HP E1442A VXI 64-Channel General Purpose Switch | |
The HP E1442A 64-Channel General-Purpose Switch is a C-size, 1- slot, register-based VXI module that contains nonlatching Form C type relays. They can be used for device actuation, as digital output modules, or to create flexible signal switching topologies.
Each product has two parts—a component card containing the relays, that plugs into the mainframe, and a removable terminal block that plugs onto the component card. The component card has 64 Form C relays with three independent axial component locations per relay. These locations can be used for jumpers, fuses, or pullup/down and protection resistors. All three relay terminals can be connected through such components to one of three internal buses.
The HP E1442A terminal block comes standard with solder lugs that
allow full access to the Form C (SPDT) capability of the component
module relays. Option 020 has a screw terminal block that allows
access to the Form A (SPST) contacts only. Option 010 gives you
complete Form C connectivity with more customization versatility.
Note: Configuration software that allows you to easily select
compatible interfaces, controller platforms, and VXI test system
components is available on CD ROM. Order the
"Test System
and VXI Products Catalog and CD ROM" from
the Test & Measurement Catalogs available on this web site.
Form: C (SPDT) A (SPST)
Number of Channels: 64
Input (HP E1442A):
Maximum voltage: 150 Vdc, 150 Vrms, 210 Vac peak High to Low, or any terminal to chassis (this is derated to 42 Vdc or 60 Vac peak if the internal buses are used for any channel or Opt 010 is used). Maximum current (noninductive, per switch): 1 A dc or ac rms Maximum power: 40 VA ac or 40 W dc per switch 320 VA or W per module
DC (HP E1442A):
Maximum thermal offset: 70 µV per channel Closed channel resistance (typical): <1.5 Ohm (inital) <3.5 Ohm (end of life) Insulation resistance (between any two points): <=40° C, <=65% RH: >10E7 Ohm <=25° C, <=40% RH: >10E8 Ohm
AC (HP E1442A):
Typical bandwidth (-3 dB): 10 MHz Crosstalk (dB, channel-to-channel): <100 kHz: <-70 <10 MHz: <-28 Closed channel capacitance: <40 pF channel-to-common <30 pF channel-to-channel
General (HP E1442A)
Typical relay life: >5x10^6 no load operations >10^5 at rated load Time to Open/Close a Channel (using register programming): 13 ms Connection types: Standard HP E1442A: Solder lug connectors HP E1442A with Opt 010: (replaces standard terminal block) Provision to add dual 96-pin DIN-C connectors or solder wires directly to terminal block pc board, 0.055 inch holes, 20 AWG maximum recommended. HP E1442A with Opt 020: (replaces standard terminal block) Screw terminals for Com and NO contacts only. 16 AWG (1.5mm) maximum.
VXI Characteristics
VXI device type: Register-based,A16, slave only Size: C Slots: 1 Connectors: P1/P2 Shared memory: None VXI busses: None C-size compatibility: n/aInstrument Drivers
Click here for driver availability and downloading.Command module firmware: Downloadable Command module firmware rev: A.06 I-SCPI Win 3.1: Yes I-SCPI Series 700: Yes C-SCPI LynxOS: Yes C-SCPI Series 700: Yes HP VEE Drivers: Yes VXIplug&play Win Framework: Yes VXIplug&play Win95/NT Framework: Yes VXIplug&play HP-UX Framework: No (not available at time of publication)
Module Current
IPM IDM +5 V: 0.1 0.11
Cooling/Slot
Watt/slot:5.00 deltaP mm H2O: 0.15 Air flow liter/s: 0.42
HP E1442A VXI 64-Channel General Purpose Switch Opt 010 Pull up/pull down resistors Opt 020 Form A contacts (64) instead of Form C Opt 0B3 Service manual Opt W01 3-yr. return to HP to 1-yr. on-site warranty HP E1442-80011 Extra terminal block (Opt 020) HP E1442-80014 Extra terminal block (E1442A) HP E1442-80010 Extra terminal block (Opt 010)HP E1442A Technical Specifications Datasheet
| If you do not have the Adobe Acrobat Reader necessary for viewing this documentation, download your Free Acrobat Reader now. |
|

Page Updated: Thursday June 19 16:15:42 UTC 1997