HP E1563A 800 kSa/s, 2-Channel Digitizer![]()
|
Each channel of the HP E1563A digitizer has its own analog electronics, including switchable 25 kHz input filters, 800 kSa/s 14 bit A/D converters, and independent channel isolation.
The HP E1563A is configured for PC SIMM memory, scaleable from 4 Mbyte to 64 Mbytes. The large memory can easily capture transients and act as a FIFO to allow continuous digitizing while unloading data with block mode transfers.
Refer to the Related Links Section below for instrument driver availability and downloading instructions, as well as for recent product updates, if applicable.
Time Based Triggering
Both channels sample simultaneously with a single internal or external time base. Triggering can be set up for either time or event modes with programmable pre- and post-trigger reading counts.
Programming
The E1563A has a simple programming model which includes a data FIFO or flat file memory model for the A/D converter, 16 bit integer data corrected for offset and gain errors, and a current value table to retrieve the current sample of data.
Accuracy
Range Zero Offset1(with filter OFF) Zero Offset1(with filter ON) Specification2 Temperature Specification2 Temperature Coefficient3 Coefficient3 0.0625 V 20 µV 1.9 µV/°C 28 µV 4.3 µV/°C 0.25 V 78 µV 6 µV/°C 110 µV 16 µV/°C 1 V 300 µV 15 µV/°C 430 µV 63 µV/°C 4 V 1.2 mV 60 µV/°C 1.7 mV 251 µV/°C 16 V 21 mV 1.3 mV/°C 21 mV 1.63 mV/°C 64 V 28 mV 1.65 mV/°C 34 mV 4.24 mV/°C 256 V 79 mV 4.28 mV/°C 110 mV 16.2 mV/°C (continued) Range Gain(% of reading) Noise(3 sigma) Specification2 Temperature Specification2 0.0625 V 0.034% 0.0061%/°C 57 µV 0.25 V 0.034% 0.0061%/°C 180 µV 1 V 0.034% 0.0061%/°C 720 µV 4 V 0.034% 0.0061%/°C 2.88 mV 16 V 0.034% 0.0061%/°C 14.7 mV 64 V 0.034% 0.0061%/°C 48 mV 256 V 0.034% 0.0061%/°C 189 mV 1 Valid within the range of 0 °C to 55 °C. A zero offset calibration for all channels must be performed if the instrument experiences a temperature <0 °C or >55 °C for these specifications to remain valid.
2 Specification is valid when tested at a temperature within ± 5 °C of the calibration temperature.
3 Amount of error that must be added for each °C outside of ± 5 °C of the calibration temperature.
Integral Non-Linearity Specification
All ranges: 2.5 LSB
Environmental Specifications
For indoor use: Pollution degree 2 Operating altitude: 3000 meters or mainframe altitude specification, whichever is lower Operating temperature: 0 °C to 55 °C Relative humidity: Up to 80% at 31 °C, decreasing to 50% at 40 °C
VXI Characteristics
VXI device type: Register based Data transfer bus: A16, slave only Size: C Slots: 1 Connectors: P1/2 Shared memory: None VXI busses: TTL
Instrument Drivers
See the HP Website(http://www.hp.com/go/inst_drivers) for driver availability and downloading. Command module firmware: Downloadable Command module firmware rev: A.01.00 I-SCPI Windows 3.1: No I-SCPI Series 700: No C-SCPI LynxOS: No C-SCPI Series 700: No HP Panel Drivers: No VXIplug&play Windows Framework: No VXIplug&play Windows 95/NT Framework: Yes VXIplug&play HP-UX Framework: 10.2 planned 1998 Note: The HP VEE application can use VXIplug&play drivers or HP panel drivers.
Cooling/Slot
Watts/slot: 20.6 DeltaP mm H2O: 0.18 Air flow liter/s: 2.8
Module Current
IPM (A) IDM (A) +5 V: 1.1 0.5 +12 V: 0.6 0.1 –12 V: 0.6 0.1 +24 V: 0 0.01 –24 V: 0 0.01 –5.2 V: 0.1 0.01 –2 V: 0.1 0.01
HP E1563A 800 kHz, 2-Channel VXI Digitizer Opt 0B1 Add Manual Set Opt W01 Convert Standard Warranty to On-site
Trademarks:
Windows is a U.S. registered trademark of Microsoft Corporation.
HP E1563A 800 kSa/s, 2 Channel Digitizer Technical Specifications
If you do not have the Adobe Acrobat Reader necessary for viewing PDF documentation, download your free Acrobat Reader now. |
![]() |
Related Information:
Related VXI Digitizers product datasheets:
Related product families available online:
Page Updated: Monday October 05 19:26:28 UTC 1998