from 1 MHz to 10 MHz) Connector: BNC or VXI-system clock AM: Input impedance: 10 kOhm Connector: BNC Max. external voltage: ±15 V PM: Input impedance: >40 kOhm Connector: BNC Max. external voltage: ±15 V
Option 001 High-Voltage Output
Frequency range: 1 µHz to 1 MHz Amplitude: Range: 4 mV to 40 V p-p in eight ranges, 4-12-40 sequence into 500 ohm, <500 pF load; ranges are four times the standard instrument ranges, without DC offset Accuracy: ±2% of full output for each range at 2 kHz Flatness: ±10% relative to programmed amplitude Sinewave harmonic distortion: Frequency Range Harmonic Level 10 Hz - 199 KHz - 60 dBc 200 kHz - 1 MHz - 40 dBc Output impedance: <3 ohm at DC, <10 ohm at 1 MHz (load impedance 500 ohm, 500 pF, max. output current 40 mA peak) DC offset range: four times the specified range of the standard instrument
VXI Characteristics
VXI device type: Message-based Data transfer bus: A16/24, D16 Master, A16/D16 Slave Size: C Slots: 1 Connectors: P1/2 Shared memory: n/a VXI busses: TTL Trigger Bus (T) C-size compatibility: Yes
Instrument Drivers
Click here for driver availability and downloading.Command module firmware: n/a Command module firmware rev: n/a I-SCPI Win 3.1: n/a I-SCPI Series 700: n/a C-SCPI LynxOS: n/a C-SCPI Series 700: n/a HP VEE Drivers: n/a VXIplug&play Win Framework: Yes VXIplug&play Win95/NT Framework: Yes VXIplug&play HP-UX Framework: No (not available at time of publication)