HP E1473A VXI Six 1x4, 50 Ohm RF Multiplexer Expander |
The HP E1473A 50 Ohm RF Multiplexer Expander is a C-size, 1- slot, register-based VXI module. It is the ideal choice to inexpensively increase system switching capacity to switch test signals from your application to your test instruments such as oscilloscope and spectrum, network, or distortion analyzers. The HP E1473A expander is connected to and requires either an HP E1472A or HP E1474A RF multiplexer in the same system. Up to two HP E1473A multiplexer expanders can be connected to one multiplexer, providing a total of eighteen 1x4 multiplexer banks. You can mix and match the HP E1473A 50 ohm and HP E1475A 75 ohm expanders with the HP E1472A 50 ohm and HP E1474A 75 ohm multiplexers.
Configuration
The HP E1473A can be installed in a VXI C-size slot, adjacent to the
HP E1472A (or HP E1474A), or rack mounted externally up to eight
meters from the VXI mainframe using remote extender cables. The
adjacent slot can be used by another module, making the multiplexer
expander a slotless device. Remote installation of the HP E1473A
expander close to the device under test allows the test cable length
to be kept as short as possible, reducing signal delay and insertion
loss.
The switching sections of the HP E1473A 50 Ohm RF multiplexer
expander and the HP E1472A 50 Ohm RF multiplexer are identical. Only
one channel in each bank can be connected to its common at any time.
The multiplexer relays are arranged in a tree-switched
configuration, providing high isolation and low VSWR. Each channel
consists of a nonlatching, armature relay. At power-on or reset,
channels 00, 10,...50 areconnected to COM 00, 10,...50, respectively,
and all other channels are open (unterminated).
Note: Configuration software that allows you to easily select
compatible interfaces, controller platforms, and VXI test system
components is available on CD ROM. Order the
"Test System
and VXI Products Catalog and CD ROM" from
the Test & Measurement Catalogs available on this web site.
Input (HP E1473A):
Maximum Voltage: 42 V (center or shield-to-center, shield or chassis) Maximum Current (per channel or common): DC: 1 A AC rms: 1 A Maximum Power (per channel common): DC: 24 W AC: 24 VA
DC (HP E1473A):
Maximum Thermal Offset: 6 µV Closed Channel Resistance (typical): <1 ohm Insulation Resistance (between any two terminals): >108 ohm <=40°C, <=95% RH
AC (HP E1473A):
Characteristic Impedance (ZL=ZS=ZO): 50 ohm Insertion Loss: <10 MHz: <0.1 dB <100 MHz: <0.4 dB <500 MHz: <0.9 dB <1.3 GHz: <1.5 dB <3 GHz (Typ): <8.0 dB Crosstalk (channel-to-channel): <10 MHz: <-90 dB <100 MHz: <-80 dB Crosstalk (one channel closed or channel-to-common): <500 MHz: <-62 dB <1.3 GHz: <-50 dB VSWR: <10 MHz: <1.05 <100 MHz: <1.15 <500 MHz: <1.35 <1.3 GHz: <1.5 Risetime: <300 ps Signal Delay: <3 ns
General (HP E1473A):
Typical Relay Life: No load 5x106 operations At rated load 105 operations Relays: Nonlatching, armature
VXI Characteristics
VXI device type: Register-based, A16, slave only Size: C Slots: 1 Connectors: P1 Shared memory: None VXI busses: None C-size compatibility: n/a
Instrument Drivers
Click here for driver availability and downloading.Command module firmware: ROM Command module firmwarerev: A.02 I-SCPI Win 3.1: Yes I-SCPI Series 700: Yes C-SCPI LynxOS: Yes C-SCPI Series 700: Yes HP VEE Drivers: Yes VXIplug&play Win Framework: Yes VXIplug&play Win95/NT Framework: Yes VXIplug&play HP-UX Framework: No (not available at time of publication)
Module Current
IPM IDM +12 V: 0.36 0
Cooling/Slot
Watt/slot: 1.00 deltaP mm H2O: 0.02 Air Flow liter/s: 0.10
HP E1473A VXI Six 1x4, 50 Ohm RF Multiplexer Expander Opt 0B3 Service manual Opt W01 3-yr. return to HP to 1-yr. on-site warranty HP E1473-80002 Cable extension kitCable Extension Kit
If the HP E1473A expander is physically located away from the mainframe, you need to order the cable extension kit. Each kit includes two 0.8 meter cables, each controlling three of the six expander banks. Up to 10 extender cables can be daisy-chained. To remotely install one HP E1473A expander at a distance of eight meters, controlling all six multiplexer banks, requires ten E1473-80002 kits.
Cables and Connectors
Various 50 Ohm cables are available from HP for connecting to the SNB connectors on the front panel of the multiplexer. Adaptors and other connectors are also available. Connectors are also available from Johnson Components:
If you do not have the Adobe Acrobat Reader necessary for viewing this documentation, download your Free Acrobat Reader now. |
![]() |
Page Updated: Thursday June 19 16:15:35 UTC 1997