HP E5050A Colloid Dielectric Probe |
The HP E5050A Colloid Dielectric Probe offers new capabilities for measuring liquids with bulk conductive properties in-situ and without sample preparation. The dielectric profile of a material often yields key parameters about the nature of colloids. Applications include: Uniformity and concentration of dispersed phase; particle size and shape; colloidal structure analysis, emulsification analysis.
Although dielectric spectroscopy is a well established technique, its use has so far been limited to non-conducting systems because of the electrode polarization problem. This limitation has been overcome because the E5050A uses an electrodeless system which provides permittivity and conductivity data from 200 kHz to 20 MHz.
The new sensing technique solves the electrode polarization problem which has limited dielectric spectroscopy to low-conductivity systems.
Perform measurements by simply putting the probe into the liquid material. The probe size and shape make it suitable for on-line monitoring and control applications.
Frequency range: 200 kHz to 20 MHz specified 75 kHz to 30 MHz operating Measurement Parameters: Relative Permittivity: e', e'' Conductivity : k', k'' Maximum Temperature 120 deg. C. Cable length 2 meters Probe size: Thickness = 14mm Outer Diameter = 40mm Inner Diameter = 7.5mm Length of Arm = 160mm Minimum distance of 20 mm between probe and vessel required.
A complete system consists of: HP E5050A Colloid Dielectric Probe Opt 001 Yeast Measurement software HP 82345B PC Automation Kit HP 4285A Precision LCR Meter HP Vectra PC or equivalent- The computer must be configured with a minimum of an 80486/33MHz or higher processor, 16MB or more of RAM, 20MB of free space on a hard disk, one 3.5" 1.44 MB floppy disk drive, and a 1024x768 super VGA monitor.
If you do not have the Adobe Acrobat Reader necessary for viewing this documentation, download your Free Acrobat Reader now. |
![]() |
Page Updated: Wednesday April 02 19:44:15 UTC 1997